STRYKER 7400-009-000Recalls & reported problems

Stryker Corporation · TPX HANDSWITCH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker TPX handswitch is a hand control used with powered surgical drills. The 34 reports mostly describe the switch staying activated or running on its own, coming apart, or overheating during testing or procedures, with no mention of patient harm. There are 34 malfunction reports, no deaths and no injuries, and no recalls name this device. Reports have dropped sharply, from 23 in the earlier 12-month period to 4 in the most recent one.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 07613327483055), received between Aug 30, 2022 and May 21, 2026 — classified as 34 malfunctions.

The most frequently reported specific device problem is “Device Remains Activated” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 16 reportsJul 2025: 2 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Remains Activated18
  • Detachment of Device or Device Component8
  • Overheating of Device4
  • Unintended Power Up3
  • Fail-Safe Did Not Operate1

Patient problems

  • No Clinical Signs, Symptoms or Conditions34

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 21, 2026Overheating of Device

    THE SALES REP FROM USER FACILITY REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25246432 ↗
  • MalfunctionMay 14, 2026Overheating of Device

    No new information.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25173266 ↗
  • MalfunctionFeb 23, 2026Overheating of Device

    THE MANUFACTURER SALES REP REPORTED THAT THE DEVICE OVERHEATED DURING TESTING PRIOR TO A PROCEDURE AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24420526 ↗
  • MalfunctionJan 8, 2026Detachment of Device or Device Component

    THE MANUFACTURER FIELD SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD MISSING COMPONENT WHILE IT WAS BEING SERVICED AT THE USER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24005810 ↗
  • MalfunctionAug 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS MISSING COMPONENT WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22903194 ↗
  • MalfunctionJul 21, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22556708 ↗
  • MalfunctionJul 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE WAS MISSING SPRINGS WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22435838 ↗
  • MalfunctionJun 24, 2025Device Remains Activated; Device Remains Activated

    THE USER FROM USER FACILITY REPORTED THAT THE DEVICE WAS RUNNING CONTINUOUSLY DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22318124 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292595 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292590 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292474 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292472 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292418 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292415 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292328 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22292324 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22289392 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22289391 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22289390 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22289389 ↗
  • MalfunctionJun 19, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, THE DEVICE WAS CONTINUOUSLY RUNNING (RUN-ON). THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY. THERE WAS NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22289388 ↗
  • MalfunctionJun 17, 2025Device Remains Activated; Device Remains Activated

    THE USER FROM USER FACILITY REPORTED THAT THE DEVICE CONTINUED TO RUN (RUN-ON) DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22235435 ↗
  • MalfunctionJun 17, 2025Device Remains Activated; Device Remains Activated

    THE USER FROM USER FACILITY REPORTED THAT THE DEVICE CONTINUED TO RUN (RUN-ON) DURING TESTING PRIOR TO A PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22235179 ↗
  • MalfunctionMar 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD MISSING BALL PLUNGER WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WERE NO ADVERSE CONSEQUENCES RELATED TO THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21539950 ↗
  • MalfunctionMar 5, 2025Device Remains Activated; Device Remains Activated

    THE MANUFACTURER SERVICE TECHNICIAN REPORTED THAT THE DEVICE HAD RUN ON WHILE IT WAS BEING SERVICED AT THE MANUFACTURER FACILITY. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21524790 ↗

Other DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices

All DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE devices →