Quadra H femoral stems 01.12.036Recalls & reported problems
Medacta International SA · QUADRA Stem Quadra Laterized HAP 12/14 S6 · Class 2 · Implant
All models of Quadra H femoral stems (32)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07630030802188), received between Dec 22, 2017 and Jul 2, 2025 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K082792, K250450); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth2
- Osseointegration Problem2
- Appropriate Device Problem Term/Code Not Available1
- Device Slipped1
- Loss of Osseointegration1
Patient problems
- Inadequate Osseointegration3
- Aseptic loosening1
- Joint Disorder1
- No Known Impact Or Consequence To Patient1
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryJul 2, 2025Loss of Osseointegration; Loss of Osseointegration
REVISION DUE TO ASEPTIC STEM LOOSENING PERFORMED ABOUT 5 YEARS AND 3 MONTHS AFTER THE PRIMARY SURGERY. STEM, HEAD, AND LINER REVISED.
Patient problems: Aseptic loosening
MAUDE report 22380967 ↗MalfunctionMar 19, 2025Osseointegration Problem; Osseointegration Problem
THE PATIENT HAS MINIMAL PAIN AND PRESENTS SIGNS OF LOOSENING/STRESS SHIELDING. NO REVISION SURGERY IS PLANNED.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 21640390 ↗InjuryNov 30, 2021Appropriate Device Problem Term/Code Not Available; Loosening of Implant Not Related to Bone-Ingrowth
THE LOOSENING OF THE CUP WAS NOTICED IN A CONTROL X-RAY, THE LOOSENING OF THE STEM WAS NOTICED DURING SURGERY. LOOSENING WAS MAYBE DUE TO A TRAUMA OF THE PATIENT. REVISION SURGERY WAS PERFORMED 15 YEARS AND 2 MONTHS AFTER PRIMARY, ALL IMPLANTS WERE REPLACED.
Patient problems: Inadequate Osseointegration
MAUDE report 12891360 ↗InjuryMar 24, 2021Osseointegration Problem; Osseointegration Problem
STEM LOOSENING 12 YEARS AND 2 MONTHS AFTER THE PRIMARY SURGERY. NO REVISION HAS BEEN PERFORMED AT THE MOMENT.
Patient problems: Inadequate Osseointegration
MAUDE report 11560724 ↗InjuryJan 13, 2020Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
REVISION SURGERY WILL BE PERFORMED IN (B)(6) 2020 DUE TO STEM LOOSENING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9580421 ↗InjuryDec 22, 2017Device Slipped
PATIENT REPORTED PAIN.THE SURGEON DETERMINED IT LOOKED LOOSE. PERFORMED REVISION SURGERY AND FOUND THAT THE STEM WAS LOOSE
Patient problems: Joint Disorder
MAUDE report 7142938 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082792, K250450. One submission often covers many models, so check the product and lot details in each record.
- Z-0791-2013 · Dec 17, 2012 · TerminatedMedacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page