Skip to content

COBRA 3910900093Recalls & reported problems

Stryker Corporation · COBRA SUTURE PASSER, CAPTURE · Class 1

All models of COBRA (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Cobra suture passer has 138 adverse event reports, with no deaths and 2 injuries. Many describe the needle bending, breaking or detaching during shoulder procedures, and the broken pieces were often retrieved. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
138
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 138 adverse event reports that list this exact device (DI 07613327413595), received between Jul 22, 2019 and Feb 24, 2026 — classified as 2 injuries, 136 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (68 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 20 reportsDec 2023: 12 reportsJan 2024: 4 reportsFeb 2024: 1 reportsMar 2024: 5 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent68
  • Material Deformation41
  • Break17
  • Fracture17
  • Material Integrity Problem4
  • Mechanical Jam4
  • Material Fragmentation3
  • Failure to Advance2

Patient problems

  • No Clinical Signs, Symptoms or Conditions136
  • Insufficient Information4
  • Foreign Body In Patient2
  • Device Embedded In Tissue or Plaque1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 138 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 24, 2026Fracture

    IT WAS REPORTED THAT WHILE TRYING TO PASS THE SUTURE THE NEEDLE BROKE WHERE IT CONNECTS TO THE PUSHING SHAFT. THE NEEDLE FELL ON THE FLOOR WHEN THE COBRA WAS PULLED OUT OF THE SHOULDER. THE CASE WAS COMPLETED BY USING A DIFFERENT PASSER. THERE WAS NO DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24435098 ↗
  • MalfunctionFeb 6, 2026Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE. ALL PIECES WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24283397 ↗
  • MalfunctionJan 9, 2026Fracture

    IT WAS REPORTED THAT THERE WAS BENT NEEDLE DEPLOYMENT/RETRACTION. BASED ON INVESTIGATION FINDINGS THERE WAS A BROKEN NEEDLE. PER ADDITIONAL INFORMATION RECEIVED, ALL PIECES WERE REMOVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24020960 ↗
  • MalfunctionJan 7, 2026Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE DURING THE CASE. ALL PIECES WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23991084 ↗
  • MalfunctionDec 17, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE HAD COMPLETELY DETACHED. THE DETACHED NEEDLE WAS IDENTIFIED AND FULLY REMOVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23832428 ↗
  • MalfunctionDec 16, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE WHILE PASSING THE SUTURE. THE END OF NEEDLE WAS RECOVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23822539 ↗
  • MalfunctionDec 11, 2025Fracture; Fracture

    IT WAS REPORTED THAT THERE WAS A BROKEN NEEDLE DURING PROCEDURE. ALL PIECES WERE REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23780964 ↗
  • MalfunctionNov 26, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE COBRA NEEDLE BROKE AND THAT ALL THE PIECES WERE RETRIEVED FROM THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23659839 ↗
  • MalfunctionOct 23, 2025Fracture; Fracture

    IT WAS REPORTED THAT A BROKEN NEEDLE OCCURRED DURING A CASE. ALL BROKEN PIECES WERE REMOVED, AND THE CASE WAS COMPLETED SUCCESSFULLY. NO ADVERSE EFFECTS TO PATIENT. THERE WAS A 10 MINUTE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23373956 ↗
  • MalfunctionJul 15, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT THE NEEDLE TIP BROKE WHICH DELAYED THE CASE 20-30 MINUTES. X-RAY WAS BROUGHT IN TO FIND THE BROKEN TIP. THERE WEAS NO MEDICAL INTERVENTION OR ADVERSE EFFECTS TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22515954 ↗
  • MalfunctionApr 7, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE OF THE PRODUCT BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21786818 ↗
  • MalfunctionJul 23, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19812578 ↗
  • MalfunctionJul 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19731843 ↗
  • MalfunctionMay 24, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE BREAKING DURING THE PROCEDURE. THE BROKEN NEEDLE PIECE WAS NOT LEFT IN PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19397811 ↗
  • MalfunctionMay 23, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE NEEDLE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19388400 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18820163 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18820146 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18816921 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18816916 ↗
  • MalfunctionMar 1, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18816910 ↗
  • MalfunctionFeb 5, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18646554 ↗
  • MalfunctionJan 31, 2024Material Fragmentation; Material Twisted/Bent; Material Fragmentation; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18616525 ↗
  • MalfunctionJan 25, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18582054 ↗
  • MalfunctionJan 25, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18581880 ↗
  • MalfunctionJan 19, 2024Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NEEDLE BECAME BENT WHILE THE DEVICE WAS IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18547270 ↗

Other Accessories, arthroscopic devices

All Accessories, arthroscopic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page