Skip to content

Trident X3 Insert 723-10-32ERecalls & reported problems

Howmedica Osteonics Corp. · Trident X3 10 Degree Polyethylene Insert · Class 2 · Implant

All models of Trident X3 Insert (22)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327339109), received between Mar 22, 2023 and Sep 21, 2026 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Dislodged or Dislocated1
  • Device-Device Incompatibility1
  • Use of Device Problem1

Patient problems

  • Unspecified Infection2
  • Joint Dislocation1
  • Wound Dehiscence1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 21, 2026Use of Device Problem

    Patient problems: Wound Dehiscence

    MAUDE report 50845448 ↗
  • InjuryJul 22, 2024Device-Device Incompatibility; Device Dislodged or Dislocated; Device-Device Incompatibility; Device Dislodged or Dislocated

    PATIENT DIAGNOSED WITH OSTEONECROSIS OF THE FEMORAL HEAD OF THE LEFT HIP, WHO UNDERWENT UNCEMENTED TOTAL ARTHROPLASTY OF THE LEFT HIP WITH A CERAMIC-POLYETHYLENE TRIBOLOGICAL PAIR ON (B)(6) 2024. DURING THE UNEVENTFUL PROCEDURE, SATISFACTORY STABILITY TESTS WERE CARRIED OUT. A 50MM ACETABULAR COMPONENT AND 36MM FEMORAL HEAD WERE USED. AFTER A CONTROL RADIOGRAPHY, THE IMAGE DREW ATTENTION TO THE SUBLUXATION OF THE FEMORAL HEAD IN RELATION TO THE ACETABULAR COMPONENT. DUE TO SUSPICION OF LOOSENING OF THE POLYETHYLENE LINER OR INTERPOSITION OF SOFT TISSUES, A SURGICAL REVIEW WAS INDICATED THE FOLLOWING DAY (B)(6) 2024 AND THE PATIENT'S HOSPITALIZATION WAS EXTENDED. DURING THE NEW SURGERY, IT WAS FOUND THAT THE POLYETHYLENE LINER WAS PERFECTLY LOCKED, BUT THE 36 MM HEAD WAS NOT FULLY FITTED. THE POLYETHYLENE LINER WAS REMOVED AND THE ENGRAVING ON THE PIECE WAS CHECKED, WHICH CONFIRMED THAT THE INTERNAL DIAMETER WAS ACTUALLY 32 MM, CONTRARY TO WHAT WAS FOUND IN THE PREVIOUS DAY'S SURGERY. THE LINER WAS REPLACED WITH A 36MM INTERNAL DIAMETER COMPONENT AND THERE WAS A GOOD REDUCTION. THE PATIENT WAS DISCHARGED FROM HOSPITAL THE FOLLOWING DAY, 06/22/2024. THE PRODUCT BOX WAS NOT STORED BEC

    Patient problems: Joint Dislocation

    MAUDE report 19800388 ↗
  • InjuryFeb 29, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT RETURNED TO THEATRE FOR A HEAD LINER EXCHANGE DUE TO INFECTION STRYKER TRITANIUM 2 SHELL AND LINE DEPUY CORAIL STEM AND HEAD.

    Patient problems: Unspecified Infection

    MAUDE report 18806108 ↗
  • InjuryMar 22, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT REQUIRED A HIP REVISION FOR INFECTION, 1ST STAGE HIP REVISION. REMOVING AN EXETER V40 STEM, TRIDENT HEMISPHERICAL ACETABULAR SHELL SIZE 52 ,SMALL ARTISAN CEMENT PLUG. UNSURE IS STILL HAD A TRIDENT X3 10 DEGREE LINER OR CERAMIC V40 32MM HEAD INSITU AS IMPLANT STICKERS SENT FROM SURGEON FOR PATIENT SAID THE PATIENT HAD A CERAMIC V40 HEAD BUT THEN THE REMOVED HEAD APPEARED TO BE METAL AND THE SURGEON SPECIFIED THE PATIENT ALREADY HAD THIS REPLACED BUT UNABLE TO GIVE FURTHER DETAILS. REPLACED WITH A COMPETITOR CUP AND LINER. STRYKER REPLACED WITH LFIT V40 28MM HEAD +4MM AND EXETER LONG STEM 44 NO2 205MM.

    Patient problems: Unspecified Infection

    MAUDE report 16591373 ↗

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates