Skip to content

Tritanium PL 48951102Recalls & reported problems

Stryker Corporation · Posterior Lumbar Cage · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613327295108), received between Nov 4, 2022 and May 21, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Difficult to Insert” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Insert1
  • Improper or Incorrect Procedure or Method1
  • Mechanical Jam1
  • Migration1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 21, 2025Difficult to Insert; Mechanical Jam; Difficult to Insert; Mechanical Jam

    IT WAS REPORTED THAT ONE TRITANIUM POSTERIOR LUMBAR CAGE BECAME JAMMED AND WAS SUBSEQUENTLY REMOVED FROM THE PATIENT INTRA-OPERATIVELY, AND THAT A SECOND TRITANIUM POSTERIOR LUMBAR CAGE MIGRATED INTRA-OPERATIVELY, "PARTIALLY ENTERED THE VERTEBRAL BODY", AND WAS ALSO REMOVED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO REPORTED SURGICAL DELAY BY INSERTING A THIRD CAGE LATERAL FROM THE INITIAL APPROACH. THIS REPORT IS FOR THE CAGE THAT BECAME JAMMED.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 22061045 ↗
  • MalfunctionNov 4, 2022Improper or Incorrect Procedure or Method; Migration; Improper or Incorrect Procedure or Method; Migration

    IT WAS REPORTED THAT A TRITANIUM PL POSTERIOR LUMBAR CAGE MIGRATED ANTERIORLY DURING FINAL IMPACTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15728057 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page