Quadra H femoral stems 01.12.22SNRecalls & reported problems
Medacta International SA · QUADRA "H" Short neck, cem.less, std, HAP S.2 · Class 2 · Implant
All models of Quadra H femoral stems (32)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 16
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 07630030802621), received between Jun 27, 2017 and Mar 19, 2025 — classified as 16 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (5 reports).
1 other recall cover devices cleared under the same FDA submission (K082792, K250450); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loose or Intermittent Connection5
- Adverse Event Without Identified Device or Use Problem2
- Osseointegration Problem2
- Appropriate Device Problem Term/Code Not Available1
- Device Operates Differently Than Expected1
- Device Slipped1
- Difficult to Remove1
- Failure to Osseointegrate1
Patient problems
- Inadequate Osseointegration8
- Bone Fracture(s)3
- Pain3
- Joint Disorder2
- Failure of Implant1
- No Information1
- Osteolysis1
- Unspecified Infection1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 19, 2025Osseointegration Problem; Osseointegration Problem
THE PATIENT HAS MODERATE PAIN AND PRESENTS SIGNS OF LOOSENING/STRESS SHIELDING. NO REVISION SURGERY IS PLANNED.
Patient problems: Inadequate Osseointegration; Pain
MAUDE report 21640394 ↗InjuryJul 26, 2024Failure to Osseointegrate; Failure to Osseointegrate
AT ABOUT 9 YEARS 8 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND COMPETITOR'S CUP AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH A COMPETITOR STEM AND HEAD AND REVISED THE COMPETITOR CUP AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 19836131 ↗InjuryNov 3, 2023Osseointegration Problem; Osseointegration Problem
REVISION SURGERY AT ABOUT 6 YEARS AND 11 MONTHS POST-PRIMARY DUE TO STEM LOOSENING. THE STEM, HEAD AND LINER HAVE BEEN REVISED. NO PROBLEM WERE AT LINER AND HEAD LEVEL.
Patient problems: Inadequate Osseointegration
MAUDE report 18063839 ↗InjuryMar 25, 2021Loose or Intermittent Connection; Loss of Osseointegration
THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH A COMPETITOR STEM AND HEAD AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER 7 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 11568319 ↗InjuryMar 17, 2021Difficult to Remove; Malposition of Device
3 MONTHS AFTER THE PRIMARY SURGERY THE STEM SUBSIDED AND IT SHOULD HAVE BEEN REVISED BUT THE SURGEON COULD NOT EXTRACT IT EVEN AFTER CONSIDERABLE EFFORT. THE STEM REMAINED IN SITU AND A BIGGER HEAD WAS IMPLANTED.
Patient problems: Failure of Implant
MAUDE report 11504350 ↗InjuryMar 16, 2021Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING PAIN. POST-OP X-RAYS REVEALED THAT THE STEM POTTED DISTALLY. THE SURGEON REVISED THE STEM AND HEAD AND ADDED A SLEEVE 3 YEARS 2 MONTHS AFTER THE PRIMARY.
Patient problems: Inadequate Osseointegration
MAUDE report 11493346 ↗InjuryJan 15, 2021Migration or Expulsion of Device
REVISION SURGERY PERFORMED AFTER 6 YEARS FROM THE PRIMARY SURGERY DUE TO THIGH PAIN & GROIN PAIN ASSOCIATED WITH RADIOLUCENT LINES AROUND THE FEMORAL COMPONENT. REVISION COMPLETED VIA A POSTERIOR APPROACH. THE STEM AND CUP WERE WELL FIXED. TRUNIONOSIS NOTED BY THE SURGEON.
Patient problems: Osteolysis; Pain
MAUDE report 11177765 ↗InjuryJul 17, 2020Loose or Intermittent Connection
THE PATIENT WAS APPROXIMATELY 3 YEARS OF POST-PRIMARY REPLACEMENT (2017) CAME BACK TO THE SURGEON WITH COMPLAINTS OF PAIN. THE STEM WAS FOUND TO BE LOOSE PROXIMALLY BUT WELL FIXED DISTALLY. THE SURGEON STATED THE PATIENT HAD BEEN WATER SKIING ON THEIR REPLACEMENT.
Patient problems: Inadequate Osseointegration
MAUDE report 10288602 ↗InjuryDec 3, 2019Migration; Migration
REVISION SURGERY 5 YEARS AFTER THE PRIMARY DUE TO STEM SUBSIDENCE. THE QUADRA STEM WAS REMOVED VIA A POSTERIOR APPROACH, THE SURGEON COMMENTED THAT IT HAD NOT BONDED. A MINIMAX STEM WAS USED TO COMPLETE THE REVISION SURGERY WITH A NEW CERAMIC HEAD. A SMALL AMOUNT OF BONE GRAFT DISTALLY WAS ALSO USED.
Patient problems: No Information
MAUDE report 9400844 ↗InjuryOct 25, 2019Loose or Intermittent Connection
REVISION FOLLOWING THE LOOSENING OF THE STEM 8 MONTHS AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Inadequate Osseointegration
MAUDE report 9237117 ↗InjuryJul 10, 2019Loose or Intermittent Connection
TWO DAYS AFTER PRIMARY SURGERY, THE STEM WAS REVISED DUE TO SUBSIDENCE. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 8775238 ↗InjuryDec 14, 2018Loose or Intermittent Connection; Patient-Device Incompatibility
THE PATIENT CAME IN COMPLAINING OF PAIN 4 YEARS AND 8 MONTHS AFTER THE PRIMARY. THE STEM WAS LOOSE AND MICRO MOVEMENT HAD CAUSED HETEROTOPIC OSSIFICATION.
Patient problems: Inadequate Osseointegration
MAUDE report 8164773 ↗InjuryNov 7, 2018Device Slipped
REVISION SURGERY PERFORMED 5 YEARS AFTER PRIMARY DUE TO INTERMITTENT SHARP THIGH PAIN. DURING SURGERY IT WAS DISCOVERED THAT THE STEM WAS BONDED DISTALLY AND LOOSE PROXIMALLY. STEM AND BALL HEAD WERE REVISED.
Patient problems: Joint Disorder
MAUDE report 8045562 ↗InjuryAug 17, 2017Fracture
AFTER PRIMARY SURGERY, IT WAS NOTICED THAT THE FEMUR HAD BEEN FRACTURED: STEM, HEAD AND LINER REVISED IN AMIS APPROACH ON THE FOLLOWING DAY WITH ITEMS OF THE SAME REFERENCES. MOREOVER, THE SURGEON ADDED 2 SCREWS TO SECURE THE CUP MORE.
Patient problems: Bone Fracture(s)
MAUDE report 6800232 ↗InjuryAug 3, 2017Device Operates Differently Than Expected
REVISION SURGERY 1 MONTH AFTER PRIMARY DUE TO BONE FRACTURE AND STEM SUBSIDENCE. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Bone Fracture(s)
MAUDE report 6763328 ↗InjuryAug 3, 2017Adverse Event Without Identified Device or Use Problem
REVISION OF ALL THE IMPLANTS DUE TO SUSPECTED INFECTION 1 YEAR AND 1 MONTH AFTER PRIMARY. THE SURGEON IMPLANTED A STEM AND DM LINER AND METAL HEAD DURING THE FIRST STEP OF THE REVISION SURGERY, AS THERE WAS NOT ANY CEMENT SPACERS AVAILABLE AT THE HOSPITAL. NO CUP WAS IMPLANTED. WHEN THE INFECTION IS CLEARED THESE COMPONENTS WILL BE REMOVED AND THE DEFINITIVE ONES IMPLANTED.
Patient problems: Unspecified Infection
MAUDE report 6762725 ↗InjuryJun 27, 2017Adverse Event Without Identified Device or Use Problem
THE PATIENT CAME IN COMPLAINING OF PAIN. THE PATIENT FRACTURED HER FEMUR WHILE WALKING. THE SURGEON DETERMINED THE PATIENT HAD POOR BONE QUALITY. THE SURGEON SWAPPED THE BIPOLAR HEAD, COCR HEAD AND THE STEM. THE SURGERY WAS COMPLETED SUCCESSFULLY. X-RAYS AND EXPLANTS ARE NOT AVAILABLE.
Patient problems: Bone Fracture(s); Pain
MAUDE report 6669388 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K082792, K250450. One submission often covers many models, so check the product and lot details in each record.
- Z-0791-2013 · Dec 17, 2012 · TerminatedMedacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page