Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 07613336121382), received between Dec 15, 2022 and Oct 21, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Incorrect Measurement1
  • Low Test Results1
  • Non Reproducible Results1
  • Unable to Obtain Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 21, 2025Incorrect Measurement; Incorrect Measurement

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH LITHIUM ON A COBAS C 703 ANALYTICAL UNIT. THE SAMPLE INITIALLY RESULTED IN A LITHIUM VALUE OF 2.43 NMOL/L AND IT REPEATED AS 0.94 NMOL/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23350214 ↗
  • MalfunctionMar 19, 2024Unable to Obtain Readings; Low Test Results; Unable to Obtain Readings; Low Test Results

    THERE WAS AN ALLEGATION OF A QUESTIONABLE LITHIUM RESULT FROM THE COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.1 MMOL/L WITH A DATA FLAG AND WAS REPORTED OUTSIDE OF THE LABORATORY. THE RESULT DID NOT MATCH THE PATIENT'S HISTORY AND THE SAMPLE WAS REPEATED. THE REPEAT RESULT WAS 2.3 MMOL/L AND WAS BELIEVED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18933199 ↗
  • MalfunctionFeb 28, 2023High Test Results; High Test Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR TWO PATIENT SAMPLES TESTED WITH LI LITHIUM ON A COBAS PRO C 503 ANALYTICAL UNIT SERIAL NUMBER (B)(4). THE INITIAL VALUES WERE REPORTED OUTSIDE OF THE LABORATORY. DUE TO PREVIOUS ISSUES, THE SAMPLES WERE REPEATED ON A SECOND ANALYZER ON 03-FEB-2023 AND THESE VALUES WERE DEEMED CORRECT. THE FIRST SAMPLE INITIALLY RESULTED IN A LITHIUM VALUE OF 1.2 MMOL/L AND REPEATED AS 0.7 MMOL/L. THE SECOND SAMPLE INITIALLY RESULTED IN A LITHIUM VALUE OF 1.9 MMOL/L AND REPEATED AS 0.6 MMOL/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16456723 ↗
  • MalfunctionDec 15, 2022Non Reproducible Results; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED FOR LITHIUM (LI) ON A COBAS PRO C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 1.28 MMOL/L WITH A DATA FLAG. THE REPEAT RESULT WAS 0.00524 MMOL/L WITH A DATA FLAG. THE CUSTOMER REPEATED THE SAMPLE 5 TIMES WITH RESULTS OF 0.0173 MMOL/L WITH A DATA FLAG, 0.0190 MMOL/L WITH A DATA FLAG, 0.0183 MMOL/L WITH A DATA FLAG, 0.0141 MMOL/L WITH A DATA FLAG, AND 0.0160 MMOL/L WITH A DATA FLAG. THE REPEAT RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. IT IS NOT CLEAR WHICH RESULTS WERE BELIEVED TO BE CORRECT. THE LI REAGENT LOT NUMBER WAS 59094201 WITH AN EXPIRATION DATE OF 30-SEP-2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15991887 ↗

Other Assay, porphyrin, spectrophotometry, lithium devices