Prime TC 1460-000-000Recalls & reported problems
Stryker Corporation · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Stryker Prime TC adjustable chair has 107 reports, no deaths and 1 injury. Many reports describe brake problems, with brakes that will not engage or engage falsely, and some describe exposed sharp metal edges or parts that do not support weight, with no reported delay in treatment. There is 1 recall, still open, started in May 2021 because chairs meant for healthcare facilities reached home users, who may not follow the use instructions.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 107
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 107 adverse event reports that list this exact device (DI 07613327282665), received between Feb 23, 2018 and Apr 23, 2026 — classified as 1 injury, 106 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (27 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- May 12, 2021Class IIOpen, Classified
Medical devices intended for use in a healthcare facility were distributed to residential customers and may be in use in a home environment. The product s labeling accurately identifies the correct use environment, but home users may not be aware of or fully understand these instructions intended for healthcare professionals
Z-2209-2021 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Mechanical Problem27
- Mechanical Jam26
- Device Fell20
- Sharp Edges14
- Device Operates Differently Than Expected5
- Unstable4
- Device Slipped2
- Device Tipped Over2
Patient problems
- No Clinical Signs, Symptoms or Conditions83
- No Patient Involvement14
- No Known Impact Or Consequence To Patient6
- No Consequences Or Impact To Patient3
- Pain2
- Abrasion1
- Bone Fracture(s)1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 107 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2026Mechanical Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 7 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966489 ↗MalfunctionApr 23, 2026Mechanical Jam
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JANUARY 1- MARCH 31, 2026. THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES CANNOT BE ENGAGED; FALSE ENGAGEMENT OF BRAKES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966321 ↗MalfunctionApr 23, 2026Sharp Edges
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) - (B)(6) 2026. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP EDGES EXPOSED (METAL). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966295 ↗MalfunctionApr 23, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) 2026. THIS REPORT SUMMARIZES 1 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOOTREST DOES NOT SUPPORT WEIGHT; INABILITY TO SUPPORT TRANSPORTING LOAD. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24966112 ↗MalfunctionJan 26, 2026Mechanical Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 4 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24184046 ↗MalfunctionJan 26, 2026Sharp Edges
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN (B)(6) - (B)(6) 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24184041 ↗MalfunctionJan 26, 2026Device Fell
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED SEAT SECTION DOES NOT SUPPORT WEIGHT; INABILITY TO SUPPORT TRANSPORTING LOAD OR FRAME DOES NOT SUPPORT WEIGHT; INABILITY TO SUPPORT TRANSPORTING LOAD. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24183890 ↗MalfunctionJan 26, 2026Mechanical Jam
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN OCTOBER 1-DECEMBER 31, 2025. THIS REPORT SUMMARIZES 17 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED BRAKES CANNOT BE ENGAGED; FALSE ENGAGEMENT OF BRAKES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24183876 ↗MalfunctionOct 27, 2025Unstable; Unstable
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 5 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED FOOTREST - DOES NOT SUPPORT WEIGHT; INABILITY TO SUPPORT TRANSPORTING LOAD. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393829 ↗MalfunctionOct 27, 2025Mechanical Problem; Mechanical Problem
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 12 MALFUNCTION EVENTS(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393822 ↗MalfunctionOct 27, 2025Mechanical Jam; Mechanical Jam
THIS REPORT NOW SUMMARIZES 15 MALFUNCTION EVENTS(S), INSTEAD OF 16.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393820 ↗MalfunctionOct 27, 2025Sharp Edges; Sharp Edges
THIS RECORD IS A CONSOLIDATION OF RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. EVENTS OCCURRED BETWEEN JULY 1-SEPTEMBER 30, 2025. THIS REPORT SUMMARIZES 2 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP EDGES EXPOSED (METAL). NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23393425 ↗MalfunctionJul 25, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22608797 ↗MalfunctionApr 28, 2025Mechanical Problem; Mechanical Problem
THIS REPORT SUMMARIZES 9 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919837 ↗MalfunctionApr 28, 2025Unstable; Unstable
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED DOES NOT SUPPORT WEIGHT; INABILITY TO SUPPORT TRANSPORTING LOAD. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919782 ↗MalfunctionApr 28, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919781 ↗MalfunctionApr 28, 2025Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 18 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED BRAKES CANNOT BE ENGAGED. NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21919404 ↗MalfunctionJan 30, 2025Unstable; Unstable
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICES EXPERIENCED A FAILURE OF THE FOOTREST TO SUPPORT WEIGHT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21277564 ↗MalfunctionJan 27, 2025Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21241328 ↗MalfunctionJan 27, 2025Mechanical Problem; Device Tipped Over; Mechanical Problem; Device Tipped Over
THIS REPORT SUMMARIZES 1 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED REDUCED BRAKE FORCE AND A TIP-OVER. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21241315 ↗MalfunctionJan 27, 2025Device Fell; Device Fell
THIS REPORT SUMMARIZES 5 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED AN INABILITY TO SUPPORT TRANSPORT WEIGHT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21241314 ↗MalfunctionJan 27, 2025Mechanical Jam; Mechanical Jam
THIS REPORT SUMMARIZES 12 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED BRAKES WHICH COULD NOT BE ENGAGED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21241313 ↗MalfunctionJan 27, 2025Mechanical Problem; Mechanical Problem
THIS REPORT SUMMARIZES 6 MALFUNCTION EVENT(S), WHERE IT WAS REPORTED THE DEVICE EXPERIENCED REDUCED BRAKE FORCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21241312 ↗MalfunctionOct 29, 2024Mechanical Problem; Mechanical Problem
THIS REPORT SUMMARIZES 13 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED REDUCED/INADEQUATE BRAKE FORCE. THERE WAS 1 EVENT WITH PATIENT INVOLVEMENT; NO ADVERSE CONSEQUENCES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20559020 ↗MalfunctionOct 25, 2024Sharp Edges; Sharp Edges
THIS REPORT SUMMARIZES 3 MALFUNCTION EVENTS, WHERE IT WAS REPORTED THE DEVICES EXPERIENCED ACCESSIBLE SHARP METAL EDGES. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20539246 ↗
Other CHAIR, ADJUSTABLE, MECHANICAL devices
- Raz Raz-APRaz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
- Raz Raz-ATRaz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
- Raz Raz-ARTRaz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
- Raz Raz-AT600Raz Design Inc · Chair, Adjustable, Mechanical1 recall0 reports
- Aquatec Ocean -VIP Kit Series 1470000Invacare Corporation · CHAIR, ADJUSTABLE, MECHANICAL1 death3 reports
- Raz Raz-CATRaz Design Inc · Chair, Adjustable, Mechanical2 reports
- DRIVEMEDICAL ATC17-BKDrive Devilbiss Healthcare · CHAIR, ADJUSTABLE, MECHANICAL1 report
- Staxi ST010Three Bears(Jiangsu)Healthcare Technology CO.,Ltd · Chair, Adjustable, Mechanical1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page