Skip to content

Scorpio 73-0110Recalls & reported problems

Howmedica Osteonics Corp. · Resurfacing Patella Component · Class 2 · Implant

All models of Scorpio (107)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

2 FDA recall records name this device's identifier, 2 still open.

12 other recalls cover devices cleared under the same FDA submission (K962152); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Jul 30, 2026Open, Classified

    Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

    Z-3197-2026 · Root cause: Process change control · Recall details · FDA record ↗

  2. Dec 1, 2023Open, Classified

    Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.

    Z-0886-2024 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K962152. One submission often covers many models, so check the product and lot details in each record.

  • Z-0890-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #11 -Intended for knee replacement Part Number: 73-2110
  • Z-0887-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #5 -Intended for knee replacement Part Number: 73-0510
  • Z-0891-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-2710
  • Z-0888-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #7 -Intended for knee replacement Part Number: 73-0710
  • Z-0892-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TOTAL KNEE CONCENTRIC DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-2910
  • Z-0889-2024 · Dec 1, 2023 · Open, Classified
    SCORPIO TOTAL KNEE MEDIALIZED DOME PATELLA-SZ #9 -Intended for knee replacement Part Number: 73-0910
  • Z-1634-2015 · Aug 15, 2014 · Terminated
    Stryker Orthopaedics Scorpio femoral component. Howmedica Osteonics Corp., A subsidiary of Stryker Corporation Made in USA Sterile. For use in total reconstructive replacement of
  • Z-0469-2010 · Oct 2, 2009 · Terminated
    Scorpio PS Femur Waffle Posts with LfIt; Stryker Orthopaedics, 325 Corporate Drive, Mahwah, NJ 07430. The Scorpio Femoral Components are a knee component that is utilized in total
  • Z-0795-2010 · Jul 30, 2007 · Terminated
    Scorpio PS Femoral Waffle (No post) and LFIT; Use only with Scorpio PS Tibial Bearing inserts and Scorpio Patellas, Stryker Orthopaedics. Intended for cemented application to repl
  • Z-1814-2008 · Nov 8, 2005 · Terminated
    Scorpio Total Knee Posteriorly Stabilized Femoral Component, #7R, Catalog No.: 71-4107R, Lot Code: K05L310; Scorpio Total Knee Posteriorly Stabilized Femoral Component, #7L, Catalo
  • Z-2079-2008 · Mar 17, 2005 · Terminated
    Osteonics Scorpio Total Knee; Posteriorly Stabilized Femoral Component; #13 with posts, low friction Ion treatment. Use only with Scorpio Posteriorly Stabilized Tibial Bearing Com
  • Z-1382-2008 · Feb 8, 2005 · Terminated
    Scorpio Total Knee Posteriorly Stabilized Femoral Component; Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis; Product Number: 71-4507R; St

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page