Skip to content

Expiratory valve set 160245Recalls & reported problems

Hamilton Medical AG · Expiratory valve set for HAMILTON-C2/C3/C6, adult/pediatric, autoclavable · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 07630002801133), received between Jan 26, 2023 and Jan 26, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Gas/Air Leak” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Gas/Air Leak1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2023Obstruction of Flow; Gas/Air Leak; Obstruction of Flow; Gas/Air Leak

    THE FOLLOWING WAS REPORTED TO HAMILTON MEDICAL AG: REUSABLE EXPIRATORY VALVE (PN:160245) FOR HAMILTON-C6 AFTER THE REPROCESSING PROCESS, THE MEMBRANE STICKS TO THE VALVE BODY, WHICH IS NOT CHECKED DURING THE LEAK TEST. AN OCCLUSION ALARM APPEARS IMMEDIATELY WHEN VENTILATION IS STARTED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16245917 ↗

Other Tubing, Pressure And Accessories devices

All Tubing, Pressure And Accessories devices →