Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327117981), received between Feb 10, 2017 and Jul 2, 2024 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Fracture1
- Material Fragmentation1
- Material Integrity Problem1
- Migration1
Patient problems
- Failure of Implant1
- Injury1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 2, 2024Migration; Migration
B5: A PATIENT WAS REVISED TO ADDRESS A MIGRATED TRITANIUM PL CAGE AT L4/L5 APPROXIMATELY FOUR MONTHS AFTER IMPLANTATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19656398 ↗InjuryNov 21, 2017Fracture
IT WAS REPORTED THAT; LUMBAR CAGE FRACTURED IN HALF INSITU AT L5-S1. AS A RESULT, PATIENT WAS BROUGHT BACK FOR REVISION ALIF SURGERY.
Patient problems: Failure of Implant; Injury; Pain
MAUDE report 7049713 ↗MalfunctionFeb 10, 2017Break; Material Fragmentation; Material Integrity Problem; Break; Material Integrity Problem
IT IS REPORTED THAT THE SURGEON DID A COUPLE TAPS DURING INSERTION OF THE CAGE AND TRIED TO TURN AND STEER THE CAGE. THE NOSE OF THE CAGE WAS IN THE DISC SPACE WAS UNDER LOAD AND THE REST OF THE CAGE WAS NOT IN. WHEN ROTATING AND STEERING THE CAGE, IT GAVE WAY BEHIND THE NOSE AND IT BROKE INTO MULTIPLE FRAGMENTS THAT DID NOT FALL IN THE PATIENT CAVITY. ALL FRAGMENTS WERE RETRIEVED. USED 1 MM SMALLER SIZED CAGE TO COMPLETE THE PROCEDURE (10 MM) PROCEDURE WAS COMPLETED SUCCESSFULLY. NO ADVERSE CONSEQUENCES OF THE PATIENT REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6322395 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page