Quick summary

This Advanced Bionics HiRes 90K cochlear implant sends sound signals to the inner ear. Of its 127 reports, 96 are injuries and 31 are malfunctions, with no deaths; some describe the magnet shifting after MRI scans or infections around the implant. Reports fell to 3 in the last 12 months from 10 in the 12 months before. 1 recall, closed and started in March 2019, concerns materials missing from a Japanese biocompatibility list.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
127
Deaths reported
0
Injuries reported
96
Recalls naming it
1

Summary

The FDA's MAUDE database contains 127 adverse event reports that list this exact device (DI 07630016832185), received between Jun 29, 2018 and Apr 13, 2026 — classified as 96 injuries, 31 malfunctions.

The most frequently reported specific device problem is “Material Protrusion/Extrusion” (16 reports).

1 FDA recall record name this device's identifier.

16 other recalls cover devices cleared under the same FDA submission (P960058); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 2 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 3 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 18, 2019Class IIITerminated

    Three materials used in the makeup of the implant were mistakenly left off the list of bio-compatible materials in the firm's previously approved submission to the Pharmaceuticals and Medical Devices Agency (PMDA) of Japan.

    Z-1259-2020 · Root cause: Other · Recall details · FDA record ↗

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem59
  • Material Protrusion/Extrusion16
  • Expulsion13
  • Appropriate Device Problem Term/Code Not Available11
  • Electrical /Electronic Property Problem10
  • Therapeutic or Diagnostic Output Failure10
  • Migration or Expulsion of Device9
  • Intermittent Communication Failure8

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available36
  • Unspecified Infection28
  • Pain24
  • Failure of Implant16
  • Insufficient Information11
  • Head Injury9
  • Discomfort8
  • No Code Available8

Recent adverse event reports

The 25 most recent of 127 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 13, 2026Device Dislodged or Dislocated

    THE RECIPIENT REPORTEDLY EXPERIENCED MAGNET DISPLACEMENT FOLLOWING AN MRI, DUE TO THE BANDAGE PROTOCOL NOT BEING FOLLOWED. THE MAGNET WAS SUCCESSFULLY REPOSITIONED THROUGH EXTERNAL MANIPULATION. THE RECIPIENT EXPERIENCED SOME PAIN, WHICH HAS SINCE RESOLVED, AND THE DEVICE IS NOW FUNCTIONING PROPERLY.

    Patient problems: Pain

    MAUDE report 24866561 ↗
  • InjuryFeb 24, 2026Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED A CHOLESTEATOMA INFECTION AND REQUIRED SURGERY TO REMOVE THE INFECTED TISSUE. THE INFECTION REPORTEDLY REACHED THE IMPLANT, AND THE RECIPIENT'S DEVICE WAS EXPLANTED DUE TO INFECTION FROM CHOLESTEATOMA. THE RECIPIENT WAS IMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE.

    Patient problems: Cholesteatoma

    MAUDE report 24437088 ↗
  • MalfunctionNov 17, 2025Intermittent Communication Failure; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING LOSS OF LOCK. EXTERNAL EQUIPMENT WAS EXCHANGED; HOWEVER, THE ISSUE DID NOT RESOLVE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; No Clinical Signs, Symptoms or Conditions

    MAUDE report 23577881 ↗
  • MalfunctionOct 17, 2025Mechanical Problem; Impedance Problem; Mechanical Problem; Impedance Problem

    A REVIEW OF THE RECIPIENT'S TEST DATA INDICATED IMPEDANCE ISSUES. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS RE-IMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant

    MAUDE report 23330240 ↗
  • InjuryJul 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE COMPANY WAS INFORMED THAT THE SURGEON MISTAKENLY EXPLANTED THE WRONG DEVICE. ADVANCED BIONICS IS CURRENTLY IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22531850 ↗
  • InjuryJul 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED A FLIPPED MAGNET FOLLOWING AN MRI. HEAD BANDAGING PROCEDURE WAS FOLLOWED. THE RECIPIENT HAS RESUMED DEVICE USE. REVISION SURGERY IN UNDER CONSIDERATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 22394649 ↗
  • InjuryJul 1, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED A HEAD TRAUMA EVENT. THE RECIPIENT REPORTEDLY RECEIVED STITCHES CLOSE TO THE IMPLANT SITE. ADVANCED BIONICS IS STILL IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Head Injury

    MAUDE report 22372675 ↗
  • InjuryJun 16, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE COMPANY WAS INFORMED THAT THE RECIPIENT'S DEVICE WAS EXPLANTED DUE TO THINNING OF THE SKIN CAUSED BY AN INFECTION. THE RECIPIENT WAS REPORTEDLY HOSPITALIZED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE.

    Patient problems: Unspecified Infection

    MAUDE report 22211257 ↗
  • MalfunctionMay 22, 2025Electrical /Electronic Property Problem; Mechanical Problem; Electrostatic Discharge; Electrical /Electronic Property Problem; Mechanical Problem; Electrostatic Discharge

    THE RECIPIENT IS REPORTEDLY EXPERIENCING PAIN AND NO SOUND WITH DEVICE USE. DEVICE TESTING REVEALED ABNORMAL RESULTS. THE RECIPIENT IS UNABLE TO WEAR THE DEVICE DUE TO PAIN. EXTERNAL EQUIPMENT AND THE EXTERNAL MAGNET WERE EXCHANGED, AND PROGRAMMING ADJUSTMENTS WERE MADE: HOWEVER, THE ISSUES WERE NOT RESOLVED. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Failure of Implant; Pain

    MAUDE report 22069663 ↗
  • InjuryApr 4, 2025Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2025, THE RECIPIENT REPORTEDLY UNDERWENT AN MRI. BANDAGING PROTOCOL WAS FOLLOWED. THE RECIPIENT IS REPORTEDLY EXPERIENCING PAIN AND A DISLODGED MAGNET. THE RECIPIENT WAS ADVISED TO CEASE DEVICE USE. ADVANCED BIONICS IS STILL IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Pain

    MAUDE report 21773074 ↗
  • InjuryApr 2, 2025Intermittent Communication Failure; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING PAIN AND LOSS OF LOCK. THE RECIPIENT REPORTEDLY WALKED INTO AN MRI SUITE AND EXPERIENCED PAIN. NO MRI WAS PERFORMED. MAGNET REPLACEMENT SURGERY HAS BEEN SCHEDULED.

    Patient problems: Pain

    MAUDE report 21751295 ↗
  • InjuryJan 7, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ON (B)(6) 2024, THE RECIPIENT REPORTEDLY UNDERWENT AND MRI AND BANDAGING PROTOCOL WAS NOT FOLLOWED. THE RECIPIENT'S MAGNET REPORTEDLY ROTATED. ON (B)(6) 2024, THE RECIPIENT'S MAGNET WAS REVISED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 21093429 ↗
  • MalfunctionDec 2, 2024Electrical /Electronic Property Problem; Electrostatic Discharge; Intermittent Communication Failure; Electrical /Electronic Property Problem; Electrostatic Discharge; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING PAIN WITH DEVICE USE AFTER GOING DOWN A SLIDE. THE RECIPIENT IS PRESENTING WITH INTERMITTENCIES. EXTERNAL EQUIPMENT WAS EXCHANGED, HOWEVER THE ISSUE DID NOT RESOLVE. DEVICE TESTING REVEALED ABNORMAL RESULTS.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Pain

    MAUDE report 20824485 ↗
  • InjuryOct 18, 2024Electrical /Electronic Property Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Communication Failure; Electrical /Electronic Property Problem; Adverse Event Without Identified Device or Use Problem; Intermittent Communication Failure

    THE RECIPIENT IS REPORTEDLY EXPERIENCING NO SOUND FOLLOWING A HEAD TRAUMA INCIDENT. EXTERNAL EQUIPMENT WAS EXCHANGED; HOWEVER, THE ISSUE DID NOT RESOLVE. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Head Injury

    MAUDE report 20482261 ↗
  • InjuryOct 3, 2024Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING A DISPLACED MAGNET FOLLOWING AN MRI. THE RECIPIENT IS PRESENTING WITH PAIN. REVISION SURGERY IS UNDER CONSIDERATION.

    Patient problems: Pain

    MAUDE report 20364232 ↗
  • InjurySep 24, 2024Expulsion; Expulsion

    THE RECIPIENT REPORTEDLY EXPERIENCED AN INFECTION AT THE IMPLANT SITE. THE RECIPIENT PRESENTED WITH DEVICE EXTRUSION. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS REIMPLANTED WITH ANOTHER ADVANCED BIONICS COCHLEAR DEVICE. THE RECIPIENT HEALED.

    Patient problems: Unspecified Infection

    MAUDE report 20294353 ↗
  • InjurySep 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING SKIN BREAKDOWN OVER MAGNET SITE. THE RECIPIENT WAS REPORTEDLY ADVISED TO CEASE DEVICE USE. THE RECIPIENT HAS EXPERIENCED SIGNIFICANT WEIGHT LOSS.

    Patient problems: Tissue Breakdown

    MAUDE report 20164358 ↗
  • InjuryJul 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS EXPERIENCING TINNITUS. DEVICE TESTING REVEALED RESULTS WITHIN NORMAL LIMITS. REVISION SURGERY IS SCHEDULED.

    Patient problems: Tinnitus

    MAUDE report 19715446 ↗
  • InjuryMay 6, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING DECREASED PERFORMANCE. EXTERNAL EQUIPMENT WAS EXCHANGED AND PROGRAMING ADJUSTMENTS WERE MADE, HOWEVER THE ISSUE DID NOT RESOLVE. THE RECIPIENT HAS GRANULATION IN THE EAR CANAL AND WAS PRESCRIBED AN OINTMENT (TYPE UNKNOWN). IT IS BELIEVED THE RECIPIENT MAY HAVE OSSIFICATION OF THE COCHLEA.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19249547 ↗
  • InjuryMay 1, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT'S DEVICE WAS EXPLANTED TO TREAT A CHOLESTEATOMA. THE RECIPIENT WILL BE REIMPLANTED AT A LATER DATE. ADVANCED BIONICS IS STILL IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19226711 ↗
  • MalfunctionApr 23, 2024Mechanical Problem; Adverse Event Without Identified Device or Use Problem; Mechanical Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED NON AUDITORY SENSATIONS. EXTERNAL EQUIPMENT WAS EXCHANGED AND PROGRAMMING ADJUSTMENTS WERE MADE, HOWEVER THE ISSUE DID NOT RESOLVE. SOME DEVICE TESTING COULD NOT BE COMPLETED DUE TO RECIPIENT DISCOMFORT. DEVICE TESTING COMPLETED REVEALED ABNORMAL RESULTS. REVISION SURGERY WILL BE SCHEDULED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 19164794 ↗
  • InjuryJan 11, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING RECURRENT INFECTIONS. A SURGERY TO RESOLVE THE INFECTION WAS REPORTEDLY PERFORMED. THE RECIPIENT IS PRESENTING WITH A LOSS OF SOUND. DEVICE TESTING WAS WITHIN NORMAL LIMITS. REVISION SURGERY WILL BE SCHEDULED.

    Patient problems: Unspecified Infection

    MAUDE report 18495906 ↗
  • InjuryNov 10, 2023Expulsion; Expulsion

    THE RECIPIENT REPORTEDLY EXPERIENCED CHOLESTEATOMA AND DEVICE EXTRUSION. THE RECIPIENT'S DEVICE WAS EXPLANTED. THE RECIPIENT WAS NOT REIMPLANTED. THE RECIPIENT HAS HEALED AFTER SURGERY.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Infection

    MAUDE report 18112854 ↗
  • InjuryOct 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT IS REPORTEDLY EXPERIENCING AN INFECTION ON HIS HEAD BEHIND THE PINNA. THE RECIPIENT RECEIVED TREATMENT (TYPE UNKNOWN) AND WAS INSTRUCTED BY A NURSE TO USE AQUAPHOR.

    Patient problems: Unspecified Infection

    MAUDE report 17968273 ↗
  • InjuryOct 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE RECIPIENT REPORTEDLY EXPERIENCED A CHOLESTEATOMA. THE RECIPIENT RECEIVED TREATMENT (TYPE UNKNOWN). ADVANCED BIONICS IS STILL IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION. WHEN ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Unspecified Infection

    MAUDE report 17961083 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P960058. One submission often covers many models, so check the product and lot details in each record.

  • Z-2283-2026 · Apr 30, 2026 · Open, Classified
    Brand Name: HiResolution" Bionic Ear System Product Name: HiRes Ultra 3D CI HiFocus Mid Scala Model/Catalog Number: CI-1601-04 Software Version: N/A Product Description: The HiReso
  • Z-1950-2026 · Mar 23, 2026 · Open, Classified
    Brand Name: HiResolution" Bionic Ear System Product Name: M Zn-Air Battery Pak Model/Catalog Number: CI-5501-110; CI-5501-120; CI-5501-130; CI-5501-140; CI-5501-150; CI-5501-190; C
  • Z-1813-2025 · Mar 5, 2025 · Open, Classified
    HiResolution" Bionic Ear System - M Battery Charger Model Number CI-5607
  • Z-2032-2020 · Feb 17, 2020 · Terminated
    AB HiResolution Bionic Ear System, HiRes Ultra 3D CI, HiFocus SlimJ Electrode - Product Usage: The device is intended to restore a level of auditory sensation to patients with prof
  • Z-2033-2020 · Feb 17, 2020 · Terminated
    AB HiResolution Bionic Ear System, HiRes Ultra CI, MS Electrode - Product Usage: The device is intended to restore a level of auditory sensation to patients with profound sensorine
  • Z-1441-2020 · Apr 4, 2019 · Terminated
    SoundWave Professional Suite Software 3.2, Ref CI-6055-014, SPN 058-023-32085, UDI: (01)07630016853227, Logical SoundWave Professional Suite 3.2, CE 0123 - Product Usage: Soundwave
  • Z-1084-2019 · Feb 6, 2019 · Terminated
    Nada CI Q70 Sound Processor, Sand Beige, Model Cl-5245-120; Silver Gray, Model Cl-5245-140; and Velvet Black, Model Cl-5245-150.
  • Z-1085-2019 · Feb 6, 2019 · Terminated
    Nada CI Q90 Sound Processor, Silver Gray, Model Cl-5280-140.
  • Z-1083-2019 · Feb 6, 2019 · Terminated
    Nada CL Q30 Sound Processor, Sand Beige, Model Cl-5260-120; Silver Gray, Model Cl-5260-140; and Velvet Black, Model Cl-5260-150.
  • Z-2039-2009 · Jun 10, 2009 · Terminated
    Advanced Bionics HiRes 90K Cochlear Implant, Model: CI-1400-01 and CI-1400-02H, Implant Magnet Case. Cochlear implant intended to restore a level of auditory sensation to patients
  • Z-0958-2009 · Jul 24, 2008 · Terminated
    S-Series Long Range Headpiece Model Number AB-5301-00 (beige) and AB-5301-10 (brown) ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The headpiece is used in
  • Z-0956-2009 · Jul 24, 2008 · Terminated
    HR90K Platinum Series Headpiece Model Number AB-7300-001 ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The HiRes90K headpiece is used in conjunction with spee
  • Z-0959-2009 · Jul 24, 2008 · Terminated
    S-Series Short Range Headpiece Model Number MMT-5300-00 (beige) and MMT-5300-10 (brown) ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The headpiece is used i
  • Z-0957-2009 · Jul 24, 2008 · Terminated
    Platinum Series Headpiece Model Number AB-7300. The headpiece is used in conjunction with speech processors manufactured by Advanced Bionics. The Harmony HiResolution Bionic Ear
  • Z-0759-06 · Mar 8, 2006 · Terminated
    The unimplanted HiRes90K Implantable Cochlear Stimulators (ICS) manufactured with feedthrus from a specific supplier are being removed from the market. Model number CI-1400-02H.

…and 1 more.

Other Implant, Cochlear devices

All Implant, Cochlear devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page