Skip to content

Elecsys Syphilis 09015051162Recalls & reported problems

Roche Diagnostics GmbH · Class 2

All models of Elecsys Syphilis (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 07613336183083), received between Nov 26, 2025 and May 29, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • False Positive Result2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2026False Positive Result

    THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS HTLV-I/II AND ELECSYS SYPHILIS RESULTS FOR 1 PATIENT AND QUESTIONABLE ELECSYS SYPHILIS RESULTS FOR 1 PATIENT ON A COBAS E 801 ANALYTICAL UNIT. PATIENT 1: THE ELECSYS SYPHILIS RESULT WAS REACTIVE ON (B)(6) 2026, THE (CAPTIA-G METHOD) CONFIRMATORY SYPHILIS TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE ELECSYS HTLV-I/II RESULT WAS REACTIVE. ON (B)(6) 2026, THE (WESTERN BLOT METHOD) CONFIRMATORY HTLV TEST RESULT WAS "CONFIRMED INDETERMINATE". PATIENT 2: ON (B)(6) 2026, THE ELECSYS SYPHILIS RESULT WAS REACTIVE THE (CAPTIA-G METHOD) CONFIRMATORY SYPHILIS TEST RESULT WAS "CONFIRMED NON-REACTIVE". THE SAMPLES WERE REPEATED BECAUSE OF THE INITIALLY REACTIVE RESULTS. THIS MEDWATCH WILL APPLY TO THE ELECSYS SYPHILIS ASSAY. PLEASE REFER TO THE MEDWATCH WITH A1. PATIENT IDENTIFIER (B)(6) FOR INFORMATION RELATED TO THE ELECSYS HTLV-I/II ASSAY RESULTS FOR PATIENT 1.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25333407 ↗
  • MalfunctionNov 26, 2025False Positive Result; False Positive Result

    THERE WAS AN ALLEGATION OF A QUESTIONABLE REACTIVE ELECSYS SYPHILIS RESULT FROM THE COBAS E 801 ANALYTICAL UNIT FOR ONE PATIENT. THE INITIAL RESULT WAS 8.18 COI (REACTIVE). THE FIRST REPEAT RESULT WAS 8.20 COI (REACTIVE). THE SECOND REPEAT RESULT WAS 8.17 COI (REACTIVE). THE SAMPLE WAS SENT FOR CONFIRMATORY TESTING USING THE CAPTIA SYPHILIS G APPLICATION, AND THE RESULT WAS NON-REACTIVE. THE NON-REACTIVE RESULT WAS DEEMED TO BE CORRECT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23657150 ↗

Other Test, donor, syphilis, antigens, treponemal devices

All Test, donor, syphilis, antigens, treponemal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page