Skip to content

Surpass Evolve FDS50015Recalls & reported problems

Stryker Corporation · Flow Diverter System · Class 3 · Implant

All models of Surpass Evolve (22)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07613327386691), received between Jun 22, 2021 and Feb 9, 2024 — classified as 4 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Deformation Due to Compressive Stress1
  • Difficult or Delayed Activation1
  • Insufficient Device Problem Information1

Patient problems

  • Headache3
  • Insufficient Information1
  • Ischemia Stroke1
  • Memory Loss/Impairment1
  • Nausea1
  • No Clinical Signs, Symptoms or Conditions1
  • Paresis1
  • Stenosis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 9, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE SUBJECT FLOW DIVERTER WAS IMPLANTED IN A PATIENT SUCCESSFULLY ON (B)(6) 2022. POST PROCEDURE ON (B)(6) 2023, PATIENT STATS THAT SHE LAST REMEMBERS SPEAKING WITH HER FRIEND THAT AFTERNOON AND HAS NO RECOLLECTION OF EVENTS THEREAFTER. PATIENT HERSELF COULD NOT PROVIDE MUCH HISTORY. AS PER FAMILY, PATIENT WAS LAST WELL KNOWN AT 5:20 PM THAT DAY. PATIENT TOOK A NAP AND WHEN WOKE UP PATIENT HAD LEFT-SIDED FACIAL DROOP AND INABILITY TO MOVE LEFT SIDE OF THE BODY. A CT (COMPUTERIZED TOMOGRAPHY) WAS PERFORMED ON (B)(6) 2023. RESULTS INDICATE 'NO NEW ANEURYSM OR ARTERIOVENOUS MALFORMATION IS IDENTIFIED. THE DURAL VENOUS SINUSES AND DEEP VENOUS SYSTEM ARE PATENT', NO STENOSIS WAS OBSERVED, THE SUBJECT STENT IMPLANTED WAS PATENT. PATIENT WAS PROVIDED WITH DIPHENHYDRAMINE, METOCLOPRAMIDE, ACETAMINOPHEN, RIBOFLAVIN MAGNESIUM FOR FOCAL IPSILATERAL HEADACHE. THE AE (ADVERSE EVENT) OUTCOME IS INDICATED RECOVERED/RESOLVED. THE AE WAS REPORTED TO BE RESOLVED ON (B)(6) 2023. THE AE WAS INDICATED AS POSSIBLY RELATED TO STUDY PROCEDURE, STUDY DEVICE, DAPT (DUAL ANTIPLATELET THERAPY) AND UNDERLYING CONDITION OR DISEASE. PATIENT HAS MEDICAL HISTORY OF HYPERTENSION, DYSLIPIDEMIA,

    Patient problems: Headache; Memory Loss/Impairment; Paresis

    MAUDE report 18677793 ↗
  • MalfunctionJan 29, 2024Deformation Due to Compressive Stress; Insufficient Device Problem Information; Deformation Due to Compressive Stress; Insufficient Device Problem Information

    IT WAS REPORTED THAT THREE DAYS POST PROCEDURE WITH THE SUBJECT FLOW DIVERTER THE PATIENT WAS RE-ADMITTED TO HOSPITAL WITH COMPLICATIONS. NO FURTHER INFORMATION IS AVAILABLE. ADDITIONAL INFORMATION RECEIVED CLARIFIED THAT THREE DAYS POST PROCEDURE WITH THE SUBJECT FLOW DIVERTER, THE PATIENT WAS READMITTED TO THE HOSPITAL WITH HEADACHE AND NAUSEA COMPLICATIONS. PROXIMAL END OF THE SUBJECT STENT WAS FOUND TO BE CHANGED ON COMPUTED TOMOGRAPHY (CT) SCAN AND ANGIOGRAPHY WHICH WAS PERFORMED ON RE-ADMISSION. THE PATIENT CONDITION IS DETAILED AS FINE.

    Patient problems: Headache; Insufficient Information; Nausea

    MAUDE report 18595680 ↗
  • MalfunctionOct 20, 2022Difficult or Delayed Activation; Difficult or Delayed Activation

    IT WAS REPORTED THAT THE PROXIMAL END OF THE STENT (SUBJECT DEVICE) FAILED TO FULLY EXPAND AND PARTIALLY DEPLOYED INSIDE THE PATIENT¿S ANATOMY, IT WAS RECAPTURED/REMOVED FROM THE PATIENT¿S ANATOMY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO CLINICAL CONSEQUENCES WERE REPORTED TO THE PATIENT DUE TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15636982 ↗
  • InjuryOct 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT ON (B)(6) 2022, THE SUBJECT FLOW DIVERTER STENT WAS SUCCESSFULLY IMPLANTED FOR A INTERNAL CAROTID ARTERY ANEURYSM IN A PATIENT. DURING 6 MONTH FOLLOW UP, 25-50% VESSEL NARROWING WITH DISTAL STENT FISH MOUTHING WAS OBSERVED. THE PHYSICIAN WAS CONCERNED WITH THE ANGLED POSITION OF THE SUBJECT STENT WITHIN THE VESSEL. THE PATIENT WAS ASYMPTOMATIC THOUGH WAS HAVING NEW HEADACHES THAT STILL ARE LIKELY NOT RELATED. THE PATIENT WAS CONTINUED ON PLAVIX AND ASPIRIN FOR 6 MORE MONTHS. ON (B)(6) 2022 BALLOON ANGIOPLASTY WITH ANOTHER STENT PLACEMENT WAS PERFORMED TO FURTHER ASSIST WITH THE SUBJECT STENT REMAINING OPEN . THE DEVICE WAS MUCH MORE OPEN POST PLASTY, THE STENOSIS WAS MUCH LESS. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Headache; Stenosis

    MAUDE report 15636443 ↗
  • InjuryAug 25, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED DURING THE PROCEDURE THE PATIENT SUFFERED A STROKE WITH UNKNOWN RELATION TO THE STENT (SUBJECT DEVICE). THE PATIENT WAS PLACED ON A VENTILATOR AND HAD A NIH SCORE OF 18 AFTER THE EVENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Stroke/CVA

    MAUDE report 12366355 ↗
  • InjuryJun 22, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PROCEDURE OF SUBJECT FLOW DIVERTING STENT DEPLOYMENT IN INTERNAL CAROTID ARTERY WAS COMPLETED SUCCESSFULLY ON (B)(6) 2021. IT WAS REPORTED THAT 12- 24 HOURS POST PROCEDURE, PATIENT HAD STROKE DUE TO CLOTTING AROUND THE SUBJECT FLOW DIVERTING STENT. ADDITIONAL SURGERY WAS PERFORMED TO TREAT THE STROKE. THE SUBJECT STENT REMAINS INSIDE THE PATIENT¿S INTERNAL CAROTID ARTERY. NO ADDITIONAL INFORMATION IS AVAILABLE

    Patient problems: Ischemia Stroke; Thrombosis/Thrombus

    MAUDE report 12043155 ↗

Other Intracranial aneurysm flow diverter devices

All Intracranial aneurysm flow diverter devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page