Skip to content

Elecsys AMH 08819378190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07613336166291), received between Feb 4, 2025 and Feb 4, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “High Test Results” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results1
  • Material Integrity Problem1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2025High Test Results; Material Integrity Problem; Non Reproducible Results; High Test Results; Material Integrity Problem; Non Reproducible Results

    THE INITIAL REPORTER COMPLAINED OF A DISCREPANT HIGH RESULT FOR 1 PATIENT SAMPLE TESTED FOR ELECSYS AMH (AMH) ON A COBAS E 402 ANALYTICAL UNIT. THE INITIAL RESULT WAS 17.4 NG/ML. THE REPEAT RESULT WAS < 0.0100 NG/ML WITH A DATA FLAG. THE PATIENT WAS TESTED IN A DIFFERENT LABORATORY AND THE RESULT FROM THE MAGLUMI METHOD WAS 0.049 NG/ML. THE ORIGINAL SAMPLE WAS REPEATED AT THE LABORATORY USING THE MAGLUMI METHOD AND THE REPEAT RESULT WAS 0.056 NG/ML. THE REPEAT RESULTS WERE BELIEVED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21305729 ↗

Other Anti-Müllerian hormone test system devices

All Anti-Müllerian hormone test system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page