Tritanium PL 48955116Recalls & reported problems
Stryker Corporation · Posterior Lumbar Cage · Class 2 · Implant
All models of Tritanium PL (45)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327118094), received between Mar 12, 2018 and Apr 19, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Break1
- Material Fragmentation1
- Migration1
- Migration or Expulsion of Device1
Patient problems
- Injury1
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
- Pain1
- Vertebral Fracture1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2024Fracture; Fracture
IT WAS REPORTED THAT A TRITANIUM POSTERIOR LUMBAR CAGE FRACTURED DURING INSERTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A 10 MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Vertebral Fracture
MAUDE report 19144567 ↗InjuryOct 8, 2019Migration or Expulsion of Device; Migration; Migration
A PHYSICIAN REPORTED REVISION SURGERY TO EXPLANT A MIGRATED TRITANIUM PL CAGE AND A FRACTURED ES2 INTEGRATED BLADE SCREW. THE PATIENT REPORTED EXPERIENCING PAIN. THIS REPORT CAPTURES THE TRITANIUM PL CAGE.
Patient problems: Injury; Pain
MAUDE report 9167052 ↗MalfunctionMar 12, 2018Fracture; Break; Fracture; Material Fragmentation
IT WAS REPORTED THAT A SURGEON WAS PERFORMING A SINGLE LEVEL T-LIFT, L2-L3. DISCECTOMY WAS COMPLETED WITH NO ISSUE. THE SURGEON PREPARED END PLATES AND USED PROPER INSERTER TO INSERT INTERBODY BY LIGHTLY TAPPING WITH MALLET INSERTER TO PUT IT IN PROPER POSITION. SURGEON CHECKED FLUORO AND OBSERVED INTERBODY APPEARED TO TURN OBLIQUELY 60 DEGREES. SURGEON REMOVED THE INSERTER WITH INTERBODY AND THE TOP 1/3 ATTACHED TO THE INSERTER, WHILE THE REMAINING 2/3 SHATTERED IN PIECES IN THE PATIENT'S BODY. ALL PIECES WERE REMOVED AND SURGEON TOOK ADDITIONAL FLUORO TO CONFIRM ALL PARTICLES WHERE REMOVED. SURGEON COMPLETED SURGERY WITH A NEW DEVICE AND COMPLETED THE SURGERY SUCCESSFULLY. THERE WAS A REPORTED SURGICAL DELAY OF 1 HOUR 15 MINUTES. THERE WERE NO ADVERSE HEALTH CONSEQUENCES TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7331387 ↗
Other Intervertebral fusion device with bone graft, lumbar devices
- T-PAL 03.812.003Synthes GmbH · Intervertebral fusion device with bone graft, lumbar152 reports
- T-PAL 03.812.004Synthes GmbH · Intervertebral fusion device with bone graft, lumbar120 reports
- Catalyft PL Expandable Interbody System 6068073Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall66 reports
- Catalyft PL Expandable Interbody System 6069076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall51 reports
- Catalyft PL Expandable Interbody System 6068076Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall46 reports
- EIT TLIF Inserter TFT30101EIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Lumbar88 reports
- T-PAL 03.812.001Synthes GmbH · Intervertebral fusion device with bone graft, lumbar76 reports
- Catalyft PL Expandable Interbody System 6069096Medtronic Sofamor Danek, Inc. · Intervertebral fusion device with bone graft, lumbar1 recall25 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page