Versafitcup CC TRIO Acetabular System 01.26.3648HCTRecalls & reported problems
Medacta International SA · Flat PE HC liner Ø 36 / F · Class 2 · Implant
All models of Versafitcup CC TRIO Acetabular System (32)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 07630030807688), received between Mar 9, 2018 and Feb 12, 2026 — classified as 10 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available4
- Adverse Event Without Identified Device or Use Problem3
- Device Dislodged or Dislocated2
- Device Markings/Labelling Problem1
- Inadequacy of Device Shape and/or Size1
- Insufficient Device Problem Information1
- Naturally Worn1
Patient problems
- Unspecified Infection3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- Joint Dislocation2
- Pain2
- Aseptic loosening1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryFeb 12, 2026Adverse Event Without Identified Device or Use Problem
REVISION SURGERY AT ABOUT 13 YEARS AND 8 MONTHS POST PRIMARY FOR STEM LOOSENING. THE LINER WAS FOUND WORN; IT IS UNKNOWN IF THE LINER WORN CONTRIBUTED TO THE STEM LOOSENING.
Patient problems: Aseptic loosening
MAUDE report 24321110 ↗InjuryAug 21, 2024Naturally Worn; Naturally Worn
AT APPROXIMATELY 12 YEARS AND 4 MONTHS AFTER PRIMARY SURGERY, FEMORAL OSTEOLYSIS AND POSSIBLE LINER WEAR WERE DETECTED. REVISION SURGERY WAS PERFORMED IN WHICH THE HEAD AND LINER WERE REMOVED AND REPLACED WITH CERAMIC HEAD AND LINER, WHILE THE STEM AND CUP WERE NOT REVISED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 20022567 ↗InjuryJul 9, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
REVISION SURGERY DUE TO PE HC LINER WEAR, AT ABOUT FIVE YEARS FROM THE PRIMARY. THE PRIMARY SURGERY DATE IS UNKNOWN. THE SURGEON REVISED SUCCESSFULLY THE LINER AND THE COMPETITOR STEM.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 19700408 ↗InjurySep 4, 2020Device Dislodged or Dislocated; Appropriate Device Problem Term/Code Not Available
REVISION SURGERY PERFORMED DUE TO INFECTION, 20 DAYS AFTER THE PREVIOUS REVISION SURGERY (PREVIOUSLY LINER AND HEAD WERE REVISED DUE TO DISLOCATION). THE INLAY WAS A MEDACTA PRODUCT.
Patient problems: Unspecified Infection
MAUDE report 10493017 ↗InjuryAug 25, 2020Appropriate Device Problem Term/Code Not Available
AMIS REVISION SURGERY PERFORMED AFTER 4 YEARS AND 9 MONTHS DUE TO PAIN. THE SURGEON REMOVED THE HEAD AND INLAY, AFTER THAT HE HIT THE STEM A FEW TIMES WITH AN PESTLE TO LOOK IF THE STEM IS LOOSENED. THE STEM WAS NOT LOOSENED. THE SURGEON IMPLANTED A NEW INLAY LINER F 32MM AND AN 32MM XXL COCR HEAD.
Patient problems: Pain
MAUDE report 10447984 ↗InjuryApr 15, 2020Adverse Event Without Identified Device or Use Problem
REVISION SURGERY PERFORMED AFTER 7 YEARS AND 3 MONTHS FROM THE PRIMARY. THE SURGEON REVISED THE CUP, HEAD AND LINER AND REPLACE THEM WITH A DUAL MOBILITY CUP DUE TO DISLOCATION THE HEAD FROM THE LINER DUE TO THE PATIENT WEIGHT. THE STEM WAS NOT REVISED.
Patient problems: Joint Dislocation
MAUDE report 9961668 ↗InjuryJan 9, 2020Adverse Event Without Identified Device or Use Problem
REVISION OF ALL HIP COMPONENTS, 8 MONTHS AFTER PRIMARY SURGERY, DUE TO STREPTOCOCCOSIS INFECTION.
Patient problems: Unspecified Infection
MAUDE report 9568608 ↗InjuryMar 29, 2019Device Dislodged or Dislocated
ABOUT 3 MONTHS AFTER PRIMARY REVISION THE SURGEON REVISED THE PATIENT FOR HIP DISLOCATION (HEAD FROM LINER) DUE TO A GLUTEAL REPAIR RUPTURE. REVISION HAS BEEN PERFORMED SUCCESSFULLY USING COMPETITOR CONSTRAINED CUP AND LINER AND MEDACTA HEAD. SURGEON COMMENTED THAT THERE WAS GREAT BONE IN GROWTH ON CUP AND NEEDED EXPLANT TO GET OUT.
Patient problems: Joint Dislocation
MAUDE report 8464662 ↗MalfunctionOct 11, 2018Inadequacy of Device Shape and/or Size; Device Markings/Labelling Problem
DURING SURGERY THE SURGEON WENT TO USE AN IMPLANT THAT WAS PACKAGED IN A BOX LABELLED AS A 36/F LINER. WHEN THE SURGEON IMPLANTED THE LINER INTO A 56MM CC TRIO CUP, THE LINER SEEMED TO BE TOO SMALL. UPON FURTHER INVESTIGATION, THE AGENT AND SURGEON NOTICED THAT THE BOX AND STICKERS WERE LABELLED WITH 36/F YET THE IMPLANT AND IMPLANT BLISTER PACK WAS LABELLED 32/E LINER. THE AGENT USED A SECOND IMPLANT TO COMPLETE THE CASE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7955804 ↗InjurySep 26, 2018Appropriate Device Problem Term/Code Not Available
REVISION SURGERY PERFORMED DUE TO INFECTION AFTER ABOUT 1 MONTH FROM THE PRIMARY. THE PATHOGEN IS NOT KNOWN. HEAD AND INLAY WERE REVISED.
Patient problems: Unspecified Infection
MAUDE report 7909562 ↗InjuryMar 9, 2018Insufficient Device Problem Information
ON (B)(6) 2018 MEDACTA HAS BEEN INFORMED BY THE SALES AGENT THAT THE PATIENT CAME IN COMPLAINING OF PAIN. THE CAUSE OF PAIN IS UNKNOWN AT THIS TIME. ON (B)(6) 2018 SURGEON REVISED THE PATIENT.
Patient problems: Pain
MAUDE report 7326632 ↗
Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page