Tina-quant Apolipoprotein A-1 ver.2 08105448190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336133521), received between Oct 25, 2023 and Aug 13, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (1 report).

1 other recall cover devices cleared under the same FDA submission (K013249); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Incorrect Measurement1
  • Incorrect, Inadequate or Imprecise Result or Readings1
  • Low Test Results1
  • No Apparent Adverse Event1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2024Low Test Results; Low Test Results

    THE INITIAL REPORTER COMPLAINED OF A DISCREPANT LOW RESULT FOR 1 PATIENT SAMPLE TESTED FOR TINA-QUANT APOLIPOPROTEIN A-1 VER.2 (APOAT) ON A COBAS C 503 ANALYTICAL UNIT. THE INITIAL RESULT WAS 0.489 G/L. THE SAMPLE WAS REPEATED BY A COMPETITOR METHOD AND THE RESULT WAS "DISCREPANT." THE SPECIFIC RESULT WAS NOT PROVIDED. THE SAMPLE WAS REPEATED ON THE C503 ANALYZER USING A 1:2 DILUTION AND THE RESULT WAS 0.85 G/L. A NEW SAMPLE WAS OBTAINED AND THE INITIAL RESULT ON THE C503 ANALYZER WAS 0.8 G/L. THE SAMPLE WAS REPEATED USING A 1:2 DILUTION AND THE RESULT WAS 0.924 G/L.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19972879 ↗
  • MalfunctionOct 25, 2023Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; No Apparent Adverse Event; Non Reproducible Results; Incorrect Measurement; Incorrect, Inadequate or Imprecise Result or Readings; High Test Results; No Apparent Adverse Event; Non Reproducible Results

    THE INITIAL REPORTER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH APOAT (APOLIPOPROTEIN A) ON A COBAS PRO C 503 ANALYTICAL UNIT. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN AN APOLIPOPROTEIN A VALUE OF 94 MG/DL. THE VALUE WAS CONSIDERED VERY HIGH SO THE SAMPLE WAS REPEATED. THE REPEAT APOLIPOROTEIN A RESULT WAS 45 MG/DL. THE SAMPLE WAS ALSO REPEATED TWICE ON A SECOND ANALYZER ON (B)(6) 2023, RESULTING IN APOLIPOROTEIN A VALUES OF 36 MG/DL AND 36 MG/DL. THE RESULTS FROM THE SECOND ANALYZER WERE DEEMED CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18007436 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K013249. One submission often covers many models, so check the product and lot details in each record.

  • Z-1124-2019 · Dec 14, 2018 · Terminated
    COBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin Gen.2 Catalog Number 04