N/A (REF D087031)Recalls & reported problems
Stryker Sustainability Solutions, Inc. · Size: 9F (3.2mm) · Class 2
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613327376067), received between May 1, 2024 and Jan 8, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem2
- Insufficient Device Problem Information1
Patient problems
- Cardiac Perforation1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- Pericardial Effusion1
- Retroperitoneal Hemorrhage1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJan 8, 2026Use of Device Problem
IT WAS REPORTED THAT AFTER A PFA ATRIAL FIBRILLATION ABLATION A PATIENT HAD A RETROPERITONEAL "BLEED". DURING THE PROCEDURE THE PHYSICIAN FELT THE ICE CATHETER WAS ADVANCING INCORRECTLY. THERE WAS TORTUOUS VASCULATURE. THE PHYSICIAN PERFORMED A CONTRAST IMAGE OF THE IVC PRIOR TO TRANSEPTAL AND IT REVEALED NO ISSUES. THE PATIENT PRESSURES WERE STABLE THROUGHOUT THE PROCEDURE. THE PHYSICIAN PERFORMED A REPEAT CONTRAST IMAGE OF THE IVC POST PROCEDURE, AND IT REVEALED NO BLEED. AFTER THE PROCEDURE THE PATIENTS¿ BLOOD PRESSURE DROPPED AND A CT REVEALED AN RP BLEED. THE PATIENT RECOVERED. THE SUSPECTED CATHETER WAS THE VIEW FLEX ICE CATHETER. MULTIPLE ATTEMPTS WERE MADE TO OBTAIN ADDITIONAL INFORMATION. NO INFORMATION WAS PROVIDED REGARDING PATIENT INJURY, MEDICAL INTERVENTION, OR EXTENDED PROCEDURE TIME REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: Retroperitoneal Hemorrhage
MAUDE report 24004590 ↗InjuryJul 2, 2025Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT FOLLOWING A PFA PROCEDURE, A PERICARDIAL EFFUSION OCCURRED WHICH REQUIRED SURGERY TO REPAIR THE PERFORATION. THE PROCEDURE HAD GONE WELL, PVI TO FIND A HIS ELECTROGRAM AND PACE IN THE RIGHT VENTRICLE (RV) WITH A GOOD OUTCOME FROM THE PVI POINT. A FEW HOURS POST-PROCEDURE WHILE IN RECOVERY, A PERICARDIAL EFFUSION WAS DIAGNOSED. SURGERY WAS PERFORMED TO REPAIR THE PERFORATION IN THE RIGHT ATRIUM. AS REPORTED, THE PATIENT WAS STABLE AND RECOVERING IN THE ICU. THE PHYSICIAN THINKS THE PERFORATION MAY HAVE BEEN CAUSED BY THE VIEWFLEX ICE CATHETER WHICH WAS REPROCESSED BY STRYKER. THERE WAS NO OTHER PATIENT INJURY OR MEDICAL INTERVENTION REPORTED AND NO EXTENDED PROCEDURE TIME WAS REPORTED. THESE ARE COMMONLY USED DEVICES THAT ARE READILY AVAILABLE.
Patient problems: Cardiac Perforation; Pericardial Effusion
MAUDE report 22383928 ↗MalfunctionMay 22, 2024
THE PATIENT WAS TORTUROUS, SO CATH TEAM UPSIZED TO 12FR X 30CM FROM 10FR X 23CM. PROCEDURALIST PUT THE CATHETER UP THE IVC (INFERIOR VENA CAVA), REMARKED ON THE IMAGE, AND THEN WHEN HE TORQUED THE CATHETER TO ENTER THE RIGHT ATRIUM, ULTRASOUND IMAGE WAS LOST. CATH TEAM PULLED THE CATHETER OUT AND UPON INSPECTION THE TIP WAS BROKEN AT A 45-DEGREE ANGLE TO THE CATHETER SHAFT WITH A VISIBLE SPLIT. MANUFACTURER RESPONSE FOR REPROCESSED INTRAVASCULAR ULTRASOUND CATHETER, N/A (PER SITE REPORTER). STRYKER DID PHONE CONFERENCE WITH HOSPITAL TO ADVISE OF STEPS THEY ARE TAKING TO INVESTIGATE CONCERN.
Patient problems: Insufficient Information
MAUDE report 19369665 ↗MalfunctionMay 1, 2024
REPROCESSED ICE (INTRACARDIAC ECHOCARDIOGRAPHY) CATHETER INSERTED INTO THE PATIENT VIA 10 FRENCH SHEATH IN THE LEFT FEMORAL VEIN. DEVICE HAD IMAGERY ON THE MONITOR, AND PROCEDURALIST NOTED THAT THE IMAGE WAS POOR. PROCEDURALIST DID FLUORO OF THE DEVICE AND NOTED THAT THE TIP OF THE ICE CATHETER WAS AT AN ACUTE NON-NORMAL ANGLE AND WAS MOVING UNDER FLUORO. DEVICE WAS RETRIEVED WITH NO PATIENT HARM AND PROCEDURE CONTINUED WITH A NEW DEVICE. MANUFACTURER RESPONSE FOR REPROCESSED INTRAVASCULAR ULTRASOUND CATHETER (PER SITE REPORTER). VENDOR REP WOULD LIKE TO PICK ITEM UP FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19219978 ↗
Other Reprocessed intravascular ultrasound catheter devices
- MEDLINE D087031RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter2 recalls5 reports
- MEDLINE 10135910RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter2 recalls0 reports
- MEDLINE 10135936RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter2 recalls0 reports
- NA ACU10135936Sterilmed, Inc. · Reprocessed intravascular ultrasound catheter1 recall11 reports
- NA ACU10135910Sterilmed, Inc. · Reprocessed intravascular ultrasound catheter1 recall6 reports
- MEDLINE 10438577RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter1 recall0 reports
- MEDLINE 10439236RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter1 recall0 reports
- MEDLINE 10439011RHMedline Industries, Inc. · Reprocessed intravascular ultrasound catheter1 recall0 reports