XIA 4.5 System - XIA CT 488515030Recalls & reported problems
Stryker Corporation · Polyaxial Cortical Screw · Class 2 · Implant
All models of XIA 4.5 System - XIA CT (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327134148), received between Mar 2, 2018 and Sep 3, 2020 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment Of Device Component1
- Device Dislodged or Dislocated1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 3, 2020Material Separation; Material Separation
A PHYSICIAN REPORTED THAT THE, "HEAD POPPED OFF THE SHANK," OF A XIA CT POLYAXIAL SCREW APPROXIMATELY FIFTEEN-MONTHS POST-OPERATIVELY. THE PATIENT HAS RECEIVED MEDICAL INTERVENTION.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Information
MAUDE report 10489416 ↗MalfunctionMar 2, 2018Device Dislodged or Dislocated; Detachment Of Device Component; Device Dislodged or Dislocated
IT WAS REPORTED THAT DURING AN L4-L5 FUSION, ALL SCREWS HAD BEEN SUCCESSFULLY IMPLANTED. THE SURGEON THEN TOOK THE PARALLEL POSTS AND POST DISTRACTOR TO DISTRACT AND PERFORM TLIF. HE THREADED THE POST ONTO THE L4 SCREW AND BEGAN TO TIGHTEN POST ONTO SCREW, AND THE TULIP HEAD ON THE SCREW DISENGAGED. A SCREWDRIVER WAS THEN USED TO REMOVE SHAFT OF SCREW AND A NEW SCREW WAS REINSERTED. SURGEON DISTRACTED OFF CONTRA-LATERAL SCREWS AND LOCKED A ROD IN PLACE TO COMPLETE DISCECTOMY AND TLIF. THE PROCEDURE WAS SUCCESSFULLY COMPLETED AND THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7311380 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page