Skip to content

AMIStem H Femoral Stems 01.18.132Recalls & reported problems

Medacta International SA · AMIStem-H Std. # 2 · Class 2 · Implant

All models of AMIStem H Femoral Stems (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medacta AMIStem-H femoral stem, a hip replacement component placed in the thigh bone, has 115 adverse event reports on file. Many describe the stem loosening or failing to bond with the bone, often leading to pain and revision surgery. The reports include 1 death and 114 injuries, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
115
Deaths reported
1
Injuries reported
114
Recalls naming it
0

Summary

The FDA's MAUDE database contains 115 adverse event reports that list this exact device (DI 07630030804083), received between Jul 13, 2017 and Jul 8, 2026 — classified as 1 death, 114 injuries.

The most frequently reported specific device problem is “Osseointegration Problem” (24 reports).

1 other recall cover devices cleared under the same FDA submission (K093944, K250450); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 3 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Osseointegration Problem24
  • Appropriate Device Problem Term/Code Not Available21
  • Loose or Intermittent Connection21
  • Loosening of Implant Not Related to Bone-Ingrowth16
  • Device Slipped8
  • Failure to Osseointegrate7
  • Fracture5
  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • Inadequate Osseointegration42
  • Joint Disorder21
  • Pain19
  • Aseptic loosening9
  • Bone Fracture(s)7
  • No Code Available7
  • Unspecified Infection6
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 115 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 8, 2026Osseointegration Problem

    AT ABOUT 12 YEARS AND 6 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THERE WAS NO INDICATION OF TRAUMA. THE SURGEON REVISED THE MEDACTA STEM AND HEAD TO A COMPETITOR STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Aseptic loosening; Implant Pain; Inadequate Osseointegration

    MAUDE report 25747720 ↗
  • InjuryMay 12, 2026Osseointegration Problem

    AT ABOUT 9 YEARS AFTER THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THERE WERE NO INDICATIONS OF TRAUMA OR FALL. THE SURGEON REVISED THE STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Aseptic loosening; Inadequate Osseointegration

    MAUDE report 25148938 ↗
  • InjuryMay 5, 2026Osseointegration Problem

    AT ABOUT 8 YEARS AND 10 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THERE WERE NO INDICATIONS OF TRAUMA OR FALL. THE SURGEON REVISED THE MEDACTA HOODED D 32/D LINER TO A FLAT D 32/D LINER AND REVISED THE MEDACTA STEM AND HEAD TO A COMPETITOR STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Aseptic loosening; Inadequate Osseointegration

    MAUDE report 25077231 ↗
  • InjuryApr 20, 2026Implant subsidence

    AT ABOUT 10 YEARS AND 4 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN PRESENTING PAIN DUE TO A SUBSIDED STEM AND THE CAUSE IS UNKNOWN. THE PATIENT REPORTED NO TRAUMA. THE SURGEON REVISED THE MEDACTA LINER TO A MEDACTA LINER AND REVISED THE MEDACTA STEM AND HEAD TO A COMPETITOR STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 24924647 ↗
  • InjuryMar 31, 2026Implant subsidence

    AT ABOUT 9 YEARS AND 1 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN DUE TO PAIN ASSOCIATED WITH A SUBSIDED STEM AND THE CAUSE IS UNKNOWN. THERE WAS NO TRAUMA INDICATED. THE SURGEON REVISED THE MEDACTA STEM AND HEAD TO A COMPETITOR STEM AND HEAD. THE SURGEON ALSO REVISED THE D 32 LINER TO A SAME SIZE LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 24738820 ↗
  • InjuryMar 16, 2026Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY DUE TO STEM LOOSENING AT ABOUT 10 YEARS AND 3 MONTHS POST PRIMARY. THE PATIENT HAD A PRIMARY SURGERY IN GERMANY ON BOTH THE HIP AND THE CONTRALATERAL HIP WAS ALREADY REVISED IN (B)(6) 2025 FOR STEM LOOSENING.

    Patient problems: Aseptic loosening

    MAUDE report 24603551 ↗
  • InjuryJan 29, 2026Device Appears to Trigger Rejection

    AT ABOUT 13 YEARS AND 9 MONTHS FROM THE PRIMARY,THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION.THE SURGEON PERFORMED A REVISION OF THE FEMORAL HEAD, STEM, ACETABULAR CUP, AND LINER.

    Patient problems: Medical device site infection

    MAUDE report 24208608 ↗
  • InjuryJan 29, 2026Osseointegration Problem

    AT ABOUT13 YEARS AND 3 MONTHSFROM THE PRIMARY, THE PATIENT CAME IN PRESENTING HIP PAIN AND A LOOSE STEM WAS DIAGNOSED AFTER X-RAYS AND A CT SCAN. THE SURGEON REVISED THE MEDACTA LINER TO A MEDACTA LINER AND REVISED THE MEDACTA STEM AND HEAD TO A COMPETITOR STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Aseptic loosening; Inadequate Osseointegration

    MAUDE report 24208602 ↗
  • InjuryNov 21, 2025Osseointegration Problem; Osseointegration Problem

    REVISION SURGERY DUE TO HIP STEM LOOSENING.THE SPECIFIC DATE OF THE PRIMARY SURGERY IS UNKNOWN; IT IS PERFORMED IN 2015 OR 2018.

    Patient problems: Aseptic loosening; Inadequate Osseointegration

    MAUDE report 23615337 ↗
  • InjuryNov 18, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    REVISION SURGERY FOR STEM LOOSENING AT ABOUT 10 YEARS AND 5 MONTHS POST PRIMARY. ALL DEVICES REVISED SUCCESSFULLY. NO SIGNS OF INFECTION FOUND IN THE HIP.

    Patient problems: Aseptic loosening

    MAUDE report 23579952 ↗
  • InjuryNov 10, 2025Osseointegration Problem; Osseointegration Problem

    AT ABOUT 10 YEARS FROM THE PRIMARY, REVISION SURGERY NEEDED DUE TO A LOOSE STEM. THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE MEDACTA STEM, HEAD AND CUP WITH COMPETITORS AND THE LINER WITH A NEW MEDACTA LINER (AN HOODED LINER INSTEAD OF THE FLAT LINER THAT WAS ORIGINALLY IMPLANTED).THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Aseptic loosening; Inadequate Osseointegration

    MAUDE report 23510644 ↗
  • InjurySep 8, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    ON (B)(6) 2021, THE PATIENT WAS ENROLLED IN THE AMIS CLINICAL TRIAL AND UNDERWENT PRIMARY SURGERY ON(B)(6) 2023, THE PATIENT REPORTED A SUDDEN CHANGE IN THE HIP, AND PROSTHETIC STEM LOOSENING WAS IDENTIFIED. ON (B)(6) 2024, THE SURGEON REVISED THE CUP AND THE STEM.

    Patient problems: Aseptic loosening

    MAUDE report 22991545 ↗
  • InjuryJul 23, 2025Loss of Osseointegration; Loss of Osseointegration

    REVISION FOR STEM LOOSENING AT ABOUT 9 YEARS AND 5 MONTHS POST-PRIMARY. STEM AND HEAD REVISED SUCCESSFULLY. NO INFECTION. NO LOOSENING OF THE ACETABULAR SHELL.

    Patient problems: Inadequate Osseointegration

    MAUDE report 22574586 ↗
  • InjuryMay 6, 2025Osseointegration Problem; Osseointegration Problem

    AT ABOUT 11 YEARS AND 8 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A POTTED STEM AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 21968714 ↗
  • InjuryMar 6, 2025Osseointegration Problem; Osseointegration Problem

    AT ABOUT 9 YEARS FROM PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH COMPETITOR COMPONENTS AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 21534691 ↗
  • InjuryNov 14, 2024Osseointegration Problem; Osseointegration Problem

    AT ABOUT 10 YEARS AND 7 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A DISTALLY POTTED STEM AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE STEM, HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Implant Pain; Inadequate Osseointegration

    MAUDE report 20681552 ↗
  • InjuryNov 6, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    AT ABOUT 7 YEARS 5 MONTHS AFTER THE PRIMARY, THE PATIENT COMPLAINED OF THIGH PAIN, AND UPON INSPECTION, A RADIOLUCENT LINE WAS FOUND TO HAVE DEVELOPED AND LOOSENING OCCURRED IN THE PROXIMAL PART. IT IS SUPPOSED THAT DISTAL LOCKING OF THE STEM HAD CAUSED BONE THICKENING, WHICH HAD RESULTED IN ATROPHY OF THE PROXIMAL PART. THE STEM, CANCELLOUS BONE SCREW, HEAD AND LINER WERE REVISED. BECAUSE THE CUP WAS SECURELY FIXED, THE SCREW WAS REMOVED AND NOT REPLACED. THE AMISTEM H WAS REMOVED AND AN AMIS-C WAS IMPLANTED, AND THE THICKENED BONE AREA WAS EXCAVATED WITH A FLEXIBLE REAMER.

    Patient problems: Implant Pain

    MAUDE report 20619234 ↗
  • InjuryMay 24, 2024Osseointegration Problem; Osseointegration Problem

    THE PATIENT HAD PAIN DUE TO A DISTALLY POTTED STEM AND THE CAUSE IS UNKNOWN. AT 10 YEARS AND 9 MONTHS POST PRIMARY THE SURGEON REVISED THE STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration; Pain

    MAUDE report 19392539 ↗
  • InjuryMay 16, 2024Failure to Osseointegrate; Failure to Osseointegrate

    AT ABOUT 7 YEARS AND 8 MONTHS AFTER PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED SUCCESSFULLY THE MEDACTA STEM AND HEAD WITH A COMPETITOR STEM AND HEAD AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER.

    Patient problems: Pain

    MAUDE report 19329133 ↗
  • InjuryMay 13, 2024Osseointegration Problem; Osseointegration Problem

    REVISION SURGERY DUE TO STEM LOOSENING, AT ABOUT 8 YEARS AFTER PRIMARY. STEM, HEAD AND INSERT SUCCESSFULLY REVISED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 19297227 ↗
  • InjuryMar 23, 2024Osseointegration Problem; Osseointegration Problem

    REVISION SURGERY WAS PERFORMED ABOUT 5 YEARS AND 9 MONTHS AFTER THE PRIMARY SURGERY DUE TO STEM LOOSENING.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18965455 ↗
  • InjuryJan 30, 2024Osseointegration Problem; Osseointegration Problem

    AT 8 YEARS AND 1 MONTH FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE STEM AND HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Pain

    MAUDE report 18604830 ↗
  • InjuryJan 5, 2024Osseointegration Problem; Osseointegration Problem

    AT ABOUT 9 YEARS AND 3 MONTHS AFTER THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A POTTED STEM AND THE CAUSE IS UNKNOWN. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH COMPETITOR PRODUCTS AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18450190 ↗
  • InjuryOct 27, 2023Osseointegration Problem; Osseointegration Problem

    AT ABOUT 8 YEARS AND 11 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE MEDACTA HEAD AND STEM WITH COMPETITOR COMPONENTS AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 18020055 ↗
  • InjuryOct 10, 2023Failure to Osseointegrate; Failure to Osseointegrate

    AT 9 YEARS AND 11 MONTHS FROM THE PRIMARY, THE PATIENT CAME IN REPORTING PAIN DUE TO A LOOSE STEM AND THE CAUSE OF THE LOOSE STEM IS UNKNOWN. THE SURGEON REVISED THE MEDACTA STEM AND HEAD WITH A COMPETITOR STEM AND HEAD AND REVISED THE MEDACTA LINER WITH A MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 17904608 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K093944, K250450. One submission often covers many models, so check the product and lot details in each record.

  • Z-0791-2013 · Dec 17, 2012 · Terminated
    Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page