Skip to content

Trident II 7030-6535Recalls & reported problems

Howmedica Osteonics Corp. · 6.5MM LOW PROFILE HEX SCREW 35MM · Class 2 · Implant

All models of TRIDENT II (42)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 07613327345100), received between Oct 25, 2018 and Nov 25, 2025 — classified as 4 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Break1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Loss of Osseointegration1
  • Material Deformation1
  • Material Twisted/Bent1
  • Migration1
  • Patient Device Interaction Problem1

Patient problems

  • Insufficient Information2
  • No Consequences Or Impact To Patient2
  • Inadequate Osseointegration1
  • Joint Laxity1
  • Neuralgia1
  • No Clinical Signs, Symptoms or Conditions1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 25, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    RIGHT HIP REVISION. PT. PRESENTED WITH SIGNS OF SCIATIC NERVE IRRITATION FROM AN ACETABULAR SCREW FOLLOWING A THA. DR. OPTED TO REMOVE THE OFFENDING SCREW(S) AND REPLACE THE HEAD AND LINER ALONG WITH A THOUGH IRRIGATION AND DEBRIDEMENT OF SCAR TISSUE. NO COMPLAINTS FILED AGAINST THE COMPONENTS THEMSELVES, NO FURTHER INFO AVAILABLE.

    Patient problems: Neuralgia

    MAUDE report 23644441 ↗
  • InjuryAug 28, 2023Migration; Migration

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED. AS REPORTED: "ACETABULAR SHELL SHIFTED FROM IMPLANTED POSITION. EXPLANTS ARE BEING RETAINED BY THE FACILITY."

    Patient problems: Insufficient Information

    MAUDE report 17629973 ↗
  • InjuryAug 25, 2022Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED DUE TO INSTABILITY. A SHELL, 6 SCREWS, HEAD AND LINER WERE REVISED TO ANOTHER SHELL, LINER, SCREWS, AND HEAD. REP CONFIRMED THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Joint Laxity

    MAUDE report 15294902 ↗
  • MalfunctionJul 8, 2022Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    PRIMARY PROCEDURE, LEFT HIP. IT WAS REPORTED THAT POST-OP X-RAY SHOWED THAT THE SCREW WAS BENT/ CURVED. SCREW WAS NOT BENT OR CURVED WHEN TAKEN OUT OF THE PACKAGING. SURGEON USED 25MM DRILL WITH GUIDE TO DRILL THE PILOT HOLE. PATIENT'S BONE QUALITY UNKNOWN BUT WAS REPORTED AS 'NOT YOUNG, STRONG BONE'. SURGEON DOES NOT HAVE ANY PLAN TO REVISE THE PATIENT. UPDATE PER SALES REP: 35MM SCREW IS BENT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14961293 ↗
  • InjuryAug 12, 2021Fracture; Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Loss of Osseointegration; Loosening of Implant Not Related to Bone-Ingrowth

    AS REPORTED: "REVISED HIP FOR LOOSE CUP. PATIENT CAME WITH LOOSE CUP AND PAIN." PATIENT'S LEFT HIP WAS REVISED. A SHELL, MDM METAL LINER, ADM/ MDM POLY INSERT, AND 2 SCREWS WERE REVISED. REP PROVIDED AN OPERATIVE REPORT FROM THE PRIOR IMPLANTATION AND AN X-RAY. UPDATE 20/JULY/2021 WG: SPOKE TO REP. THERE ARE NO ALLEGATIONS AGAINST THE REVISED MDM METAL LINER OR ADM/ MDM POLY INSERT. THE SHELL HAD SHIFTED THROUGH THE PATIENT'S PELVIS. NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 12312017 ↗
  • MalfunctionFeb 11, 2019Break; Break

    REVISION OF COMPETITOR DEVICES, LEFT HIP. WHILE IMPLANTING A SCREW THROUGH A TRIDENT II SHELL, THE SCREW BROKE. THE SURGEON HAD USED A 3.3 X 25 DRILL BIT TO PREPARE THE ACETABULUM. 6.5 X 35MM SCREW WAS LESS THAN HALFWAY INTO THE PATIENT. PATIENT REPORTEDLY HAS SCLEROTIC BONE (HARD/ DENSE BONE QUALITY). THE PORTION OF THE SCREW WAS NOT SITTING PROUD AND THE SURGEON DECIDED TO LEAVE THAT PORTION IN THE PATIENT. SURGEON PROCEEDED TO USE A 4.0 X 25 DRILL BIT TO PREPARE THE ACETABULUM, AND WAS ABLE TO IMPLANT 3 SCREWS OF SMALLER LENGTH (6.5 X 20 AND 6.5 X 30MM). REP CONFIRMED THERE WAS NO SURGICAL DELAY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8325060 ↗
  • MalfunctionOct 25, 2018Fracture; Fracture

    IT WAS REPORTED THAT THE SCREW BROKE IN HALF DURING INSERTION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8006232 ↗

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates