HOFFMANN 983031Recalls & reported problems
Stryker GmbH · LRF Web Application · Class 2
All models of HOFFMANN (45)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327297751), received between Jul 26, 2018 and Jul 26, 2018 — classified as 1 malfunction.
The most frequently reported specific device problem is “Application Program Problem” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K163656); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Application Program Problem1
- Improper or Incorrect Procedure or Method1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 26, 2018Improper or Incorrect Procedure or Method; Application Program Problem
WHEN COMPUTING A CORRECTION PLAN WITH MORE THAN ONE CORRECTION PER DAY AND "CORRECT AXIAL FIRST" SELECTED, THE HOFFMANN LRF SOFTWARE ERRONEOUSLY COMPUTES THE AXIAL PORTION OF THE CORRECTION, RESULTING IN A CORRECTION PLAN THAT WOULD BE FASTER THAN INTENDED. THE SURGEON OVERRODE THE (ERRONEOUS) CORRECTION TIME WITH HIS OWN CALCULATIONS. WE PROVIDE A SURGEON OVERRIDE FEATURE IN CASE THEY WISH TO DO THAT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7723510 ↗
FDA clearance and classification
Cleared through 510(k) K163656 on Apr 14, 2017 (decision: Substantially Equivalent). Submitted by Stryker GmbH as “Hoffmann LRF System”. The FDA review took 112 days. FDA record ↗
The FDA regulates product code KTT (appliance, fixation, nail/blade/plate combination, multiple component) under 21 CFR 888.3030 as a Class II device. 21 CFR 888.3030 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K163656. One submission often covers many models, so check the product and lot details in each record.
- Z-0288-2019 · Aug 28, 2018 · TerminatedHoffman LRF Hexapod Strut, Extra Short 89mm-109mm, Model Number 4935-0-010. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment
- Z-0289-2019 · Aug 28, 2018 · TerminatedHoffman LRF Hexapod Strut, Extra Short 105mm-139mm, Model Number 4935-0-020. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment
- Z-0291-2019 · Aug 28, 2018 · TerminatedHoffman LRF Hexapod Strut, Extra Short 183mm-295mm, Model Number 4935-0-040. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment
- Z-0290-2019 · Aug 28, 2018 · TerminatedHoffman LRF Hexapod Strut, Extra Short 131mm-191mm, Model Number 4935-0-030. Product Usage - The Hoffmann LRF System is indicated in pediatric patients and adults for the treatment
- Z-3240-2018 · Jul 26, 2018 · TerminatedHoffmann LRF Hexapod Software Catalog #: 983031 Software Version #: HLRF 1.0-2/002, HLRF 1.1-2/002
Common questions about the HOFFMANN 983031
Has the HOFFMANN 983031 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613327297751), based on the FDA recall file dated Nov 27, 2026. However, 5 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the HOFFMANN 983031?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jul 26, 2018. That report lists device problems “Application Program Problem” (1) and “Improper or Incorrect Procedure or Method” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the HOFFMANN 983031?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the HOFFMANN 983031 FDA approved?
It was cleared through 510(k) K163656 on Apr 14, 2017. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the HOFFMANN 983031, and is it still sold?
The FDA's device database lists Stryker GmbH as the company responsible for it (brand HOFFMANN, model 983031). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-11030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15020Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15060Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page