ONLINE TDM Acetaminophen Gen.2 08056684190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336120934), received between May 16, 2025 and Dec 8, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2025High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ONLINE TDM ACETAMINOPHEN GEN. 2 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS C 303 ANALYTICAL UNIT. THE INITIAL RESULT WAS 48.5 UG/ML. THE NURSE QUESTIONED THE RESULT, WHICH PROMPTED THE CUSTOMER TO REPEAT THE SAMPLE. THE SAMPLE WAS REPEATED TWICE ON ANOTHER MODULE, AND BOTH REPEAT RESULTS WERE <5 UG/ML WITH A FLAG. THE REPEAT RESULTS WERE DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23744963 ↗MalfunctionMay 16, 2025High Test Results; Non Reproducible Results; High Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ACETAMINOPHEN RESULTS FOR 11 PATIENT SAMPLES FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULTS RANGED FROM >200 TO 12 UG/ML. THE REPEAT RESULTS WERE ALL <5 UG/ML. A SPECIFIC EXAMPLE WAS PROVIDED ON AN INITIAL RESULT OF 50 UG/ML AND A REPEAT RESULT OF <5 UG/ML. SOME PATIENTS WERE TREATED WITH AN ANTI-ACETAMINOPHEN DRUG, BUT THERE WERE NO ADVERSE EFFECTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22035908 ↗
Other COLORIMETRY, ACETAMINOPHEN devices
- Atellica CH Acet 11097522Siemens Healthcare Diagnostics Inc. · COLORIMETRY, ACETAMINOPHEN3 reports
- ONLINE TDM Acetaminophen Gen2 6769942190Roche Diagnostics GmbH · COLORIMETRY, ACETAMINOPHEN2 reports
- Dimension Vista® Flex® reagent cartridge ACTM 10445110Siemens Healthcare Diagnostics Inc. · COLORIMETRY, ACETAMINOPHEN2 reports
- VITROS 1037274Ortho-Clinical Diagnostics, Inc. · COLORIMETRY, ACETAMINOPHEN1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page