Elecsys BRAHMS PCT 08828644190Recalls & reported problems

Roche Diagnostics GmbH · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336166635), received between Jan 2, 2026 and Sep 23, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “High Test Results” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Test Results2
  • Incorrect Measurement1
  • Low Test Results1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026High Test Results

    The initial reporter received a questionable Elecsys BRAHMS PCT assay result for one patient sample tested on the cobas 6000 e601 module. Initial analysis of the patient sample yielded 2.4 ng/mL. Repeat analysis yielded 0.09 ng/mL.The repeat result was reported, as the patient's previous result was 0.41 ng/mL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50865648 ↗
  • MalfunctionJan 2, 2026Incorrect Measurement; High Test Results; Low Test Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS BRAHMS PCT RESULTS FROM THE COBAS 6000 E601 MODULE. THE INITIAL RESULT WAS >100 NG/ML. THE PHYSICIAN REQUESTED MANUAL DILUTION, AND THE CUSTOMER PERFORMED A 1:4 MANUAL DILUTION. THE RESULT WAS 15 NG/ML (X4=60 NG/ML). THE CUSTOMER PERFORMED ANOTHER 1:2 MANUAL DILUTION OF THE ORIGINAL SAMPLE, PRODUCING A RESULT OF 52 NG/ML. THE CUSTOMER DID NOT KNOW WHICH RESULT WAS CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23942961 ↗