Skip to content

Gamma-Glutamyltransferase ver.2 Standardized against IFCC / Szasz 08057796190Recalls & reported problems

Roche Diagnostics GmbH · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336121252), received between Jul 24, 2024 and Sep 23, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026High Test Results

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50870255 ↗
  • MalfunctionJul 24, 2024Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR GAMMA-GLUTAMYLTRANSFERASE (GGT) ON TWO COBAS C 503 ANALYTICAL UNITS. THE INITIAL RESULT WAS 91.4 UI/L ON C503 ANALYZER A. THE SAMPLE WAS REPEATED ON C503 ANALYZER A WITH RESULTS OF 55.6 UI/L, 57 UI/L, AND 66 UI/L. THE SAMPLE WAS REPEATED ON C503 ANALYZER B WITH A RESULT OF 91.4 UI/L ACCOMPANIED BY AN INVALIDATING DATA FLAG. THE REPEAT RESULTS ON C503 ANALYZER B WERE 91.4 UI/L AND 56 UI/L. THE RESULTS CLOSE TO 56 UI/L WERE BELIEVED TO BE CORRECT. ANOTHER SAMPLE FROM THE PATIENT WAS REPEATED 5 TIMES WITH RESULTS OF 51 UI/L AND 52 UI/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19819604 ↗

Other KINETIC METHOD, GAMMA-GLUTAMYL TRANSPEPTIDASE devices

All KINETIC METHOD, GAMMA-GLUTAMYL TRANSPEPTIDASE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page