Skip to content

Versafitcup CC TRIO Acetabular System 01.26.45.1154Recalls & reported problems

Medacta International SA · Versafitcup Metalback CC Trio No-hole Ø54 mm · Class 2 · Implant

All models of Versafitcup CC TRIO Acetabular System (32)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630030807909), received between Sep 8, 2017 and Jul 26, 2023 — classified as 5 injuries.

The most frequently reported specific device problem is “Loose or Intermittent Connection” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Loose or Intermittent Connection3
  • Appropriate Device Problem Term/Code Not Available2
  • Osseointegration Problem1

Patient problems

  • Hip Fracture1
  • Inadequate Osseointegration1
  • Joint Dislocation1
  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 26, 2023Osseointegration Problem; Osseointegration Problem

    AT ABOUT 4 MONTHS AFTER THE PRIMARY, REVISION SURGERY DUE TO CUP MOBILIZATION IDENTIFIED ON POST-OPERATIVE X-RAYS. THE SURGEON REMOVED THE CUP, REAMED UP, IMPLANTED SUCCESSFULLY A SIZE BIGGER VERSAFITCUP CC TRIO CUP AND SECURED IT WITH 1 SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17396540 ↗
  • InjurySep 15, 2021Loose or Intermittent Connection; Appropriate Device Problem Term/Code Not Available

    5 YEARS AND 9 MONTHS AFTER THE PRIMARY SURGERY THE PATIENT CAME IN DUE TO A LOOSE CUP AND THE CAUSE OF THE LOOSE CUP IS UNKNOWN. THE SURGEON REVISED THE CUP AND LINER.

    Patient problems: Inadequate Osseointegration

    MAUDE report 12473172 ↗
  • InjuryDec 18, 2019Loose or Intermittent Connection

    5 DAYS POST-OP PATIENT PRESENTED WITH ACETABULAR FRACTURE OF THE MEDIAL WALL CAUSING THE CUP LOOSENING. ALL IMPLANTS WERE REVISED. THERE IS NO DEFINITIVE REASON FOR THE FRACTURE.

    Patient problems: Hip Fracture

    MAUDE report 9489460 ↗
  • InjuryMar 6, 2019Loose or Intermittent Connection

    REVISION SURGERY PERFORMED 1 YEAR AND 8 MONTHS AFTER PRIMARY DUE TO CUP AND STEM LOOSENING CAUSED BY INFECTION, PROPIONIBACTERIUM AVIDUM.

    Patient problems: Unspecified Infection

    MAUDE report 8394838 ↗
  • InjurySep 8, 2017Appropriate Device Problem Term/Code Not Available

    REVISION SURGERY DUE TO HIP LUXATION.

    Patient problems: Joint Dislocation

    MAUDE report 6851064 ↗

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page