Skip to content

GMRS 6495-2-040Recalls & reported problems

Howmedica Osteonics Corp. · Distal Femoral Component · Class 2 · Implant

All models of GMRS (32)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 07613327048780), received between Nov 5, 2018 and May 19, 2026 — classified as 14 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of Osseointegration” (5 reports).

2 other recalls cover devices cleared under the same FDA submission (K023087, K222056, K233261, K241716); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Loss of Osseointegration5
  • Corroded1
  • Insufficient Device Problem Information1
  • Mechanical Problem1
  • Migration1
  • Packaging Problem1
  • Unstable1

Patient problems

  • Inadequate Osseointegration5
  • Injury2
  • Insufficient Information2
  • Pain2
  • Bone Fracture(s)1
  • Joint Laxity1
  • Loss of Range of Motion1
  • Metal Related Pathology1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 19, 2026Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING INFORMATION WAS PROVIDED: REVISION OF GMRS DISTAL FEMORAL REPLACEMENT DUE TO FRACTURE OF THE PROXIMAL FEMUR.

    Patient problems: Insufficient Information

    MAUDE report 25225374 ↗
  • InjuryMay 11, 2026Corroded

    THE FOLLOWING INFORMATION WAS PROVIDED: IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO PATIENT COMPLAINT OF PAIN AND SWELLING. INTRA-OPERATIVELY, SURGEON REPORTED CORROSION BETWEEN THE GMRS FEMORAL COMPONENT AND THE STEM. THE FEMORAL COMPONENT AND AXLE WERE REVISED. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON. Q.

    Patient problems: Metal Related Pathology; Pain; Swelling/ Edema

    MAUDE report 25134399 ↗
  • InjuryNov 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION OF KNEE WITH GMRS DISTAL FEMUR INSITU DUE TO SURGEON REPORT OF SYNOVITIS. SEVERAL COMPONENTS WERE REPLACED, BUT NO REPORT OF COMPONENT INVOLVEMENT.

    Patient problems: Synovitis

    MAUDE report 23454881 ↗
  • InjuryAug 5, 2025Mechanical Problem; Mechanical Problem

    INFORMATION RECEIVED FROM (B)(4) INDICATES THE FOLLOWING: MECHANICAL FAILURE, PAIN. ALL COMPONENTS REMOVED. RIGHT KNEE.

    Patient problems: Pain

    MAUDE report 22713669 ↗
  • InjuryJul 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION RECEIVED FROM IIS 2023-001 INDICATES THE FOLLOWING: PERIPROSTHETIC FEMORAL FRACTURE. FEMORAL COMPONENT REVISED. RIGHT KNEE.

    Patient problems: Bone Fracture(s)

    MAUDE report 22680680 ↗
  • InjuryOct 11, 2023Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO LOOSENING AND SUBSIDENCE OF THE FEMORAL COMPONENT AND LOOSENING OF THE TIBIAL BASEPLATE. THE PATIENT'S DISTAL FEMORAL REPLACEMENT CONSTRUCT WAS REMOVED AND A COMPETITOR COMPRESS WAS IMPLANTED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17910725 ↗
  • InjuryJul 20, 2023Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED. AS REPORTED: "PATIENT REVISED DUE TO PATELLAR INSTABILITY. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY." A DISTAL FEMORAL COMPONENT, ROTATING HINGE, AXLE, BUSHINGS, BUMPER INSERT, TIBIAL INSERT, BASEPLATE, TIBIAL SLEEVE, AND STEM WERE REVISED (REP CONFIRMED PATELLAR COMPONENT WAS RETAINED).

    Patient problems: Joint Laxity

    MAUDE report 17359498 ↗
  • InjuryFeb 2, 2023Loss of Osseointegration; Loss of Osseointegration

    A PATIENT SPECIFIC IMPLANT REQUEST WAS RECEIVED FOR THE PATIENT'S RIGHT DISTAL FEMUR. NOTED ON THE FORM: "S/P RESECTION AND RECONSTRUCTION RIGHT DISTAL FEMUR WITH STRYKER GMRS AND NOW SECOND EPISODE OF ASEPTIC LOOSENING RETAINING 8MM LARGE ALL0POLY TIBIA." PROPOSED DATE OF SURGERY IS ON (B)(6) 2023.

    Patient problems: Inadequate Osseointegration

    MAUDE report 16282762 ↗
  • InjuryNov 22, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO DECREASED ROM. AN ARTHROFIBROSIS EXCISION (SOFT TISSUE REMOVAL) WAS PERFORMED AND THE DISTAL FEMORAL COMPONENT WAS DOWNSIZED FROM STANDARD TO SMALL.

    Patient problems: Loss of Range of Motion

    MAUDE report 15842760 ↗
  • InjuryNov 7, 2022Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THROUGH THE SUBMISSION OF A REVISION USAGE SHEET THAT THE PATIENT'S RIGHT KNEE WAS REVISED. A GMRS/ MRH CONSTRUCT WAS IMPLANTED. UPDATE (B)(6) 2022: SPOKE TO REP. REASON FOR REVISION WAS MIGRATION OF THE STEM AND SUSPECTED LOOSENING OF THE DISTAL FEMORAL COMPONENT (LOOSENING CONFIRMED INTRA-OPERATIVELY). THE STEM AND DISTAL FEMUR WERE REVISED TO THE SAME CATALOG NUMBERS. OTHER INTERNAL COMPONENTS (INSERTS, BUSHES, AXLE, ETC) WERE ALSO REVISED. REP CONFIRMED THERE ARE NO ALLEGATIONS AGAINST ANY OF THE OTHER REVISED COMPONENTS.

    Patient problems: Inadequate Osseointegration

    MAUDE report 15742473 ↗
  • InjuryJun 25, 2021Loss of Osseointegration; Migration; Loss of Osseointegration; Migration

    ASEPTIC MOBILIZATION DURING A SECONDARY IMPLANT FOR KNEE PROSTHESIS REVISION.

    Patient problems: Inadequate Osseointegration; Insufficient Information

    MAUDE report 12065105 ↗
  • InjuryDec 18, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S RIGHT KNEE WAS REVISED DUE TO STIFFNESS. ALL IMPLANTS EXCEPT THE MRH TIBIAL BASEPLATE AND STEM EXTENDER WERE REVISED; HOWEVER THERE ARE NO ALLEGATIONS AGAINST THE IMPLANTS AND STIFFNESS WAS ATTRIBUTED ONLY TO SCAR TISSUE. REP PROVIDED AN EXPLANT PICTURE AND PRIMARY AND REVISION USAGE SHEETS. THE REP CONFIRMED THAT NO FURTHER INFORMATION WOULD BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Scar Tissue

    MAUDE report 11035152 ↗
  • InjuryJan 24, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THROUGH THE SUBMISSION OF A REVISION IMPLANT SHEET THAT PATIENT'S RIGHT KNEE WAS REVISED. A NOTE AT THE BOTTOM OF THE SHEET REPORTS "REVISED FOR INSTABILITY AND POSSIBLE INFECTION". AN MRH KNEE CONSTRUCT WAS REVISED TO AN MRH/ GMRS KNEE CONSTRUCT.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 9628571 ↗
  • InjuryDec 13, 2019Loss of Osseointegration; Loss of Osseointegration

    PATIENT HAD A DISTAL FEMORAL REPLACEMENT 14 MONTHS AGO AND FEMORAL COMPONENTS BECAME LOOSE. PATIENT HAD PAIN. PATIENT REQUIRED REVISION SURGERY. UPDATE 21/NOVEMBER/2019 (B)(6): SPOKE TO REP. RIGHT FEMUR. THE FEMORAL COMPONENT, EXTENSION PIECE, FLUTED STEM, BUSHINGS AND BUMPER INSERT WERE REVISED. THERE ARE NO ALLEGATIONS AGAINST ANY DEVICES BESIDES THE LOOSE FEMORAL COMPONENT. REP PROVIDED THE PRIMARY IMPLANT REPORT AND REPORTED THAT NO FURTHER INFORMATION WILL BE AVAILABLE.

    Patient problems: Inadequate Osseointegration; Injury

    MAUDE report 9466841 ↗
  • MalfunctionNov 5, 2018Packaging Problem; Packaging Problem

    THIS PI IS FOR THE INTRA-OPERATIVE EVENT ON (B)(6) 2018. IT WAS REPORTED THAT DURING REVISION OF PATIENT'S RIGHT KNEE, THE OUTER RIM OF THE OUTER BLISTER WAS ATTACHED TO THE TYVEK AS THE IMPLANT WAS OPENED. A CORNER OF THE INNER BLISTER WAS NOT ADHERED TO THE TRAY. DUE TO STERILITY CONCERNS, THE SURGEON ELECTED TO RE-IMPLANT THE EXPLANTED COMPONENTS AS NO BETTER ALTERNATIVES WERE AT HAND.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8039223 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K023087, K222056, K233261, K241716. One submission often covers many models, so check the product and lot details in each record.

  • Z-2732-2024 · Aug 5, 2024 · Open, Classified
    Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;
  • Z-2624-2010 · May 5, 2005 · Terminated
    GMRS Extension Piece, 60 mm; Sterile Stryker Howmedica Osteonics; Howmedica Internation S. de. R.L. Limerick, Ireland. The Global Modular Replacement System is intended to be used

Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page