Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336170212), received between Aug 15, 2025 and Nov 18, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Negative Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Negative Result1
  • Non Reproducible Results1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2025False Negative Result; False Negative Result

    THE CUSTOMER STATED THEY RECEIVED DISCREPANT RESULTS FOR ONE PATIENT SAMPLE TESTED WITH ELECSYS HIV COMBI PT ON A COBAS 8000 E 602 MODULE. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED IN AN HIV COMBI PT VALUE OF 0.2 COI (NON-REACTIVE). THE SAMPLE WAS TESTED ON AN ABBOTT POINT OF CARE TEST STRIP FOR HIV 1/2 AG/AB COMBO AND FREE HIV-1 P24; IT WAS POSITIVE FOR THE P24 ANTIGEN. THE SAMPLE WAS REPEATED ON THE E 602 ANALYZER AND THE HIV COMBI PT VALUE WAS NEGATIVE. THE SAMPLE WAS SENT TO ANOTHER LABORATORY FOR CONFIRMATORY TESTING AND THE RESULTS FOR HIV-1 AND HIV-2 WERE BOTH NEGATIVE. HIV 1 AND HIV 2 RNA WERE UNDETECTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23585685 ↗
  • MalfunctionAug 15, 2025Non Reproducible Results; Non Reproducible Results

    THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HIV COMBI PT RESULTS FROM THE COBAS E 602 MODULE. THE INITIAL RESULT WAS 1.4 COI (REACTIVE). THE SAMPLE WAS REPEATED TWICE, WITH A RESULT OF >5000 COI (REACTIVE) EACH TIME. THE SAMPLE WAS REPEATED ON ANOTHER ANALYZER AT THE SITE, AND THE RESULT WAS >1.0 COI (REACTIVE). NO QUESTIONABLE RESULT WAS REPORTED OUTSIDE OF THE LABORATORY. THE REPEAT RESULT WAS BELIEVED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22816285 ↗

Other Test, HIV detection devices