Skip to content

Chromophare CH00000001Recalls & reported problems

Berchtold Holding GmbH · Surgical Light System ? Ceiling Mounted · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Chromophare ceiling-mounted surgical light illuminates the operating area during surgery. The reports mostly describe parts coming loose, falling, cracking, or breaking, with covers shifting downward; 50 reports mention it, with no deaths and 6 injuries. Three recalls are open, including one that started in February 2026 saying the light assembly may not support the weight of the ceiling cover, one about stress lines and chipping in the powder coating, and one about mounting that cannot guarantee stability. Reports went from 0 to 10 in the last 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
50
Deaths reported
0
Injuries reported
6
Recalls naming it
3

Summary

The FDA's MAUDE database contains 50 adverse event reports that list this exact device (DI 07613327296167), received between Jun 19, 2018 and Aug 28, 2026 — classified as 6 injuries, 44 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (12 reports).

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 8 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 23, 2026Open, Classified

    Surgical light assembly may not adequate support the weight of the ceiling cover.

    Z-1688-2026 · Root cause: Unknown/Undetermined by firm · Recall details · FDA record ↗

  2. Jul 15, 2025Open, Classified

    Surgical lights have stress lines forming on the powder coating, which indicates that powder coating chipping can occur.

    Z-2620-2025 · Root cause: Component change control · Recall details · FDA record ↗

  3. Feb 8, 2024Open, Classified

    Due to insufficient mounting force the stability of surgical light system cannot be guaranteed.

    Z-1506-2024 · Root cause: Employee error · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component12
  • Device Fell11
  • Break8
  • Crack8
  • Mechanical Problem8
  • Material Separation7
  • Component Missing3
  • Patient Device Interaction Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions25
  • No Patient Involvement16
  • No Known Impact Or Consequence To Patient3
  • Burn(s)2
  • Bone Fracture(s)1
  • Bruise/Contusion1
  • Head Injury1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 50 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2026Detachment of Device or Device Component

    It was reported that the cosmetic cover of the SLX surgical light system shifted downward.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50488632 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OR WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OR#5 AND INOR#8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.HTM WENT TO INSPECT THE COVERS AND IN O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557886 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OPERATING ROOM WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OPERATING ROOM #5 AND IN OPERATING ROOM #8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557870 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OR WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OR#5 AND INOR#8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.HTM WENT TO INSPECT THE COVERS AND IN O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557860 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OR WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OR#5 AND INOR#8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.HTM WENT TO INSPECT THE COVERS AND IN O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557835 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OR WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OR#5 AND INOR#8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.HTM WENT TO INSPECT THE COVERS AND IN O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557805 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OR WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OR#5 AND INOR#8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.HTM WENT TO INSPECT THE COVERS AND IN O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557777 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OR WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OR#5 AND INOR#8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.HTM WENT TO INSPECT THE COVERS AND IN O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557759 ↗
  • MalfunctionMar 10, 2026Break; Crack

    (EVENT 1) AGGREGATED EVENTS: (B)(6) 2026: DEEP DIVE RESULT: OR (OPERATING ROOM) #1 BEING CLOSED FOR A WEEK FROM (B)(6) 2026 THROUGH (B)(6) 2026. DURING THE WEEK HTM (HEALTHCARE TECHNOLOGY MANAGEMENT) HAD ONLY OBSERVED TWO SMALL CRACKS ON THE SIDE BY THE HANDLE OF STRYKER SURGICAL BOOM LIGHT COVERS, HTM WAS OBSERVING THEM TWICE A DAY FOR THE WEEK WITH NO CHANGE WHEN I LAST CHECKED THEM ON FRIDAY (B)(6) 2026. WHEN HTM CHECKED THE LIGHT COVER ON TUESDAY MORNING AT 8AM THERE WERE ADDITIONAL CRACKS, PLS. SEE THE ATTACHED PICTURES THAT HTM TOOK ON FRIDAY AND THEN ONES HTM TOOK ON TUESDAY. OR WAS CLOSED X 1 WEEK FOR THIS EXPERIMENT WITH THE APPROVAL OF CHIEF OF SURGERY. HTM WAS TOLD NOBODY USED THE ROOM, EMS DID NOT CLEAN THE ROOM, SO THERE WAS NO OTHER HUMAN INTERACTION WITH THE LIGHTS OR EQUIPMENT IN THE ROOM. BASED ON OUR EXPERIMENT IT LEADS ME TO BELIEVE THAT THE PLASTIC IS EITHER NOT GOOD QUALITY, OR THE COVERS DO NOT FIT CORRECTLY CAUSING TOO MUCH STRESS AND THEN THE COVERS TO CRACK. (EVENT 2) (B)(6) 2026: NURSING OBSERVED IN OR#5 AND INOR#8 THERE ARE BROKEN LIGHT COVERS. THIS IS AFTER EMS HAS CHANGED TO THE STRYKER APPROVED CLEANING SOLUTIONS.HTM WENT TO INSPECT THE COVERS AND IN O

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24557727 ↗
  • MalfunctionFeb 24, 2026Detachment of Device or Device Component

    IT WAS REPOTTED THAT THE CEILING COVER FOR THE LIGHT/FP SUSPENSION ON PATIENT LEFT IN OR 7 HAS SLID DOWN AND THERE IS A GAP. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24436840 ↗
  • MalfunctionMay 10, 2024Material Separation; Material Separation

    IT WAS REPORTED THAT THE FLAT PANEL MONITOR IS ONLY HANGING ONTO THE ARM BY A FEW CABLES IN OR 4. IT IS BELIEVED THAT SOMETHING INTERNAL IN THE ARM IS DAMAGED. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19291756 ↗
  • MalfunctionMar 13, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT LIGHT 1 IN OR 7 HAS SEPARATED FROM THE SUSPENSION AND FALLEN TO THE GROUND. THE CONNECTION TO THE SPRING ARM STILL PRESENT ON THE SUSPENSION AND THE LC BEARING HAS BROKEN AND SLID OUT OF THE PROXIMAL END OF THE CARDANIC. THERE WAS ONE EMPLOYEE AND NO PATIENTS IN THE ROOM WHEN THIS HAPPENED. NO ONE WAS STRUCK BY THE FALLING LIGHT. PRIOR TO THIS EVENT THE LIGHT DID NOT HAVE CONTROL FROM THE CFWP, BUT WOULD TURN ON FROM THE LIGHT HEAD. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18899400 ↗
  • InjuryFeb 1, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT THE LIGHT DRIFTED DOWN AND THE SURGEON WAS HIT IN HEAD WALKING INTO IT. THE SURGEON REQUIRED STITCHES.

    Patient problems: Head Injury

    MAUDE report 18625082 ↗
  • MalfunctionNov 22, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE D660 LIGHT HEAD DISCONNECTED FROM ARM. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES AND NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18196517 ↗
  • MalfunctionSep 18, 2023Material Separation; Material Separation

    IT WAS REPORTED THAT REPORT THE BLUE COVER AT THE TOP OF THE LIGHT CAME OFF. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17769930 ↗
  • MalfunctionSep 1, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT ONE OF THE OR LIGHTS CAME COMPLETELY OFF THE ARM. THE LIGHT FELL AFTER A CASE WHILE THE LIGHT WAS BEING CLEANED. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17672983 ↗
  • MalfunctionMay 11, 2023Material Separation; Material Separation

    IT WAS REPORTED BY A NURSE THAT A BLUE DOME COVER HAS FULLY SEPARATED. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16911687 ↗
  • MalfunctionSep 15, 2022Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE LIGHT HEAD FELL OUT OF SPRING ARM AND CRACKED IN OR 3. IT WAS FURTHER REPORTED THE LIGHT HEAD HIT A USER RESULTING IN A SMALL CUT BUT DID NOT REQUIRE MEDICAL INTERVENTION.

    Patient problems: Pain

    MAUDE report 15432655 ↗
  • MalfunctionJan 11, 2022Material Separation; Material Separation

    IT WAS REPORTED AN OR 9 BLUE DOME HAS COME OFF OF THE LIGHT HEAD. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13233178 ↗
  • MalfunctionDec 21, 2021Mechanical Problem; Detachment of Device or Device Component; Mechanical Problem; Detachment of Device or Device Component

    IT WAS REPORTED A D660 LIGHT ON AN NC SUSPENSION FELL IN OPERATING ROOM 17. THE CARDANIC SEPARATED FROM THE SPRING ARM AND STRUCK A NURSE IN THE SHOULDER AS IT FELL. THEY TURNED OFF THE POWER TO THE SUSPENSION AND CUT THE WIRES TO REMOVE IT FROM THE AREA. THE NURSE REPORTED ONLY BRUISING AT THE TIME AND NO MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: Bruise/Contusion

    MAUDE report 13055979 ↗
  • MalfunctionAug 31, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE OR 1 BLUE DOME COVER COMING OFF OF LIGHT HEAD. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12397980 ↗
  • MalfunctionAug 31, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE BLUE DOME FELL OFF OF THE F-GEN LIGHT HEAD IN OR 1. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12397748 ↗
  • MalfunctionAug 3, 2021Material Separation; Material Separation

    IT WAS REPORTED THE BLUE DOME FELL OFF OF THE F-GEN LIGHT HEAD IN OR 1. IT WAS FURTHER REPORTED THIS OCCURRED OUTSIDE OF A PROCEDURE AND NO PATIENT INVOLVEMENT. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12268431 ↗
  • MalfunctionJun 30, 2021Material Separation; Material Separation

    IT WAS REPORTED THE LIGHT HEAD DOME COVERS IN CVOR W HAS POPPED OFF AND NEEDS TO BE REPLACED. IT WAS FURTHER REPORTED THE CORRECT LOCATION WAS CVOR X. THERE WERE NO REPORTED INJURIES OR ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12092156 ↗
  • MalfunctionJun 22, 2021Material Separation; Material Separation

    IT WAS REPORTED THE DOME COVER FELL OFF IN R&D 2. THERE WERE NO REPORTED INJURIES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12040484 ↗

Other DEVICE, MEDICAL EXAMINATION, AC POWERED devices

All DEVICE, MEDICAL EXAMINATION, AC POWERED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page