The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336166635), received between Jan 2, 2026 and Sep 23, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “High Test Results” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results2
- Incorrect Measurement1
- Low Test Results1
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2026High Test Results
The initial reporter received a questionable Elecsys BRAHMS PCT assay result for one patient sample tested on the cobas 6000 e601 module. Initial analysis of the patient sample yielded 2.4 ng/mL. Repeat analysis yielded 0.09 ng/mL.The repeat result was reported, as the patient's previous result was 0.41 ng/mL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50865648 ↗MalfunctionJan 2, 2026Incorrect Measurement; High Test Results; Low Test Results; Non Reproducible Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS BRAHMS PCT RESULTS FROM THE COBAS 6000 E601 MODULE. THE INITIAL RESULT WAS >100 NG/ML. THE PHYSICIAN REQUESTED MANUAL DILUTION, AND THE CUSTOMER PERFORMED A 1:4 MANUAL DILUTION. THE RESULT WAS 15 NG/ML (X4=60 NG/ML). THE CUSTOMER PERFORMED ANOTHER 1:2 MANUAL DILUTION OF THE ORIGINAL SAMPLE, PRODUCING A RESULT OF 52 NG/ML. THE CUSTOMER DID NOT KNOW WHICH RESULT WAS CORRECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23942961 ↗
Other Procalcitonin assay devices
- Elecsys BRAHMS PCT 08828679190Roche Diagnostics GmbH · Procalcitonin assay7 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page