Aero-C 48892066Recalls & reported problems
Stryker Corporation · Cervical Cage · Class 2 · Implant
All models of Aero-C (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327124651), received between Mar 9, 2016 and Jun 9, 2021 — classified as 1 injury, 8 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Mechanical Problem4
- Detachment of Device or Device Component3
- Device Dislodged or Dislocated2
- Device Slipped2
- Migration or Expulsion of Device2
- Detachment Of Device Component1
- Device Operates Differently Than Expected1
- Failure to Advance1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- No Known Impact Or Consequence To Patient4
- Injury1
- No Code Available1
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 9, 2021Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING INTRA-OPERATIVE IMPLANT PLACEMENT, THE BLADES OF THREE AERO C CAGES WERE NOT ENGAGING PROPERLY. AT C5/6 THE BLADES OF TWO CAGES WOULD NOT SEAR, AND AT C6/7 THE SUPERIOR BLADE PLUNGED DEEPER THAN EXPECTED. ALL THREE CAGES WERE REMOVED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ALTERNATIVELY AVAILABLE IMPLANTS. THIS RECORD CAPTURES THE FIRST OF THE THREE CAGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11971307 ↗MalfunctionJun 9, 2021Detachment of Device or Device Component; Unintended Movement; Detachment of Device or Device Component; Unintended Movement
IT WAS REPORTED THAT DURING INTRA-OPERATIVE IMPLANT PLACEMENT, THE BLADES OF THREE AERO C CAGES WERE NOT ENGAGING PROPERLY. AT C5/6 THE BLADES OF TWO CAGES WOULD NOT SEAR, AND AT C6/7 THE SUPERIOR BLADE PLUNGED DEEPER THAN EXPECTED. ALL THREE CAGES WERE REMOVED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ALTERNATIVELY AVAILABLE IMPLANTS. THIS RECORD CAPTURES THE THIRD OF THE THREE CAGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11971273 ↗MalfunctionJun 9, 2021Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING INTRA-OPERATIVE IMPLANT PLACEMENT, THE BLADES OF THREE AERO C CAGES WERE NOT ENGAGING PROPERLY. AT C5/6 THE BLADES OF TWO CAGES WOULD NOT SEAR, AND AT C6/7 THE SUPERIOR BLADE PLUNGED DEEPER THAN EXPECTED. ALL THREE CAGES WERE REMOVED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ALTERNATIVELY AVAILABLE IMPLANTS. THIS RECORD CAPTURES THE SECOND OF THE THREE CAGES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11971190 ↗MalfunctionJun 22, 2020Material Separation; Material Separation
IT WAS REPORTED THAT AN AERO C CAGE ANCHOR DISENGAGED INTRA-OPERATIVELY. AN X-RAY WAS TAKEN DURING THE PROCEDURE TO CONFIRM THE POSITION OF THE AERO-C ANCHOR AND IT WAS DISCOVERED THAT THE ANCHOR WAS IMPLANTED BEYOND THE POSTERIOR END OF THE AERO C CAGE. THE CAGE AND THE ANCHOR WERE REMOVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY BY USING ALTERNATE DEVICES.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10182033 ↗MalfunctionJun 30, 2017Mechanical Problem; Mechanical Problem; Migration or Expulsion of Device; Retraction Problem
IT WAS REPORTED THAT; THE BLADE WOULD NOT GO INTO THE BODY AND CAGE KEPT MIGRATING INTERIORLY. CAGE WAS LEFT IMPLANTED VERY ANTERIOR AND HAD TO USE EXCESSIVE MALLETING. 2 ANCHORS WERE IMPLANTED. AFTER EXCESSIVE FORCE THE BLADE WAS FINALLY IMPLANTED. ADVERSE CONSEQUENCES REMAINS TO BE SEEN. PATIENT HAD HARD BONE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6681579 ↗MalfunctionFeb 3, 2017Detachment Of Device Component; Mechanical Problem; Device Slipped; Device Operates Differently Than Expected; Detachment Of Device Component; Mechanical Problem
IT WAS REPORTED THAT; USING AN IMPLANT TODAY FOR A CERVICAL FUSION, ONE OF THE ANCHORS WENT THROUGH THE STOP FEATURE THAT PREVENTS THE ANCHOR FROM MIGRATING POSTERIORLY, THE DOCTOR LEFT AS IS, NO REPLACEMENT DEVICE WAS USED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6301226 ↗InjuryAug 25, 2016Device Slipped; Device Dislodged or Dislocated; Positioning Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT; THE BLADE FROM CAME OUT OF THE IMPLANT AND WAS RESTING CLOSE TO THE PATIENT'S ESOPHAGUS.
Patient problems: Injury; No Information
MAUDE report 5903384 ↗MalfunctionJun 3, 2016Migration or Expulsion of Device; Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT WHILE IMPACTING THE SUPERIOR BLADE OF THE CAGE, IT WAS BEING PUSHED POSTERIORLY. AT THE TIME THE BLADE LOCKED INTO THE CAGE, THE POSTERIOR PORTION OF THE CAGE WAS NOT FLUSH W/ THE VERTEBERBRAL BODY, AS IT WAS WHEN THE PUNCH AWLS WERE ALL THE WAY IMPACTED.
Patient problems: No Code Available; No Known Impact Or Consequence To Patient
MAUDE report 5699055 ↗MalfunctionMar 9, 2016Mechanical Jam; Failure to Advance; Mechanical Jam
IT WAS REPORTED THAT DURING THE INSERTION OF THE IMPLANT, THE BLADE/ ANCHOR WOULD NOT ADVANCE. THE SURGEON WAS FORCIBLY MALLETING THE ANCHOR TAMP AND THE BLADE WOULD NOT ADVANCE INTO THE SUPERIOR VERTEBRAL BODY (C5). THE ENTIRE DEVICE HAD TO BE REMOVED USING THE REMOVAL TOOL AND AN ALTERNATE DEVICE WAS USED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5489814 ↗
Other Intervertebral fusion device with bone graft, cervical devices
- EIT CIF cage CUI8070SEIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Cervical1 recall4 reports
- ROI-C MC1005TLDR Spine Usa, Inc. · Intervertebral Fusion Device With Bone Graft, Cervical54 reports
- EIT CIF cage CUI8060LEIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Cervical1 recall1 report
- CONSTRUX - MINI PTC 37-5012SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI Ti 37-6211SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI Ti 37-7308SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI Ti 37-9207SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI Ti 37-6106SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page