Operon OT 8602165Recalls & reported problems
Berchtold Holding GmbH · Berchtold Surgical Table, D860 · Class 1
All models of Operon (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327170122), received between Nov 30, 2022 and Jul 29, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement2
- Defective Component1
- Device Fell1
- Key or Button Unresponsive/not Working1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Neck Pain1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2025Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE D860 TABLE MOVED ON ITS OWN IN A TILT POSITION WITH THE HAND PENDANT DETACHED AND THE POWER CORD UNPLUGGED. THE PATIENT'S HEAD WAS SECURED WITH A HEAD-HOLDING APPARATUS. AS THE TABLE TILTED AND SLOWLY LOWERED, THE PATIENT'S HEAD WAS STILL SECURED, RESULTING IN THE PATIENT'S NECK BEING STRAINED. THE PATIENT IS DOING OKAY BUT WOKE UP WITH A SORE NECK. NO MEDICAL INTERVENTION WAS REQUIRED FOR THE REPORTED SORE NECK.
Patient problems: Neck Pain
MAUDE report 22654290 ↗MalfunctionSep 8, 2023Unintended Movement; Device Fell; Unintended Movement; Device Fell
IT WAS REPORTED THAT IN THE MIDDLE OF THE CASE WHILE THE TABLE WAS IN TREND, WHEN THEY WENT TO LEVEL THE TABLE IT DROPPED TO LOWEST SETTING. IT WAS CLOSE TO A FOOT DROP. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17715955 ↗MalfunctionNov 30, 2022Mechanical Problem; Defective Component; Key or Button Unresponsive/not Working; Defective Component
IT WAS REPORTED THAT AFTER PROCEDURE STARTED, WHEN PUTTING PATIENT IN REVERSE TRENDELENBURG, THE BED MALFUNCTIONED, AND THE CONTROL COMMAND BUTTON FAILED TO WORK PROPERLY. PATIENT WAS TRANSFERRED TO ANOTHER BED TO FINISH PROCEDURE. NO HARM TO PATIENT. BED REMOVED FROM SERVICE. THERE WERE NO REPORTED ADVERSE CONSEQUENCES DUE TO THE DELAY OR ANY REPORTED PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15889876 ↗
Other TABLE, OPERATING-ROOM, ELECTRICAL devices
- NA 72203299Smith & Nephew, Inc. · TABLE, OPERATING-ROOM, ELECTRICAL298 reports
- TS 7000 1723633Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall22 reports
- ALLEN 107003Hill-Rom, Inc. · Table, operating-room, electrical1 recall9 reports
- Head-Positioning accessories 1739994Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall1 report
- TS 7500 2067886Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- ALLEN 107003Hill-Rom, Inc. · Table, operating-room, electrical1 recall0 reports
- TS 7500 4091000Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 1717021Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page