Skip to content

INSTRUMENT 6264-8-236RRecalls & reported problems

Howmedica Osteonics Corp. · FEMORAL HEAD TRIAL · Class 2

All models of INSTRUMENT (131)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327250398), received between Jul 18, 2018 and Jul 18, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • Foreign Body In Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 18, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    PRIMARY PROCEDURE, RIGHT HIP. IT WAS REPORTED THAT WHEN TRIALLING THE FEMORAL HEAD, THE HEAD TRIAL GOT PINCHED BY THE FRONT EDGE OF THE SHELL, FORCING THE HEAD TRIAL OFF AND INTO THE ANTERIOR AREA OF PATIENT'S PELVIS. SURGEON WAS ABLE TO LOCATE THE TRIAL USING A C-ARM AND ATTEMPTED TO RETRIEVE THE TRIAL, CAUSING A 10 MINUTE SURGICAL DELAY. HEAD WAS NOT RETRIEVED AND SURGEON REPORTEDLY DECIDED TO LEAVE THE DEVICE IN THE PATIENT PENDING DISCUSSION OF OPTIONS WITH THE PATIENT. THE PROCEDURE WAS OTHERWISE COMPLETED SUCCESSFULLY. UPDATE (B)(6) 2018: MED REVIEW CONFIRMED THAT THE SHELL WAS INITIALLY MALPOSITIONED. THIS WAS CORRECTED AFTER THE TRIAL HEAD HAD DISASSOCIATED FROM THE STEM.

    Patient problems: Foreign Body In Patient

    MAUDE report 7699472 ↗

Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous devices

All Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page