Elecsys Anti-Tg 09005021190Recalls & reported problems
Roche Diagnostics GmbH · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07613336171479), received between Jun 14, 2024 and Dec 22, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “High Test Results” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- High Test Results5
- Low Test Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 22, 2025High Test Results; Low Test Results; High Test Results; Low Test Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT¿S SAMPLE TESTED WITH ELECSYS FT4 IV ASSAY, ELECSYS FT3 III ASSAY, ELECSYS VITAMIN D TOTAL III ASSAY, ELECSYS SHBG ASSAY, ELECSYS TESTOSTERONE II ASSAY, ELECSYS CORTISOL II ASSAY, ELECSYS PROLACTIN II ASSAY, ELECSYS FSH ASSAY, AND ELECSYS ANTI-TG ASSAY ON A COBAS E 801 ANALYTICAL UNIT. THIS MEDWATCH WILL APPLY TO THE ANTI-TG ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE VITAMIN D TOTAL III ASSAY, TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE SHBG ASSAY, TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE TESTOSTERONE II ASSAY, TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE FT3 III ASSAY, TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE CORTISOL II ASSAY, TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE PROLACTIN II ASSAY, TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE FSH ASSAY, AND TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE FT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23866557 ↗MalfunctionNov 21, 2025High Test Results; High Test Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR AN UNSPECIFIED NUMBER OF PATIENTS' SAMPLES TESTED WITH ELECSYS ANTI-TG ASSAY ON A COBAS E 801 ANALYTICAL UNIT, NOT CORRELATING WITH THE PATIENTS' CLINICAL STATUS. THE CUSTOMER ALLEGED THAT THE INITIAL RESULTS FOR THE THYROIDECTOMIZED PATIENTS WERE ESTIMATED TO BE 20 TO 29 IU/ML (WHICH WERE HIGHER THAN THE EXPECTED VALUES) AND THAT THEY DID NOT MATCH THE PATIENTS' PREVIOUS RESULTS OF <18 IU/ML. NO EXACT VALUES FOR THE INITIAL RESULTS WERE PROVIDED. DISCREPANT RESULTS FOR 1 PATIENT SAMPLE WERE PROVIDED: ON (B)(6) 2025: INITIAL RESULT: 21 IU/ML. ON (B)(6) 2025: REPEAT RESULT: 28 IU/ML.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23621616 ↗MalfunctionFeb 4, 2025High Test Results; High Test Results
THE INITIAL REPORTER QUESTIONED THYROID RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT COMPARED TO A COMPETITOR METHOD. OF THE DATA PROVIDED, THE RESULTS FOR FT3, T4, T3, ANTI-TG AND ANTI-TPO WERE DISCREPANT. REFER TO THE ATTACHED DATA FOR THE PATIENT RESULTS. THIS MEDWATCH WILL COVER ANTI-TG. REFER TO MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE T4 RESULTS, MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE T3 RESULTS, MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE FT3 RESULTS AND MEDWATCH WITH A1 PATIENT IDENTIFIER (B)(6) FOR INFORMATION ON THE ANTI-TPO RESULTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21307378 ↗MalfunctionSep 5, 2024High Test Results; High Test Results
THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS ANTI-TPO, ELECSYS ANTI-TG, ELECSYS ANTI-TSHR, AND ELECSYS TESTOSTERONE II RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT. THIS MEDWATCH COVERS THE ALLEGED RESULTS FOR THE ELECSYS ANTI-TG ASSAY. PLEASE REFER TO MEDWATCH WITH A1 - PATIENT IDENTIFIER = (B)(6) FOR THE ALLEGED ELECSYS ANTI-TSHR ASSAY RESULTS, MEDWATCH WITH A1 - PATIENT IDENTIFIER = (B)(6) FOR THE ALLEGED ELECSYS TESTOSTERONE II ASSAY RESULTS, AND MEDWATCH WITH A1 - PATIENT IDENTIFIER = (B)(6) FOR THE ALLEGED ELECSYS ANTI-TPO ASSAY RESULTS. REFER TO THE HIGHLIGHTED SECTION OF ATTACHMENT (B)(6) FOR PATIENT RESULTS. THE SAMPLES WERE REPEATED DUE TO THE RESULTS NOT MATCHING THE CLINICAL STATUS OF THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20146073 ↗MalfunctionJun 14, 2024High Test Results; High Test Results
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH ELECSYS ANTI-TG (ANTI-TG) ASSAY ON A COBAS E801 IMMUNOASSAY ANALYZER WHEN COMPARED TO A COMPETITOR'S ANALYZER (BECKMAN COULTER) AT A DIFFERENT FACILITY. INITIAL RESULT: 275 IU/ML AND WAS INTERPRETED AS POSITIVE. THE PHYSICIAN QUESTIONED THE RESULT AS IT DID NOT MATCH THE PATIENT'S CLINICAL HISTORY. ON (B)(6) 2024 THE ORIGINAL SAMPLE WAS SENT OUT TO ANOTHER LABORATORY AND REPEATED. REPEAT RESULT: 2.5 IU/ML (TESTED ON BECKMAN COULTER AND WAS INTERPRETED AS NEGATIVE). THE REPEAT RESULT MATCHED THE PATIENT'S CLINICAL HISTORY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19543023 ↗
Other SYSTEM, TEST, THYROID AUTOANTIBODY devices
- IMMULITE® 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) 10381659Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall17 reports
- Alinity 09P3521Abbott Ireland Diagnostics · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- Atellica IM aTPO 10995466Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- ADVIA Centaur® aTPO 10630886Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- ARCHITECT 2K47-27Abbott Ireland Diagnostics · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- ADVIA Centaur® aTPO +0 10630887Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- IMMULITE® 2000 Systems ATG Anti-TG Ab (Anti-Thyroglobulin Ab) 10381655Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
- IMMULITE® 2000 TSI 10876421Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, THYROID AUTOANTIBODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page