Skip to content

GMRS 6495-6-080Recalls & reported problems

Howmedica Osteonics Corp. · Extension Piece · Class 2 · Implant

All models of GMRS (32)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613327049220), received between Jan 26, 2018 and Jun 24, 2026 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K023087, K222056, K233261, K241716); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Degraded1
  • Loss of Osseointegration1

Patient problems

  • Injury3
  • No Consequences Or Impact To Patient2
  • Foreign Body Reaction1
  • Inadequate Osseointegration1
  • Insufficient Information1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 24, 2026Degraded; Loss of Osseointegration

    THE FOLLOWING INFORMATION WAS PROVIDED: THE PRIMARY PROCEDURE OF GMRS WAS DONE IN 2004. REVISION SURGERY WAS CARRIED OUT ON (B)(6) 2026 DUE TO SUSPECTED TIBIAL LOOSENING. HOWEVER, WHEN DR KC WONG OPEN THE JOINT CAPSULE, MASSIVE AND EXTENSIVE METALLOSIS WAS SEEN AROUND THE PROSTHESIS. WHEN WE TRY TO IMPACT A NEW 80MM EXTENSION PIECE, WE FOUND THAT THE MALE TAPER WAS MUCH LARGER THAN THE FEMALE TAPER OF THE EXT. PIECE. WE STRUGGLED THAT IF WE NEED TO REMOVE THE IN-SITU STEM SINCE WE HAD NO IDEA WHY THE TAPER LOOKED DIFFERENT. SOMEHOW A NURSING STAFF SUGGESTED THAT PART OF THE FEMALE TAPER OF THE IN-SITU 80MM EXT. PIECE MIGHT BE PEELED OFF AND COLDWELDED TO THE STEM'S MALE TAPER. THEREFORE THE SURGEONS TRIED TO USE METAL CUTTING DEVICE TO CUT A BIT ON THE TAPER. FINALLY WE MANAGED TO TAKE AWAY THE PEELED OFF PART AND EXPOSED THE ORIGINAL MALE TAPER OF THE STEM BUT IT EXTENDED THE SURGERY FOR ALMOST 3/4 HOURS. CONCLUSION IS THAT EXTENSION PIECE MIGHT HAVE A FAILURE TRUNNION.

    Patient problems: Foreign Body Reaction; Inadequate Osseointegration; Insufficient Information

    MAUDE report 25610317 ↗
  • InjuryJan 24, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THROUGH THE SUBMISSION OF A REVISION IMPLANT SHEET THAT PATIENT'S RIGHT KNEE WAS REVISED. A NOTE AT THE BOTTOM OF THE SHEET REPORTS "REVISED FOR INSTABILITY AND POSSIBLE INFECTION". AN MRH KNEE CONSTRUCT WAS REVISED TO AN MRH/ GMRS KNEE CONSTRUCT.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 9628290 ↗
  • MalfunctionFeb 27, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    EXPLANTATION WAS NECESSARY BECAUSE OF A FISSURE OF THE DISTAL FEMUR. DURING EXPLANTATION, THE INNER BUSHING LOOSENED FROM THE GMRS EXTENSION PIECE. EXTENSION WAS COATED IN (B)(6) 2017. NEW EXTENSION WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY.

    Patient problems: Injury; No Consequences Or Impact To Patient

    MAUDE report 7300580 ↗
  • MalfunctionFeb 27, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    DURING A REVISION, THE INNER BUSHING LOOSENED FROM THE GMRS EXTENSION PIECE. NEW EXTENSION WAS USED TO COMPLETE THE PROCEDURE SUCCESSFULLY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7300567 ↗
  • InjuryJan 26, 2018Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    IN REPORTING A REVISION OF PATIENT'S LEFT KNEE ON (B)(6) 2018, REP WAS MADE AWARE OF A PREVIOUS REVISION ON (B)(6) 2016 (REP WAS NOT PRESENT FOR PROCEDURE). IN THAT PROCEDURE, AN 80MM EXTENSION PIECE WAS IMPLANTED. ON (B)(6) 2018, A 100MM EXTENSION PIECE WAS EXPLANTED. REP IS UNAWARE OF WHEN THE 80MM EXTENSION PIECE WAS REVISED. THIS EVENT IS FOR REVISION 1 WHICH TOOK PLACE SOME TIME BETWEEN (B)(6) 2016 AND (B)(6) 2018.

    Patient problems: Injury

    MAUDE report 7222700 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K023087, K222056, K233261, K241716. One submission often covers many models, so check the product and lot details in each record.

  • Z-2732-2024 · Aug 5, 2024 · Open, Classified
    Custom Made Device - GMRS EXTENSION Piece 40MM LEFT, 35 Deg EXT Rotated Tabs (Patient Specific); Catalog Number: C-MM100434-00;
  • Z-2624-2010 · May 5, 2005 · Terminated
    GMRS Extension Piece, 60 mm; Sterile Stryker Howmedica Osteonics; Howmedica Internation S. de. R.L. Limerick, Ireland. The Global Modular Replacement System is intended to be used

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources and updates