Trident II 702-04-46CRecalls & reported problems
Howmedica Osteonics Corp. · TRIDENT II TRITANIUM CLUSTER 46C · Class 2 · Implant
All models of TRIDENT II (42)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 07613327343809), received between Mar 7, 2022 and Jul 11, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Unstable” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Unstable2
- Failure to Advance1
- Insufficient Device Problem Information1
Patient problems
- Joint Laxity2
- Ambulation or Postural Difficulties1
- Bone Fracture(s)1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 11, 2025Failure to Advance; Insufficient Device Problem Information; Failure to Advance; Insufficient Device Problem Information
DURING MAKO THA CASE, SURGEON COMMENTED THAT THERE HAD BEEN A SMALL FRACTURE TO THE POSTERIOR ACETABULUM. MAKO PLAN WAS ORIGINALLY DISCUSSED AND SHOWED SURGEON PLAN WITH 46 CUP, SURGEON ASKED, IF POSSIBLE, TO UPSIZE CUP, SHOWED BIGGER SIZE AND SAID NOT AN APPROPRIATE OPTION DUE TO BEING OVERSIZED, SURGEON HAPPY TO CONTINUE WITH 46 CUP PLAN, WELL SIZED. REAMING HAD BEEN COMPLETED WITH MAKO AS PER USUAL, WHEN TRYING TO IMPACT WITH MAKO, SURGEON WAS HAVING TROUBLE GETTING CUP INTO POSITION, WAS REPORTING THAT THE PATIENT HAS EXTENSIVE SOFT TISSUE AND WAS HAVING TROUBLE GETTING MAKO REAMING HANDLE INTO POSITION. AT THIS POINT HAD RECHECKED PELVIC CHECKPOINT AND TOOL CHECKPOINT AND BOTH HAD PASSED. SURGEONS STATED THAT HE DID NOT WANT TO EXTEND INCISION TO HELP ACCESS, SURGEON HAPPY TO TRY OFFSET IMPACTOR MAKO HANDLE. CHANGED TO OFFSET IMPACTION HANDLE AND CHECKED TOOL CHECKPOINT WHICH PASSED, BUT SURGEON STILL HAVING ISSUES WITH ACCESS AND GETTING MAKO IMPACTION HANDLE INTO GOOD POSITION, STATING THAT PATIENT SOFT TISSUE WAS PUSHING OFF PLANE. SURGEON DECIDED TO MANUALLY IMPACT WITHOUT MAKO IMPACTION HANDLE. SURGEON MANUALLY IMPACTED AND THEN CHECKED SURGICAL RESULTS WITH BLUE PROBE, S
Patient problems: Bone Fracture(s)
MAUDE report 22478214 ↗InjuryAug 17, 2023Unstable; Unstable
THIS PI IS FOR THE REVISION ON (B)(6) 2022. PATIENT HAD AN INDEX LEFT THAT ON (B)(6) 2022. PATIENT HAD INSTABILITY AND DISLOCATED, REQUIRING CLOSED REDUCTION UNDER ANESTHESIA ON (B)(6) 2022. PATIENT CONTINUED WITH INSTABILITY AND UNDERWENT REVISION ON (B)(6) 2022. ALL COMPONENTS WERE EXCHANGED EXCEPT THE STEM.
Patient problems: Ambulation or Postural Difficulties; Joint Laxity
MAUDE report 17562836 ↗InjuryMar 7, 2022Unstable; Unstable
PATIENT HAD AN INDEX LEFT THA ON (B)(6) 2021. PATIENT PRESENTED WITH INSTABILITY AND DISLOCATION OF LEFT HIP. UNDERWENT REVISION FOR INSTABILITY ON (B)(6) 2021. CUP, LINER, AND HEAD WERE EXCHANGED.
Patient problems: Joint Laxity
MAUDE report 13677953 ↗
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page