Homocysteine Enzymatic Assay 08057826190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
All models of Homocysteine Enzymatic Assay (2)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336121290), received between Aug 7, 2024 and Aug 7, 2024 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Difficult to Setup or Prepare” (1 report).
3 other recalls cover devices cleared under the same FDA submission (K113793); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Device Difficult to Setup or Prepare1
- Low Test Results1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2024Device Difficult to Setup or Prepare; Low Test Results; Non Reproducible Results; Device Difficult to Setup or Prepare; Low Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE LOW RESULTS FOR AN UNSPECIFIED NUMBER OF PATIENT SAMPLES TESTED FOR HOMOCYSTEINE ENZYMATIC ASSAY (HCYS) ON A COBAS PURE ANALYZER COMPARED TO A HIGH-PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) METHOD. OF THE DATA PROVIDED FOR 11 PATIENT SAMPLES, THE RESULTS FOR 10 PATIENT SAMPLES WERE DISCREPANT. REFER TO THE ATTACHED DATA FOR THE PATIENT RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19930753 ↗
FDA clearance and classification
Cleared through 510(k) K113793 on Jun 5, 2012 (decision: Substantially Equivalent). Submitted by Roche as “Homocysteine Test”. The FDA review took 165 days. FDA record ↗
The FDA regulates product code LPS (urinary homocystine (nonquantitative) test system) under 21 CFR 862.1377 as a Class II device. 21 CFR 862.1377 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K113793. One submission often covers many models, so check the product and lot details in each record.
- Z-0489-2025 · Oct 17, 2024 · Open, Classifiedcobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine
- Z-0679-2020 · Oct 4, 2019 · TerminatedRoche Homocysteine-In vitro test for the quantitative determination of total L-homocysteine in human serum and plasma on Roche/Hitachi cobas c systems Catalog Number: 06542921190
- Z-1622-2015 · Mar 23, 2015 · TerminatedHomocysteine test system - Homocysteine Reagent. 05385415190; Homocysteine 100 Tests, cobas c, COBAS INTEGRA Homocysteine Enzymatic Assay is an in vitro test for the quantitative
Common questions about the Homocysteine Enzymatic Assay 08057826190
Has the Homocysteine Enzymatic Assay 08057826190 been recalled?
No FDA recall record in our data names this exact device identifier (DI 07613336121290), based on the FDA recall file dated Nov 27, 2026. However, 3 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the Homocysteine Enzymatic Assay 08057826190?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Aug 7, 2024. That report lists device problems “Device Difficult to Setup or Prepare” (1), “Low Test Results” (1) and “Non Reproducible Results” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Homocysteine Enzymatic Assay 08057826190?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Homocysteine Enzymatic Assay 08057826190 FDA approved?
It was cleared through 510(k) K113793 on Jun 5, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Homocysteine Enzymatic Assay 08057826190, and is it still sold?
The FDA's device database lists Roche Diagnostics GmbH as the company responsible for it (brand Homocysteine Enzymatic Assay, model 08057826190). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other URINARY HOMOCYSTINE (NONQUANTITATIVE) TEST SYSTEM devices
- Homocysteine Enzymatic Assay 6542921190Roche Diagnostics GmbH · URINARY HOMOCYSTINE (NONQUANTITATIVE) TEST SYSTEM1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page