Tina-quant Apolipoprotein A-1 ver.2 08105448190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336133521), received between Oct 25, 2023 and Aug 13, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (1 report).
1 other recall cover devices cleared under the same FDA submission (K013249); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results1
- Incorrect Measurement1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Low Test Results1
- No Apparent Adverse Event1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 13, 2024Low Test Results; Low Test Results
THE INITIAL REPORTER COMPLAINED OF A DISCREPANT LOW RESULT FOR 1 PATIENT SAMPLE TESTED FOR TINA-QUANT APOLIPOPROTEIN A-1 VER.2 (APOAT) ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.489 G/L. THE SAMPLE WAS REPEATED BY A COMPETITOR METHOD AND THE RESULT WAS "DISCREPANT." THE SPECIFIC RESULT WAS NOT PROVIDED. THE SAMPLE WAS REPEATED ON THE C503 ANALYZER USING A 1:2 DILUTION AND THE RESULT WAS 0.85 G/L. A NEW SAMPLE WAS OBTAINED AND THE INITIAL RESULT ON THE C503 ANALYZER WAS 0.8 G/L. THE SAMPLE WAS REPEATED USING A 1:2 DILUTION AND THE RESULT WAS 0.924 G/L.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19972879 ↗MalfunctionOct 25, 2023Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; No Apparent Adverse Event; Non Reproducible Results; Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; No Apparent Adverse Event; Non Reproducible Results
THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH APOAT (APOLIPOPROTEIN A) ON A COBAS PRO C 503 ANALYTICAL UNIT. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN AN APOLIPOPROTEIN A VALUE OF 94 MG/DL. THE VALUE WAS CONSIDERED VERY HIGH SO THE SAMPLE WAS REPEATED. THE REPEAT APOLIPOROTEIN A RESULT WAS 45 MG/DL. THE SAMPLE WAS ALSO REPEATED TWICE ON A SECOND ANALYZER ON (B)(6) 2023, RESULTING IN APOLIPOROTEIN A VALUES OF 36 MG/DL AND 36 MG/DL. THE RESULTS FROM THE SECOND ANALYZER WERE DEEMED CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18007436 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K013249. One submission often covers many models, so check the product and lot details in each record.
- Z-1124-2019 · Dec 14, 2018 · TerminatedCOBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin Gen.2 Catalog Number 04
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.