Skip to content

Insignia 7000-6607Recalls & reported problems

Howmedica Osteonics Corp. · Hip Stem - High Offset · Class 2 · Implant

All models of Insignia (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613327511451), received between Feb 23, 2025 and Feb 23, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1

Patient problems

  • Foreign Body Reaction1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 23, 2025Degraded; Degraded

    AS REPORTED: "A PATIENT THAT HAD A LEFT PRIMARY DIRECT ANTERIOR HIP REPLACEMENT ON (B)(6) 2023. OVER 15 MONTHS LATER HE REPORTED PAIN AND LIMITED RANGE OF MOTION AND SUBSEQUENTLY FOLLOWED UP WITH DR., DR. SAW ON X-RAY THAT THERE WERE WHAT LOOKED TO BE FRACTURED PIECES OF THE PATIENTS BIOLOX DELTA V40 FEMORAL HEAD. THE PATIENT WAS THEN BROUGHT TO SURGERY FOR A HEAD AND LINER EXCHANGE REVISION SURGERY. INTRAOPERATIVELY: THE SURGEON NOTICED FRAGMENTED PIECES OF THE BIOLOX DELTA V40 AND THE FEMORAL HEAD WAS SPLIT IN HALF. THERE WERE SOME DARK STAINING OF THE SOFT TISSUES AND A SLIGHT METALLOSIS APPEARANCE. THE SURGEON REMOVED THE FEMORAL HEAD, REMAINING CERAMIC PIECES IN THE JOINT, AS WELL AS THE LINER, THE STEM AND CUP WERE WELL FIXED, NOT DAMAGED, AND THEN RETAINED. THE SURGEON THEN PUT A V40 TO C- TAPER ADAPTOR SLEEVE ON AND A NEW BIOLOX C-TAPER FEMORAL HEAD." REP REPORTED THAT THE (CONSULTANT) SURGEON WOULD LIKE INVESTIGATION RESULTS, AND THAT NO FURTHER INFORMATION WILL BE RELEASED BY THE HOSPITAL OR SURGEON.

    Patient problems: Foreign Body Reaction

    MAUDE report 21442650 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices

All PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page