HiRes 90K™ CI HiFocus™ 1j Electrode CI-1400-01Recalls & reported problems

Advanced Bionics AG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Advanced Bionics HiFocus cochlear implant is a surgically placed device that restores hearing through electrical signals delivered to the inner ear. The 55 reports mostly describe implants being removed or replaced, along with pain at the implant site, signs of infection, sensations other than sound, and intermittent communication dropouts where the external equipment loses lock or stops producing sound. No deaths were reported; 33 injuries and 22 malfunctions were described. There are no recalls naming this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
54
Deaths reported
0
Injuries reported
32
Recalls naming it
0

Summary

The FDA's MAUDE database contains 54 adverse event reports that list this exact device (DI 07630016816253), received between Jun 29, 2018 and Aug 25, 2025 — classified as 32 injuries, 22 malfunctions.

The most frequently reported specific device problem is “Intermittent Communication Failure” (8 reports).

17 other recalls cover devices cleared under the same FDA submission (P960058); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem23
  • Appropriate Device Problem Term/Code Not Available9
  • Intermittent Communication Failure8
  • Electrostatic Discharge7
  • Material Protrusion/Extrusion7
  • Electrical /Electronic Property Problem6
  • Expulsion4
  • Migration or Expulsion of Device4

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available13
  • Unspecified Infection12
  • Pain10
  • Failure of Implant9
  • No Code Available6
  • Insufficient Information5
  • Inflammation4
  • Skin Infection4

Recent adverse event reports

The 25 most recent of 54 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 25, 2025Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available; Adverse Event Without Identified Device or Use Problem; Appropriate Device Problem Term/Code Not Available

    THE RECIPIENT REPORTEDLY ELECTED DEVICE EXPLANT DUE TO A TECHNOLOGY UPGRADE. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE. THE EXPLANTED DEVICE WILL NOT RETURN TO ADVANCED BIONICS FOR ANALYSIS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22891156 ↗
  • MalfunctionMay 30, 2025Electrical /Electronic Property Problem; Electrostatic Discharge; Adverse Event Without Identified Device or Use Problem; Electrical /Electronic Property Problem; Electrostatic Discharge; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING NON-AUDITORY SENSATIONS WITH DEVICE USE. THE RECIPIENT IS REPORTEDLY UNABLE TO USE THE DEVICE. EXTERNAL EQUIPMENT HAS BEEN EXCHANGED AND PROGRAMMING ADJUSTMENTS HAVE BEEN MADE; HOWEVER, THE ISSUE DID NOT RESOLVE. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Undesired Nerve Stimulation

    MAUDE report 22115300 ↗
  • MalfunctionApr 24, 2025Electrical /Electronic Property Problem; Electrostatic Discharge; Adverse Event Without Identified Device or Use Problem; Electrical /Electronic Property Problem; Electrostatic Discharge; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCES PAIN WITH DEVICE USE. PROGRAMMING ADJUSTMENTS WERE MADE; HOWEVER, THE ISSUE DID NOT IMPROVE. DEVICE TESTING COULD NOT BE PERFORMED DUE TO PAIN. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Failure of Implant; Pain

    MAUDE report 21900468 ↗
  • MalfunctionApr 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY ELECTED REVISION SURGERY FOR A TECHNOLOGY UPGRADE. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21743464 ↗
  • MalfunctionMar 31, 2025Intermittent Communication Failure; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING NO LOCK AND NO SOUND. EXTERNAL EQUIPMENT WAS EXCHANGED; HOWEVER, THE ISSUE DID NOT RESOLVE. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant

    MAUDE report 21735464 ↗
  • InjuryMar 20, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE COMPANY WAS INFORMED THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE. ADVANCED BIONICS IS STILL IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21652196 ↗
  • MalfunctionAug 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING OPEN ELECTRODES. DEVICE TESTING YIELDED ABNORMAL RESULTS. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19908015 ↗
  • InjuryJun 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED CHRONIC MIDDLE EAR INFECTIONS. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS NOT REIMPLANTED. ADVANCED BIONICS IS STILL IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Unspecified Infection

    MAUDE report 19511004 ↗
  • MalfunctionMar 8, 2024Loose or Intermittent Connection; Mechanical Problem; Intermittent Communication Failure; Loose or Intermittent Connection; Mechanical Problem; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING INTERMITTENT LOCK, DESPITE DEVICE TESTING WITHIN NORMAL LIMITS. EXTERNAL EQUIPMENT WAS EXCHANGED, HOWEVER, THE ISSUE DID NOT RESOLVE. A CT SCAN REVEALED NO ABNORMALITIES WITH THE IMPLANT PROPER LEAD PLACEMENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18871535 ↗
  • InjuryOct 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING A SKIN FLAP INFECTION. THE RECIPIENT IS PRESENTING WITH NECROSIS WHICH REQUIRED SURGICAL INTERVENTION. ADVANCED BIONICS IS STILL IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Necrosis; Unspecified Infection

    MAUDE report 17877472 ↗
  • InjurySep 27, 2023Electrical /Electronic Property Problem; Electrostatic Discharge; Intermittent Communication Failure; Electrical /Electronic Property Problem; Electrostatic Discharge; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING LOSS OF LOCK. EXTERNAL EQUIPMENT WAS EXCHANGED; HOWEVER, THE ISSUE WAS NOT RESOLVED. INTEGRITY TESTING COULD NOT BE COMPLETED DUE TO NO LOCK. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 17827689 ↗
  • InjurySep 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING SWELLING AND PAIN AT IMPLANT SITE DUE TO A FLIPPED MAGNET FOLLOWING AN MRI. BANDAGING PROTOCOL WAS NOT FOLLOWED. THE RECIPIENT IS REPORTEDLY USING ANTI-INFLAMMATORY MEDICATION (TYPE UNKNOWN). THE RECIPIENT IS REPORTEDLY USING THE DEVICE.

    Patient problems: Pain; Swelling/ Edema

    MAUDE report 17719886 ↗
  • MalfunctionJul 14, 2023Electrical /Electronic Property Problem; Electrostatic Discharge; Intermittent Communication Failure; Electrical /Electronic Property Problem; Electrostatic Discharge; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING PAIN. THE RECIPIENT REFUSES TO WEAR DEVICE. EXTERNAL EQUIPMENT WAS EXCHANGED AND PROGRAMMING ADJUSTMENTS WERE MADE, HOWEVER, THE ISSUES DID NOT RESOLVE. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE.

    Patient problems: Pain

    MAUDE report 17328330 ↗
  • MalfunctionApr 3, 2023Adverse Event Without Identified Device or Use Problem; Intermittent Communication Failure; Adverse Event Without Identified Device or Use Problem; Intermittent Communication Failure

    THE RECIPIENT IS, REPORTEDLY, EXPERIENCING LOSS OF SOUND AND OPEN ELECTRODES. EXTERNAL EQUIPMENT WAS EXCHANGED AND PROGRAMMING ADJUSTMENTS WERE MADE; HOWEVER, THE ISSUE DID NOT RESOLVE. DEVICE TESTING REVEALED ABNORMAL RESULTS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 16663612 ↗
  • InjuryJan 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem; Output Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED OVERLY LOUD SOUND WITH DEVICE USE. DEVICE TESTING WAS WITHIN NORMAL LIMITS, PROGRAMMING ADJUSTMENTS WERE MADE AND IMPROVEMENT WAS NOTED. THE RECIPIENT'S DEVICE WAS EXPLANTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 16132770 ↗
  • MalfunctionDec 7, 2022Mechanical Problem; Output Problem; Intermittent Communication Failure; Mechanical Problem; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING LOSS OF LOCK. DEVICE TESTING COULD NOT BE COMPLETED DUE TO LOSS OF LOCK. REVISION SURGERY WILL BE SCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 15936241 ↗
  • MalfunctionDec 7, 2022Mechanical Problem; Output Problem; Intermittent Communication Failure; Mechanical Problem; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING LOSS OF LOCK. DEVICE TESTING COULD NOT BE COMPLETED DUE TO LOSS OF LOCK. REVISION SURGERY WILL BE SCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 15936204 ↗
  • InjuryOct 4, 2022Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING RECURRENT CHRONIC OTITIS MEDIA FOR THE PAST FIVE MONTHS. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS NOT REIMPLANTED. THE RECIPIENT'S INFECTION HAS IMPROVED.

    Patient problems: Infectious Otitis Media; Skin Infection

    MAUDE report 15536012 ↗
  • MalfunctionAug 26, 2022Expulsion; Expulsion; Material Protrusion/Extrusion; Output Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED EXTRUSION AND INFLAMMATION THAT WAS TREATED (TYPE UNKNOWN). THE RECIPIENT IS REPORTEDLY EXPERIENCING DECREASED PERFORMANCE. A REVIEW OF THE TEST DATA INDICATES POTENTIAL IMPEDANCE ISSUES. THE RECIPIENT WILL CONTINUE TO MONITORED.

    Patient problems: Inflammation; Unspecified Infection

    MAUDE report 15307532 ↗
  • InjuryAug 24, 2022Adverse Event Without Identified Device or Use Problem; Material Protrusion/Extrusion; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED RECURRENT MIDDLE EAR INFECTION AND DRAINAGE SINCE 2020. THE RECIPIENT WAS PRESCRIBED ANTIBIOTICS (TYPE UNKNOWN) AND TREATMENT. A CT SCAN REVELED EXTRUSION. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WILL NOT BE REIMPLANTED.

    Patient problems: Fluid Discharge; Skin Infection; Unspecified Infection

    MAUDE report 15283666 ↗
  • InjuryMay 27, 2022Expulsion; Material Protrusion/Extrusion; Expulsion

    THE RECIPIENT REPORTEDLY EXPERIENCED SWELLING, PAIN AND DEVICE EXTRUSION STARTING IN (B)(6) 2021. THE RECIPIENT'S DEVICE WAS EXPLANTED.

    Patient problems: Pain; Swelling/ Edema; Unspecified Infection

    MAUDE report 14522159 ↗
  • MalfunctionFeb 2, 2022Electrical /Electronic Property Problem; Electrostatic Discharge; Electrical /Electronic Property Problem; Loose or Intermittent Connection; Electrostatic Discharge; Appropriate Device Problem Term/Code Not Available

    THE RECIPIENT IS REPORTEDLY EXPERIENCING NO LOCK. DEVICE TESTING COULD NOT BE COMPLETED DUE TO NO LOCK. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE.

    Patient problems: Insufficient Information; Pain; Undesired Nerve Stimulation

    MAUDE report 13428690 ↗
  • InjuryFeb 1, 2022Intermittent Communication Failure; Connection Problem; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING LOSS OF LOCK. EXTERNAL EQUIPMENT WAS EXCHANGED, HOWEVER THE ISSUES DID NOT RESOLVE. REVISION SURGERY WILL BE SCHEDULED

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 13416989 ↗
  • InjuryJun 18, 2021Expulsion; Adverse Event Without Identified Device or Use Problem; Expulsion; Material Protrusion/Extrusion; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED TRAUMA AT THE IMPLANT SITE. ON (B)(6) 2019, THE RECIPIENT PRESENTED WITH EDEMA AND WAS TREATED WITH AMOXICILLIN AND SULBACTAM. THE EDEMA RESOLVED. ON (B)(6) 2020, THE EDEMA RECURRED. THE RECIPIENT WAS TREATED WITH CETIBUTEN, HOWEVER, THE ISSUE DID NOT RESOLVE. ON (B)(6) 2020, THE RECIPIENT PRESENTED WITH AN EXTRUDED DEVICE. THE RECIPIENT UNDERWENT SKIN FLAP REVISION SURGERY. ON (B)(6) 2021, THE RECIPIENT PRESENTED WITH AN EXTRUDED DEVICE AGAIN, AND THE RECIPIENT UNDERWENT EXPLANT SURGERY. THE RECIPIENT WAS NOT REIMPLANTED. THE RECIPIENT RECOVERED WELL.

    Patient problems: Failure of Implant; Skin Infection; Swelling/ Edema

    MAUDE report 12022519 ↗
  • MalfunctionApr 2, 2021Electrical /Electronic Property Problem; Electrostatic Discharge; Appropriate Device Problem Term/Code Not Available; Electrical /Electronic Property Problem; Electrostatic Discharge

    THE RECIPIENT REPORTEDLY EXPERIENCED AN ELECTROSTATIC DISCHARGE (ESD) EVENT. THE RECIPIENT PRESENTS WITH PAIN AND NON-AUDITORY SENSATIONS. PROGRAMMING ADJUSTMENTS WERE MADE, HOWEVER, THE ISSUE DID NOT RESOLVE. DEVICE TESTING REVEALED ABNORMAL RESULTS. REVISION SURGERY WILL BE SCHEDULED.

    Patient problems: Pain; Undesired Nerve Stimulation

    MAUDE report 11611311 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P960058. One submission often covers many models, so check the product and lot details in each record.

  • Z-2283-2026 · Apr 30, 2026 · Open, Classified
    Brand Name: HiResolution" Bionic Ear System Product Name: HiRes Ultra 3D CI HiFocus Mid Scala Model/Catalog Number: CI-1601-04 Software Version: N/A Product Description: The HiReso
  • Z-1950-2026 · Mar 23, 2026 · Open, Classified
    Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; C
  • Z-1813-2025 · Mar 5, 2025 · Open, Classified
    HiResolution" Bionic Ear System - M Battery Charger Model Number CI-5607
  • Z-2033-2020 · Feb 17, 2020 · Terminated
    AB HiResolution Bionic Ear System, HiRes Ultra CI, MS Electrode - Product Usage: The device is intended to restore a level of auditory sensation to patients with profound sensorine
  • Z-2032-2020 · Feb 17, 2020 · Terminated
    AB HiResolution Bionic Ear System, HiRes Ultra 3D CI, HiFocus SlimJ Electrode - Product Usage: The device is intended to restore a level of auditory sensation to patients with prof
  • Z-1441-2020 · Apr 4, 2019 · Terminated
    SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave
  • Z-1259-2020 · Mar 18, 2019 · Terminated
    HiResolution Bionic Ear System HiRes 90K Advantage CI HiFocus MS Electrode REF CI-1500-04. Cochlear Implant.
  • Z-1085-2019 · Feb 6, 2019 · Terminated
    Nada CI Q90 Sound Processor, Silver Gray, Model Cl-5280-140.
  • Z-1083-2019 · Feb 6, 2019 · Terminated
    Nada CL Q30 Sound Processor, Sand Beige, Model Cl-5260-120; Silver Gray, Model Cl-5260-140; and Velvet Black, Model Cl-5260-150.
  • Z-1084-2019 · Feb 6, 2019 · Terminated
    Nada CI Q70 Sound Processor, Sand Beige, Model Cl-5245-120; Silver Gray, Model Cl-5245-140; and Velvet Black, Model Cl-5245-150.
  • Z-2039-2009 · Jun 10, 2009 · Terminated
    Advanced Bionics HiRes 90K Cochlear Implant, Model: CI-1400-01 and CI-1400-02H, Implant Magnet Case. Cochlear implant intended to restore a level of auditory sensation to patients
  • Z-0956-2009 · Jul 24, 2008 · Terminated
    HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The HiRes90K headpiece is used in conjunction with spee
  • Z-0959-2009 · Jul 24, 2008 · Terminated
    S-Series Short Range Headpiece Model Number MMT-5300-00 (beige) and MMT-5300-10 (brown) ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The headpiece is used i
  • Z-0957-2009 · Jul 24, 2008 · Terminated
    Platinum Series Headpiece Model Number AB-7300. The headpiece is used in conjunction with speech processors manufactured by Advanced Bionics. The Harmony HiResolution Bionic Ear
  • Z-0958-2009 · Jul 24, 2008 · Terminated
    S-Series Long Range Headpiece Model Number AB-5301-00 (beige) and AB-5301-10 (brown) ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The headpiece is used in

…and 2 more.