Howmedica Osteonics Corp. · Trident II Tritanium Multihole Acetabular Shell 50D · Class 2 · Implant

All models of TRIDENT II (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327343953), received between Feb 26, 2020 and Apr 3, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Malposition of Device” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Malposition of Device3
  • Adverse Event Without Identified Device or Use Problem1
  • Corroded1
  • Device Dislodged or Dislocated1
  • Fracture1
  • Loss of Osseointegration1
  • Noise, Audible1
  • Osseointegration Problem1

Patient problems

  • Insufficient Information2
  • Joint Dislocation2
  • Ambulation or Postural Difficulties1
  • Bone Fracture(s)1
  • Inadequate Osseointegration1
  • Injury1
  • Loss of Range of Motion1
  • Pain1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 3, 2026Fracture

    THE FOLLOWING INFORMATION WAS PROVIDED: THE PRIMARY SURGERY WAS CONDUCTED ON (B)(6) 2023. THE REVISION SURGERY FOR A CUP LOOSENING REPLACEMENT WAS CONDUCTED ON (B)(6) 2026, AND DURING WHICH IT IS FOUNDED THAT THE CUP BROKEN AND THE SCREW LOOSENING. THE SCREW HAD COME LOOSE AND WAS CUTTING INTO THE LINER SLIDING SURFACE. THE SURGERY WAS COMPLETED WITH IMPLANTED NON-STRYKER PRODUCTS.

    Patient problems: Insufficient Information

    MAUDE report 24776141 ↗
  • InjuryNov 25, 2025Malposition of Device; Device Dislodged or Dislocated; Malposition of Device; Device Dislodged or Dislocated

    REVISION SURGERY DUE TO CUP DISLOCATION WITH SCREW FRACTURE.

    Patient problems: Insufficient Information; Joint Dislocation

    MAUDE report 23642900 ↗
  • InjuryAug 12, 2025Loss of Osseointegration; Osseointegration Problem; Loss of Osseointegration; Osseointegration Problem

    NO NEW INFORMATION.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22780526 ↗
  • InjuryAug 15, 2023Corroded; Positioning Problem; Noise, Audible; Corroded; Positioning Problem; Noise, Audible

    PATIENT HAD AN INDEX RIGHT THA ON (B)(6) 2021. PATIENT PRESENTED WITH RIGHT HIP PAIN AND SQUEAKING. PATIENT UNDERWENT REVISION ON (B)(6) 2022. SOME CORROSION WAS FOUND ON THE TAPER-HEAD UNION. ALL COMPONENTS WERE EXCHANGED EXCEPT THE FEMORAL STEM.

    Patient problems: Loss of Range of Motion; Pain

    MAUDE report 17553613 ↗
  • InjuryJul 3, 2023Malposition of Device; Adverse Event Without Identified Device or Use Problem; Malposition of Device; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "DR. REVISED A LEFT HIP HE ORIGINALLY DID VIA MAKO TO A COMPETITOR REVISION CUP/CAGE CONSTRUCT AND CHANGED THE HEAD. THIS WAS DUE TO PELVIC FRACTURE DURING THE ORIGINAL SURGERY AND SUBSEQUENT FAILURE POST-OP. DURING THE ORIGINAL SURGERY, DR. NOTICED DURING CUP IMPACTION THE ANTERIOR WALL OF THE ACETABULUM HAD A FRACTURE. THIS IS WHEN HE CHANGED FROM A 48 PRIMARY CUP TO A SZ 50 MULTI-HOLE INTRA-OP. POST-OP HE SAID THE PATIENT COULDN'T WALK AND THEREFORE DECIDED TO REVISE THE CUP SIDE. HE NOTED THE BONE QUALITY WAS VERY POOR. THE REVISION SURGERY HE KEPT THE ORIGINAL EXETER STEM IN PLACE." SPOKE TO REP. REP CONFIRMED THAT ON (B)(6) /2023 THE SIZE 48 SHELL WAS IMPACTED, ACETABULUM FRACTURE WAS NOTED, 48 WAS REMOVED AND A 50 IMPLANTED. 5 DAYS LATER ON (B)(6) 2023, THE 50 SHELL WAS REMOVED AND COMPETITOR ACETABULAR COMPONENTS IMPLANTED. THA 4.1

    Patient problems: Ambulation or Postural Difficulties; Bone Fracture(s)

    MAUDE report 17245872 ↗
  • InjuryFeb 26, 2020Malposition of Device; Malposition of Device

    IT WAS REPORTED THE PATIENT'S RIGHT HIP WAS REVISED. AS REPORTED BY REP: "PATIENT REVISED DUE TO DISLOCATION." SPOKE TO REP. THE FEMORAL HEAD CAME OUT OF THE POLY LINER. THE HEAD AND LINER WERE REVISED. THE SHELL AND 2 SCREWS WERE ALSO REVISED TO CHANGE THE SHELL'S POSITIONING.

    Patient problems: Injury; Joint Dislocation

    MAUDE report 9753573 ↗

Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices

All Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices →