Operon OT 8602165Recalls & reported problems

Berchtold Holding GmbH · Berchtold Surgical Table, D860 · Class 1

All models of Operon (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327170122), received between Nov 30, 2022 and Jul 29, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Unintended Movement” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Unintended Movement2
  • Defective Component1
  • Device Fell1
  • Key or Button Unresponsive/not Working1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Neck Pain1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 29, 2025Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE D860 TABLE MOVED ON ITS OWN IN A TILT POSITION WITH THE HAND PENDANT DETACHED AND THE POWER CORD UNPLUGGED. THE PATIENT'S HEAD WAS SECURED WITH A HEAD-HOLDING APPARATUS. AS THE TABLE TILTED AND SLOWLY LOWERED, THE PATIENT'S HEAD WAS STILL SECURED, RESULTING IN THE PATIENT'S NECK BEING STRAINED. THE PATIENT IS DOING OKAY BUT WOKE UP WITH A SORE NECK. NO MEDICAL INTERVENTION WAS REQUIRED FOR THE REPORTED SORE NECK.

    Patient problems: Neck Pain

    MAUDE report 22654290 ↗
  • MalfunctionSep 8, 2023Unintended Movement; Device Fell; Unintended Movement; Device Fell

    IT WAS REPORTED THAT IN THE MIDDLE OF THE CASE WHILE THE TABLE WAS IN TREND, WHEN THEY WENT TO LEVEL THE TABLE IT DROPPED TO LOWEST SETTING. IT WAS CLOSE TO A FOOT DROP. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17715955 ↗
  • MalfunctionNov 30, 2022Mechanical Problem; Defective Component; Key or Button Unresponsive/not Working; Defective Component

    IT WAS REPORTED THAT AFTER PROCEDURE STARTED, WHEN PUTTING PATIENT IN REVERSE TRENDELENBURG, THE BED MALFUNCTIONED, AND THE CONTROL COMMAND BUTTON FAILED TO WORK PROPERLY. PATIENT WAS TRANSFERRED TO ANOTHER BED TO FINISH PROCEDURE. NO HARM TO PATIENT. BED REMOVED FROM SERVICE. THERE WERE NO REPORTED ADVERSE CONSEQUENCES DUE TO THE DELAY OR ANY REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15889876 ↗

Other TABLE, OPERATING-ROOM, ELECTRICAL devices

All TABLE, OPERATING-ROOM, ELECTRICAL devices →