Skip to content

AMIStem-H Proximal Coating Femoral Stems 01.18.174Recalls & reported problems

Medacta International SA · AMIStem-H Proximal Coating Std. #4 · Class 2 · Implant

All models of AMIStem-H Proximal Coating Femoral Stems (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 07630030804397), received between Dec 7, 2017 and Sep 9, 2026 — classified as 6 injuries.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available2
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1
  • Loose or Intermittent Connection1
  • Loosening of Implant Not Related to Bone-Ingrowth1
  • Malposition of Device1

Patient problems

  • Hip Fracture2
  • Inadequate Osseointegration2
  • Bone Fracture(s)1
  • Joint Disorder1
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem

    Revision surgery due to a periprosthetic fracture after about 9 years and 2 months from primary. The surgeon cabled the fracture and revised the Flat PE HC liner D 32/F to a Flat PE HC liner D 40/F at his discretion.

    Patient problems: Bone Fracture(s)

    MAUDE report 50692303 ↗
  • InjuryAug 3, 2020Appropriate Device Problem Term/Code Not Available

    THE HIP OF THE PATIENT WAS REVISED ON THE (B)(6) 2020 (2 MONTHS AFTER PRIMARY) DUE TO THE LOOSENING OF THE HIP STEM. INITIALLY, HE WANTED TO CHANGE ONLY THE STEM, BUT SINCE THE SURGEON DID A PUNCTURE AND DISCOVERED A LOW-GRADE INFECTION, HE DECIDED ALSO TO CHANGE THE CUP AND THE LINER. NO MORE INFORMATION ABOUT THE INFECTION AGENT IS AVAILABLE. IT IS UNKNOWN IF THE LOOSENING IS RELATED TO THE INFECTION.

    Patient problems: Inadequate Osseointegration; Unspecified Infection

    MAUDE report 10354963 ↗
  • InjurySep 19, 2019Loose or Intermittent Connection

    THE PATIENT CAME IN COMPLAINING OF PAIN DUE TO A LOOSE STEM. THE SURGEON REVISED THE LINER AND THE STEM 1 YEAR 9 MONTHS AND 1 WEEK AFTER PRIMARY. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 9086482 ↗
  • InjuryJul 30, 2019Malposition of Device; Loosening of Implant Not Related to Bone-Ingrowth

    ON THE (B)(4) 2019 WE WERE INFORMED OF A REVISION SURGERY PERFORMED ON THE (B)(6) 2019, ABOUT 4 MONTHS AND A HALF FROM THE PRIMARY FOR STEM LOOSENING AND CUP MALPOSITIONING. FEMUR STRESS SHIELDING WAS DETECTED. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Inadequate Osseointegration

    MAUDE report 8840207 ↗
  • InjuryOct 26, 2018Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY 9 DAYS AFTER PRIMARY DUE TO PROXIMAL FEMUR BONE FRACTURE. THE SURGEON REVISED THE STEM AND HEAD WITH ANOTHER COMPANY'S PRODUCT AND THE LINER WITH ANOTHER MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Hip Fracture

    MAUDE report 8008972 ↗
  • InjuryDec 7, 2017Fracture

    DURING STEM IMPLANTATION THE BONE FEMUR WAS FRACTURED, THE FRACTURE WAS TREATED WITH A CERCLAGE. THE AMISTEM-H PROXIMAL COATING WAS REPLACED WITH A CEMENTED STEM.

    Patient problems: Hip Fracture

    MAUDE report 7095271 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →

Sources and updates