Skip to content

Insignia 7000-6606Recalls & reported problems

Howmedica Osteonics Corp. · Hip Stem - High Offset · Class 2 · Implant

All models of Insignia (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327511406), received between Aug 30, 2022 and Sep 25, 2026 — classified as 9 injuries.

The most frequently reported specific device problem is “Loss of Osseointegration” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Loss of Osseointegration2
  • Device Dislodged or Dislocated1
  • Fracture1
  • Material Erosion1
  • Migration1

Patient problems

  • Bone Fracture(s)3
  • Pain3
  • Inadequate Osseointegration2
  • Periprosthetic fracture2
  • Ambulation or Postural Difficulties1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    The following information was provided: Patient revised due to periprosthetic fracture. No other information available due to hospital policy.

    Patient problems: Periprosthetic fracture

    MAUDE report 50896024 ↗
  • InjuryFeb 25, 2026Loss of Osseointegration; Migration

    THE FOLLOWING INFORMATION WAS PROVIDED: REVISED A RIGHT HIP INSIGNIA STEM DUE TO SUBSIDENCE WHICH CAUSED PAIN FOR THE PATIENT. HE REVISED THE LINER FROM A 44 TO A 40 AND REMOVED THE ACETABULAR SCREW. HE REVISED THE STEM TO A COMPETITOR STEM. UPDATE REGARDING REVISED LINER AND SCREW: NO ALLEGATIONS AGAINST THE IMPLANT. HE REMOVED THE SCREW BECAUSE IT WAS NO LONGER NEEDED. HE CHANGED FROM A 44 TO A 40 LINER BECAUSE COMPETITOR DIDN'T HAVE A 44 HEAD OPTION.

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 24443598 ↗
  • InjuryFeb 24, 2026Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING INFORMATION WAS PROVIDED: PATIENT'S HIP REVISED DUE TO PERIPROSTHETIC FEMORAL FRACTURE. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY.

    Patient problems: Periprosthetic fracture

    MAUDE report 24433782 ↗
  • InjuryNov 28, 2025Loss of Osseointegration; Loss of Osseointegration

    SURGEON SUSPECTED A LOOSE FEMORAL COMPONENT, REVISION SURGERY WAS PERFORMED AND THE DIAGNOSIS WAS CONFIRMED. HE PROCEEDED TO REVISE AN INSIGNIA STEM AND PUT IN A RESTORATION MODULAR STEM.

    Patient problems: Inadequate Osseointegration

    MAUDE report 23669978 ↗
  • InjuryJun 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE FOLLOWING WAS REPORTED: "LEFT HIP. PATIENT REVISED DUE TO PERIPROSTHETIC FRACTURE. NO OTHER INFORMATION AVAILABLE DUE TO HOSPITAL POLICY.".

    Patient problems: Bone Fracture(s)

    MAUDE report 22144372 ↗
  • InjuryMay 13, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT'S RIGHT HIP WAS REVISED. DUE TO THE CERAMIC HEAD FRACTURING IN HALF, THE HEAD DISSOCIATED FROM THE STEM. THE HEAD AND LINER WERE REVISED. REP CONFIRMED THAT THERE WERE NO ALLEGATIONS AGAINST THE REVISED LINER, AND THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Ambulation or Postural Difficulties

    MAUDE report 22012649 ↗
  • InjuryJun 12, 2024Material Erosion; Fracture; Material Erosion; Fracture

    IT WAS REPORTED THAT THE PATIENT'S LEFT HIP WAS REVISED AFTER PATIENT COMPLAINT OF PAIN. INTRA-OPERATIVELY, THE FOLLOWING WAS NOTED: THE CERAMIC HEAD WAS BROKEN INTO AT LEAST 2 PIECES. THE INSIDE OF THE HEAD SHOWED CONCENTRIC WEAR, AND THE TRUNNION WAS DAMAGED. REVISING SURGEON REPORTED HIS ASSESSMENT: THE SURGEON FOR THE INDEX PROCEDURE (DIFFERENT SURGEON) WASTED A -5 CERAMIC HEAD, AND IMPLANTED A +7.5 CERAMIC HEAD WITHOUT A SLEEVE. REVISING SURGEON BELIEVES THE HEAD WAS NOT FULLY IMPACTED ON THE STEM, LEADING TO MOTION OF THE HEAD ON THE STEM. THE CERAMIC HEAD AND NEUTRAL POLY LINER WERE REVISED TO ANOTHER CERAMIC HEAD AND NEUTRAL ECENTRIC POLY LINER (REASON FOR THE LINER CHANGE WAS THE SURGEON WANTED TO TIGHTEN UP THE CONSTRUCT AND DID NOT WANT TO USE A +7.5 OR +10 HEAD).

    Patient problems: Pain

    MAUDE report 19517457 ↗
  • InjuryApr 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISED A HEAD/STEM DUE TO PERI-PROSTHETIC FRACTURE.

    Patient problems: Bone Fracture(s)

    MAUDE report 19184812 ↗
  • InjuryAug 30, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    AS REPORTED: "DIAGNOSIS: RIGHT HIP PAIN. RIGHT HIP REVISION ARTHROPLASTY FEMORAL COMPONENT ONLY WITH OPEN REDUCTION INTERNAL FIXATION. NO FURTHER INFORMATION AVAILABLE PER HOSPITAL." A STEM AND HEAD WERE REVISED.

    Patient problems: Bone Fracture(s); Pain

    MAUDE report 15318317 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page