Operon OT 8602065Recalls & reported problems
Berchtold Holding GmbH · Berchtold Surgical Table, D 860 · Class 1
All models of Operon (11)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07613327170160), received between Apr 23, 2021 and Jun 24, 2026 — classified as 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Unintended Movement” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Unintended Movement3
- Use of Device Problem2
- Component or Accessory Incompatibility1
- Unintended Collision1
- Unintended System Motion1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Laceration(s)1
- Unspecified Tissue Injury1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 24, 2026Unintended Collision
IT WAS REPORTED THAT IN A SURGERY WHEN TRYING TO LOWER THE BED IN REVERSE TREND, THE D860 TABLE HEIGHT DROPPED OUT FROM UNDER THE PATIENT ABRUPTLY TO THE LOWEST HEIGHT. AFTER THAT HAPPENED, THEY WERE ABLE TO CONTINUE TO USE THE TABLE, ALL FUNCTIONS APPEARED TO WORK AND THEY WERE ABLE TO COMPLETE THE PROCEDURE. IT WAS LATER DETERMINED THAT THEY HAD A MISTRAL-AIR MACHINE UNDER THE LEG SECTION OF THE TABLE AND LOWERED THE TABLE AND ONLY THE HEAD SIDE LOWERED UNTIL THEY KICKED OUT THE MACHINE AND THE TABLE DROPPED. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25612520 ↗MalfunctionJul 3, 2024Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE D860 TABLE RAISED AND LEVELED WITHOUT ANYONE TOUCHING THE TABLE WHILE THE PATIENT WAS ON THE TABLE AND WITH TROCARS INSIDE THE BODY. THE FOOT END OF THE TABLE (THE TABLE WAS IN REVERSE TRENDELENBURG) STARTED TO LEVEL AND AFTER IT LEVELED OUT THE ENTIRE TABLE ROSE SLIGHTLY. THE ANESTHESIOLOGIST DID NOT HAVE THE CONTROL PENDANT IN THEIR HAND AND AS THE FOOT END STARTED TO RISE, THE ANESTHESIOLOGIST TOOK THE CONTROL PIECE AND TRIED TO STOP IT BUT IT DID NOT WORK. THE PHYSICIAN TOOK ABOUT 5 TO 8 MINUTES TO VERIFY THAT NO DAMAGE HAD BEEN DONE AND THAT THE CASE COULD BE COMPLETED SAFELY. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19667553 ↗InjuryJan 3, 2024Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE TABLE MOVED UNEXPECTEDLY. THEY WERE PUTTING THE TABLE IN REVERSE MODE AND WHEN THEY MOVED IT, THE TABLE JERKED. THE DOCTOR HAD BEEN HOLDING A CAMERA UNDER THE KIDNEY WHEN IT HAPPENED AND IT PIERCED THE KIDNEY WHICH REQUIRED MEDICAL INTERVENTION.
Patient problems: Laceration(s)
MAUDE report 18440129 ↗MalfunctionDec 21, 2023Unintended Movement; Unintended Movement
IT WAS REPORTED THAT THE D860 TABLE (S/N: (B)(6) IS MOVING INTO TREND POSITION AT RANDOM ON IT'S OWN. THERE WAS CURRENTLY A PATIENT ON THE TABLE, BUT THE SURGEON DECIDED TO CONTINUE THE CASE WITH THIS TABLE. PER ADDITIONAL INFORMATION RECEIVED, THE PATIENT HAD AN OPEN INCISION. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18374804 ↗MalfunctionFeb 23, 2023Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A PROCEDURE THERE WAS UNINTENDED MOVEMENT OF THE TABLE. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16433394 ↗MalfunctionApr 15, 2022Component or Accessory Incompatibility; Component or Accessory Incompatibility
IT WAS REPORTED THE BACK SECTION OF THE D860 TABLE CAME OFF WHEN PUTTING IT INTO THE TREND POSITION. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14117602 ↗MalfunctionDec 1, 2021Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING A PROCEDURE THERE WAS UNINTENDED MOVEMENT OF THE TABLE AFTER BEING LOWERED ON AN EXTERNAL PIECE OF EQUIPMENT. THERE WERE NO REPORTED INJURIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12909411 ↗InjuryApr 23, 2021Use of Device Problem; Use of Device Problem
IT WAS REPORTED WHEN THE TABLE WAS MOVED INTO POSITION THE TABLE EXPERIENCED UNINTENDED MOVEMENT AFTER BEING CAUGHT ON AN EXTERNAL PIECE OF EQUIPMENT DURING A PROCEDURE. THE UNINTENDED MOVEMENT RESULTED IN PATIENT INJURY.
Patient problems: Unspecified Tissue Injury
MAUDE report 11713583 ↗
Other TABLE, OPERATING-ROOM, ELECTRICAL devices
- NA 72203299Smith & Nephew, Inc. · TABLE, OPERATING-ROOM, ELECTRICAL298 reports
- TS 7000 1723633Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall22 reports
- ALLEN 107003Hill-Rom, Inc. · Table, operating-room, electrical1 recall9 reports
- Head-Positioning accessories 1739994Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall1 report
- TS 7500 2067886Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 1717021Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 4091000Baxter Medical Systems GmbH + Co. KG · TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- ALLEN 107003Hill-Rom, Inc. · Table, operating-room, electrical1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page