Skip to content

ONLINE TDM Acetaminophen Gen.2 08056684190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336120934), received between May 16, 2025 and Dec 8, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 8, 2025High Test Results; High Test Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ONLINE TDM ACETAMINOPHEN GEN. 2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 303 ANALYTICAL UNIT. THE INITIAL RESULT WAS 48.5 UG/ML. THE NURSE QUESTIONED THE RESULT, WHICH PROMPTED THE CUSTOMER TO REPEAT THE SAMPLE. THE SAMPLE WAS REPEATED TWICE ON ANOTHER MODULE, AND BOTH REPEAT RESULTS WERE <5 UG/ML WITH A FLAG. THE REPEAT RESULTS WERE DEEMED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23744963 ↗
  • MalfunctionMay 16, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ACETAMINOPHEN RESULTS FOR 11 PATIENT SAMPLES FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULTS RANGED FROM >200 TO 12 UG/ML. THE REPEAT RESULTS WERE ALL <5 UG/ML. A SPECIFIC EXAMPLE WAS PROVIDED ON AN INITIAL RESULT OF 50 UG/ML AND A REPEAT RESULT OF <5 UG/ML. SOME PATIENTS WERE TREATED WITH AN ANTI-ACETAMINOPHEN DRUG, BUT THERE WERE NO ADVERSE EFFECTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22035908 ↗

Other COLORIMETRY, ACETAMINOPHEN devices

All COLORIMETRY, ACETAMINOPHEN devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page