ONLINE TDM Theophylline 08253153190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336147931), received between Dec 17, 2025 and Dec 17, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Low Test Results” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 17, 2025Low Test Results; Low Test Results
THERE WAS AN ALLEGATION OF A QUESTIONABLE ONLINE TDM THEOPHYLLINE RESULT FROM THE COBAS C 503 ANALYTICAL UNIT FOR ONE PATIENT. THE INITIAL RESULT WAS NO RESULT GENERATED, ACCOMPANIED BY A DATA FLAG. THE FIRST REPEAT RESULT WAS 0.8 G/ML, ACCOMPANIED BY A DATA FLAG. THE SECOND REPEAT RESULT WAS 12.7 G/ML. THE THIRD REPEAT RESULT WAS 12.9 G/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23826772 ↗
Other ENZYME IMMUNOASSAY, THEOPHYLLINE devices
- Syva® Emit® 2000 Theophylline Assay 10445324Siemens Healthcare Diagnostics Inc. · ENZYME IMMUNOASSAY, THEOPHYLLINE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page