Tritanium PL 48950096Recalls & reported problems
Stryker Corporation · Posterior Lumbar Cage · Class 2 · Implant
All models of Tritanium PL (45)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327117318), received between Mar 18, 2018 and Jun 9, 2023 — classified as 5 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Fracture5
- Migration3
- Device Dislodged or Dislocated2
- Appropriate Device Problem Term/Code Not Available1
- Improper or Incorrect Procedure or Method1
- Insufficient Device Problem Information1
- Unintended Movement1
Patient problems
- Injury3
- No Clinical Signs, Symptoms or Conditions3
- Failure of Implant1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
- Numbness1
- Pain1
- Peritoneal Laceration(s)1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2023Improper or Incorrect Procedure or Method; Insufficient Device Problem Information; Improper or Incorrect Procedure or Method; Insufficient Device Problem Information
IT WAS REPORTED THAT, DURING IMPLANTATION OF A TRITANIUM POSTERIOR LUMBAR CAGE, THE SURGEON PENETRATED THE ANTERIOR LONGITUDINAL LIGAMENT. THE CAGE WAS LEFT IN PLACE. THE SURGEON HAS PLANNED ADDITIONAL IMAGING AND ANGIOGRAPHY.
Patient problems: Peritoneal Laceration(s)
MAUDE report 17097521 ↗InjuryJun 5, 2023Migration; Migration
A COMPANY REPRESENTATIVE REPORTED THAT A PATIENT UNDERWENT REVISION SURGERY APPROXIMATELY 6 MONTHS POST-OPERATIVELY TO ADDRESS A MIGRATED TRI PL CAGE AND A 'LOOSENED' XIA POLYAXIAL SCREW. THIS REPORT IS FOR THE TRI PL CAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17059246 ↗MalfunctionMar 4, 2022Fracture; Fracture
A PHYSICIAN REPORTED THAT A TRITANIUM PL CAGE AT L5 FRACTURED ONE YEAR POSTOPERATIVELY. REVISION SURGERY HAS NOT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13667299 ↗MalfunctionNov 10, 2020Fracture; Fracture
IT WAS REPORTED THAT A TRITANIUM PL CAGE FRACTURED INTRA-OPERATIVELY DURING IN SITU CAGE PLACEMENT AT L5-S1. THE REPRESENTATIVE REPORTED THAT THE DISC SPACE WAS NOT DISTRACTED AND TRIALING WAS NOT PERFORMED PRIOR TO IMPLANT PLACEMENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AND ALL THE FRACTURED PIECES WERE REMOVED FROM THE SURGICAL SITE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 25 MINUTE SURGICAL DELAY BY USING AN ALTERNATIVELY AVAILABLE CAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10819823 ↗MalfunctionJun 15, 2020Fracture; Fracture
IT WAS REPORTED THAT AFTER INTRA-OPERATIVE PLACEMENT OF A TRITANIUM PL CAGE, THE SURGEON DECIDED TO CHANGE THE CAGE SIZE. DURING REMOVAL OF THE CAGE WITH THE INSERTER, THE CAGE FRACTURED AND PART OF IT REMAINED INSIDE THE PATIENT WHILE THE OTHER PART CAME OUT. THE FRACTURED CAGE WAS LEFT IN THE PATIENT AND THE INTERVERTEBRAL SPACE WAS PACKED WITH AUTOLOGOUS BONE TO PREVENT IT FROM MOVING. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10155564 ↗InjuryNov 12, 2019Migration; Migration
PHYSICIAN REPORTED THAT A PATIENT EXPERIENCED NUMBNESS AND IT WAS CONFIRMED VIA X-RAY THAT A TRITANIUM PL CAGE PLACED IN L4/5 MIGRATED POST-OPERATIVELY. REVISION SURGERY WAS PERFORMED.
Patient problems: Injury; Numbness
MAUDE report 9312058 ↗InjuryJun 21, 2019Fracture; Device Dislodged or Dislocated
PRIMARY SURGERY USING ES2 SYSTEM WAS PERFORMED (B)(6) 2017. IT WAS REPORTED THAT THE SCREW DEVIATED FROM THE VERTEBRAL BODY POSTOPERATIVELY AND THE CAGE FRACTURED AT L3/4. PATIENT WAS REVISED ON (B)(6) 2019. THIS RECORD REPRESENTS THE SCREW.
Patient problems: Injury
MAUDE report 8724911 ↗MalfunctionJun 8, 2018Device Dislodged or Dislocated; Unintended Movement; Migration; Migration
IT WAS REPORTED THAT WHEN THE CAGE WAS INSERTED, IT PROTRUDED EXTERIORLY BECAUSE IT WAS INSERTED TOWARDS THE EXTERIOR. THE INSERTER WAS RE-ATTACHED TO THE CAGE AND REPOSITIONED. HOWEVER, THE CAGE WAS DROPPED OUT FROM THE VERTEBRAL BODY POST OPERATIVELY. CURRENTLY THE PATIENT HAS NO SUBJECTIVE SYMPTOMS, SO THE SURGEON DECIDED NOT TO TAKE ADDITIONAL TREATMENT.
Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient
MAUDE report 7583768 ↗InjuryMar 18, 2018Fracture; Fracture; Appropriate Device Problem Term/Code Not Available
A PATIENT WHO WAS IMPLANTED WITH A CAGE IN (B)(6), REPORTED PAIN AFTER THE OPERATION. IT WAS FURTHER REPORTED THAT THE PATIENT COULD NOT WALK AT ALL, A "BLACK SHADOW LIKE A FOREIGN OBJECT" RESPONSE WAS SEEN ON CT IMAGES, SYMPTOMS IMPROVED AS A RESULT OF BLOCK INJECTIONS ON THE INTERVERTEBRAL DISC AND THAT "THE INTERVERTEBRAL DISCS ARE DOING BAD THINGS". THE CAGE WAS REMOVED ON (B)(6) DURING A REVISION SURGERY. IT WAS OBSERVED THAT THE CASE DID NOT FUSE. AFTER PALPATING THE AFFECTED AREA, AFTER REMOVAL, THERE WAS NO CAVITY AND NORMAL.
Patient problems: Injury; Pain
MAUDE report 7347922 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page