Adverse event reports
13
Deaths reported
0
Injuries reported
12
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 07613327048797), received between Sep 12, 2017 and Sep 9, 2026 — classified as 12 injuries, 1 malfunction.

The most frequently reported specific device problem is “Loss of Osseointegration” (5 reports).

2 other recalls cover devices cleared under the same FDA submission (K023087, K222056, K233261, K241716); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loss of Osseointegration5
  • Adverse Event Without Identified Device or Use Problem3
  • Degraded2
  • Device Contamination with Chemical or Other Material1
  • Failure to Osseointegrate1
  • Insufficient Device Problem Information1
  • Material Erosion1

Patient problems

  • Inadequate Osseointegration5
  • Arthrofibrosis2
  • Foreign Body Reaction2
  • Ambulation or Postural Difficulties1
  • Injury1
  • Muscle/Tendon Damage1
  • No Known Impact Or Consequence To Patient1
  • Pain1

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2026Degraded

    It was reported that the patient presented with poly failure and wear as well as apparent metallosis of GMRS taper junctions which was noticed upon patella fracture treatment.

    Patient problems: Foreign Body Reaction

    MAUDE report 50690352 ↗
  • InjuryAug 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION RECEIVED FROM (B)(4) INDICATES THE FOLLOWING: ARTHROFIBROSIS. ABOVE KNEE AMPUTATION PERFORMED. LEFT KNEE.

    Patient problems: Arthrofibrosis

    MAUDE report 22735532 ↗
  • InjuryAug 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS PI IS FOR THE SURGICAL INTERVENTION FOR ARTHROFIBROSIS. INFORMATION RECEIVED FROM IIS 2023-001 INDICATES THE FOLLOWING: WOUND INFECTION SUSPECTED (B)(6) 2013, IV VANC, NO RETURN TO OR. LATER NEEDED TO BE ON LIFELONG ORAL ANTIBIOTICS FOR CHRONIC SUPPRESION. SURGICAL INTERVENTION FOR ARTHROFIBROSIS (B)(6) 2015. LEFT KNEE.

    Patient problems: Arthrofibrosis

    MAUDE report 22724382 ↗
  • InjuryAug 5, 2025Loss of Osseointegration; Loss of Osseointegration

    INFORMATION RECEIVED FROM (B)(4) INDICATES THE FOLLOWING: ASEPTIC LOOSENING. ALL COMPONENTS REMOVED. LEFT KNEE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22713830 ↗
  • InjuryAug 5, 2025Loss of Osseointegration; Loss of Osseointegration

    INFORMATION RECEIVED FROM (B)(4) INDICATES THE FOLLOWING: ASEPTIC LOOSENING. ALL COMPONENTS REMOVED. LEFT KNEE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22713679 ↗
  • InjuryJul 31, 2025Loss of Osseointegration; Loss of Osseointegration

    INFORMATION RECEIVED FROM IIS 2023-001 INDICATES THE FOLLOWING: ASEPTIC LOOSENING. FEMORAL COMPONENT REVISED. LEFT KNEE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22680674 ↗
  • InjuryJul 31, 2025Loss of Osseointegration; Loss of Osseointegration

    THE FOLLOWING WAS REPORTED: "ASEPTIC LOOSENING. ALL COMPONENTS REMOVED. LEFT KNEE."

    Patient problems: Inadequate Osseointegration

    MAUDE report 22680520 ↗
  • InjuryMar 19, 2025Loss of Osseointegration; Loss of Osseointegration

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED. AS REPORTED: PATIENT WAS REVISED DUE TO LOOSEN FEMORAL COMPONENT. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21641660 ↗
  • InjuryJan 30, 2024Degraded; Material Erosion; Degraded; Material Erosion

    IT WAS REPORTED THAT DURING A REVISION OF GMRS/MRH, TRUNNIONOSIS WAS SEEN ON THE GMRS TAPER JUNCTIONS. PRIOR TO REVISION, PREOPERATIVE JOINT ASPIRATION YIELDED DARK FLUID CONSISTENT WITH METALLOSIS.

    Patient problems: Ambulation or Postural Difficulties; Foreign Body Reaction

    MAUDE report 18605226 ↗
  • InjuryJun 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED. AS REPORTED: "PT. PRESENTED WITH A TENDON-RUPTURE OF THE PATELLAR-TENDON/EXTENSOR-MECHANISM. THE DOCTOR OPTED TO SWAP OUT SOME COMPONENTS, WHILE PERFORMING AN I&D, AND EXTENSOR-MECHANISM REPAIR. NO COMPLAINTS HAVE BEEN FILED AGAINST THE COMPONENTS THEMSELVES. NO FURTHER INFO AVAILABLE."

    Patient problems: Muscle/Tendon Damage

    MAUDE report 17069109 ↗
  • InjuryJun 16, 2022Failure to Osseointegrate; Failure to Osseointegrate

    IT WAS REPORTED THAT THE PATIENT'S LEFT KNEE WAS REVISED. AS REPORTED: "PT. PRESENTED WITH TIBIAL TENDERNESS/PAIN CONSIDERED INDICATIVE OF A LOOSE CEMENTED-IMPLANT. IMAGING RESULTS LEAD SURGEON TO REVISE THE PATIENT¿S GMRS DISTAL-FEMUR CONSTRUCT. NO COMPLAINTS FILED AGAINST THE IMPLANTS THEMSELVES." SPOKE TO REP, WHO PROVIDED USAGE SHEETS. NOTHING WAS OBSERVED INTRA-OPERATIVELY TO BE WORN OR BROKEN. REP CONFIRMED NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Pain

    MAUDE report 14723084 ↗
  • MalfunctionMay 10, 2018Device Contamination with Chemical or Other Material; Device Contamination with Chemical or Other Material

    PRIMARY PROCEDURE, LEFT FEMUR. IT WAS REPORTED THAT IN OPENING THE IMPLANT, A FINE WHITE POWDER WAS WITNESSED ON THE IMPLANT. DUE TO STERILITY CONCERNS, THE IMPLANT WAS NOT USED. ANOTHER IMPLANT WAS SECURED AND OPENED WHILE THE CEMENT WAS SETTING, WITH NO SURGICAL DELAY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7502468 ↗
  • InjurySep 12, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    PATIENT UNDERWENT AN INDEX AND REVISION SURGERIES OUTSIDE ROTHMAN INSTITUTE. ON (B)(6) 2017, THE PATIENT WAS REVISED FOR TKA LOOSENING. WITH EXCHANGE OF ALL COMPONENTS TO A HINGE TKA. 5 MONTHS LATER A NEW REVISION IS PERFORMED FOR INSTABILITY, EXCHANGED ALL ROTATING HINGE COMPONENTS, AND DISTAL FEMORAL COMPONENT.

    Patient problems: Injury

    MAUDE report 6859829 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K023087, K222056, K233261, K241716. One submission often covers many models, so check the product and lot details in each record.

  • Z-2732-2024 · Aug 5, 2024 · Open, Classified
    Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;
  • Z-2624-2010 · May 5, 2005 · Terminated
    GMRS Extension Piece, 60 mm; Sterile Stryker Howmedica Osteonics; Howmedica Internation S. de. R.L. Limerick, Ireland. The Global Modular Replacement System is intended to be used

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page