STRYKER 7400-037-000Recalls & reported problems
Stryker Corporation · TPX RECIPROCATING SAW · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Stryker TPX reciprocating saw is a powered surgical handpiece used to cut bone. Most of its 66 reports describe the tool overheating, often noticed during testing before a procedure, plus fluid or oil leaking, parts coming loose, and the blade ejecting or the saw switching on by itself. There are no deaths and no injuries, and most reports say the patient was not affected. No recalls name this device. Reports have slowed, with 5 in the past 12 months compared with 15 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 66
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 66 adverse event reports that list this exact device (DI 07613327483000), received between Jun 19, 2020 and Mar 10, 2026 — classified as 66 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (43 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Overheating of Device43
- Fluid/Blood Leak12
- Detachment of Device or Device Component5
- Device Remains Activated2
- Unintended Ejection2
- Unintended Power Up2
- No Apparent Adverse Event1
Patient problems
- No Clinical Signs, Symptoms or Conditions63
- No Consequences Or Impact To Patient3
- Burn(s)1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 66 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 10, 2026Overheating of Device
THE MANUFACTURER SALES REP FROM THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24562277 ↗MalfunctionMar 10, 2026Overheating of Device
THE MANUFACTURER SALES REP FROM THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24561344 ↗MalfunctionJan 8, 2026Overheating of Device
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 24002497 ↗MalfunctionNov 11, 2025Overheating of Device; Overheating of Device
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523448 ↗MalfunctionNov 3, 2025Overheating of Device; Overheating of Device
NO NEW INFORMATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23452268 ↗MalfunctionAug 27, 2025Overheating of Device; Overheating of Device
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22904020 ↗MalfunctionJun 17, 2025Overheating of Device; Overheating of Device
THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22225327 ↗MalfunctionJun 4, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE WAS LEAKING BLACK OIL DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22142061 ↗MalfunctionJun 3, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE WAS LEAKING OIL DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22132936 ↗MalfunctionJun 2, 2025Fluid/Blood Leak; Fluid/Blood Leak
THE USER FACILITY REPORTED THAT THE DEVICE WAS FOUND TO BE LEAKING DURING TESTING AFTER THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22122377 ↗MalfunctionMay 13, 2025Overheating of Device; Overheating of Device
THE USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22012315 ↗MalfunctionApr 2, 2025Overheating of Device; Overheating of Device
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21754009 ↗MalfunctionMar 6, 2025Overheating of Device; Overheating of Device
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING A PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21535821 ↗MalfunctionMar 6, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21535815 ↗MalfunctionFeb 13, 2025Overheating of Device; Overheating of Device
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21375790 ↗MalfunctionFeb 13, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21375785 ↗MalfunctionJan 16, 2025Overheating of Device; Overheating of Device
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21168940 ↗MalfunctionDec 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD SPRING AND WASHER MISSING WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21019773 ↗MalfunctionDec 9, 2024Fluid/Blood Leak; Fluid/Blood Leak
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE IS LEAKING OIL DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20884323 ↗MalfunctionNov 19, 2024Overheating of Device; Overheating of Device
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE OVERHEATED WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20725269 ↗MalfunctionOct 30, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20566894 ↗MalfunctionOct 22, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE FRONT END OF THE DEVICE WAS DISASSEMBLED WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20510694 ↗MalfunctionJul 29, 2024Overheating of Device; Overheating of Device
IT WAS REPORTED THAT DURING TESTING BY THE MANUFACTURER SALES REPRESENTATIVE AT THE USER FACILITY, THE DEVICE OVERHEATED. AS THE EVENT OCCURRED DURING TESTING, THERE WAS NO PATIENT INVOLVEMENT, NO DELAY TO A SURGICAL PROCEDURE, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED RELATED TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19858150 ↗MalfunctionJul 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE DISASSEMBLED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19774346 ↗MalfunctionJul 17, 2024Overheating of Device; Overheating of Device
THE USER FROM USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19761721 ↗
Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices
- RemB 6400034000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,618 reports
- NA XMAXAnspach Effort, The · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1,225 reports
- NA EG1ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE859 reports
- NA CRANI-ASynthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE1 recall772 reports
- NA 5450400000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE775 reports
- NA 05.001.010Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE770 reports
- RemB 6400015000Stryker Corporation · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE603 reports
- NA QD11Synthes GmbH · DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE580 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page