Skip to content

Double Mobility Liner 01.26.2246MHCRecalls & reported problems

Medacta International SA · HC PE Liner 22 / DMC · Class 2 · Implant

All models of Double Mobility Liner (20)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07630030806834), received between Jul 27, 2018 and Dec 27, 2024 — classified as 8 injuries.

The most frequently reported specific device problem is “Device Appears to Trigger Rejection” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available3
  • Adverse Event Without Identified Device or Use Problem2
  • Device Appears to Trigger Rejection2
  • Microbial Contamination of Device1

Patient problems

  • Unspecified Infection5
  • Bone Fracture(s)1
  • Joint Dislocation1
  • Joint Laxity1
  • Medical device site infection1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 27, 2024Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    ALMOST ONE MONTH AFTER PRIMARY THE PATIENT DEVELOPED AN INFECTION. A WASHOUT WAS PERFORMED AND LINER AND HEAD WERE REVISIONED. SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Medical device site infection

    MAUDE report 21026659 ↗
  • InjuryJun 25, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    HIP REVISION SURGERY AT ABOUT 9 YEARS AND 10 MONTHS. THE PATIENT REPORTED THAT THE JOINT WAS NOT STABLE WHEN RIDING A BIKE. DURING THE REVISION, A MICRO-FRACTURE WAS FOUND IN THE ACETABULUM WHEN THE CUP WAS REMOVED AND SIGNS OF WEAR ON THE POLY WERE DETECTED. NO INFECTION WAS DETECTED. THE SURGEON REVISED THE CUP, LINER, AND HEAD, TO ADD OFFSET AND A BIT OF LEG LENGTH TO INCREASE PATIENT'S STABILITY.

    Patient problems: Bone Fracture(s); Joint Laxity

    MAUDE report 19606526 ↗
  • InjuryJun 13, 2023Device Appears to Trigger Rejection; Device Appears to Trigger Rejection

    THE PATIENT CAME IN DUE TO SIGNS OF AN INFECTION AND THE PATHOGEN IS UNKNOWN. ABOUT 1 MONTH AFTER THE PRIMARY SURGERY, THE SURGEON PERFORMED A WASHOUT AND REVISED THE HEAD AND LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 17117954 ↗
  • InjuryDec 7, 2020Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY FOR DISLOCATION OF THE DM POLY LINER FROM THE DM CONVERTER (DM POLY LINER AND COCR HEAD REMAINED COUPLED). THE SURGEON REVISED SUCCESSFULLY THE LINERS AND COCR HEAD.

    Patient problems: Joint Dislocation

    MAUDE report 10959367 ↗
  • InjuryJul 5, 2019Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY PERFORMED AFTER 1 YEAR AND 7 MONTHS FORM THE PRIMARY DUE TO AN INFECTION (THE TYPE OF THE PATHOGEN IS UNKNOWN). THE INLAY HAS BEEN REVISED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 8764331 ↗
  • InjuryMay 27, 2019Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY PERFORMED 1 YEAR AFTER PRIMARY DUE TO INFECTION. CUP AND LINER HAVE BEEN REMOVED, THE PATHOGEN IS UNKNOWN.

    Patient problems: Unspecified Infection

    MAUDE report 8644140 ↗
  • InjuryApr 25, 2019Adverse Event Without Identified Device or Use Problem

    1 MONTH AFTER PRIMARY REVISION SURGERY FOR INFECTION. THE SURGEON PERFORMED AN I&D, REVISED ANOTHER COMPANY'S HEAD WITH ANOTHER COMPANY'S HEAD AND REVISED THE LINER WITH A MEDACTA LINER. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 8548594 ↗
  • InjuryJul 27, 2018Microbial Contamination of Device

    ABOUT 1 MONTH AFTER PRIMARY THE SURGEON REVISED THE PATIENT FOR A INFECTION CASE. LINER SWAP. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 7726664 ↗

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →

Sources and updates