Skip to content

Cardiac C-Reactive Protein (Latex) High Sensitive 08057605190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336121207), received between Jul 31, 2025 and Jul 31, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 31, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THERE WAS AN ALLEGATION OF A QUESTIONABLE CARDIAC C-REACTIVE PROTEIN (LATEX) HIGH SENSITIVE RESULT FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 76 MG/L, ACCOMPANIED BY A DATA FLAG. WHEN RUN ON DECREASED SAMPLE VOLUME MODE, AND THE RESULT WAS 0.000 MG/L, ACCOMPANIED BY A DATA FLAG. MANUAL DILUTIONS OF 1:5, 1:10, AND 1:20 WERE COMPLETED, AND THE RESULTS WERE 0.000 MG/L, ACCOMPANIED BY A DATA FLAG. THEY TRIED TO WARM THE SAMPLE AND STILL RECEIVED A RESULT OF 76 MG/L, ACCOMPANIED BY A DATA FLAG. THEY RAN THE SAMPLE ON CRP4 AND RECEIVED A RESULT OF 8.68 MG/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22674630 ↗

Other CARDIAC C-REACTIVE PROTEIN, ANTIGEN, ANTISERUM, AND CONTROL devices

All CARDIAC C-REACTIVE PROTEIN, ANTIGEN, ANTISERUM, AND CONTROL devices →

Sources and updates