Skip to content

Versafitcup Double Mobility Shells 01.26.58MBRecalls & reported problems

Medacta International SA · Acetabular shell cementless Ø 58 · Class 2 · Implant

All models of Versafitcup Double Mobility Shells (9)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630030808203), received between Nov 24, 2017 and May 13, 2026 — classified as 5 injuries.

The most frequently reported specific device problem is “Device Slipped” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K083116); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Device Slipped1
  • Difficult to Remove1
  • Malposition of Device1

Patient problems

  • Joint Disorder2
  • Implant Pain1
  • No Code Available1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 13, 2026Malposition of Device

    AT 1 YEAR 9 MONTHS FROM MEDACTA PRIMARY SURGERY, THE PATIENT PRESENTED WITH PAIN. PROBABLE ILIOPSOAS IMPINGEMENT WAS SUSPECTED DUE TO A SLIGHTLY OVERHANGING ACETABULAR CUP. THERE WAS NO EVIDENCE OF COMPONENT LOOSENING OR INFECTION. THE SURGEON REVISED THE MEDACTA CUP AND DM LINER WITH A REINFORCEMENT CAGE (SIZE 62) WITH SCREWS, VERSACEM 56 CUP AND DMH LINER.

    Patient problems: Implant Pain

    MAUDE report 25160331 ↗
  • InjurySep 3, 2020Difficult to Remove

    THE SURGEON WANTED TO IMPLANT A VERSAFITCUP DM S58 AND AFTER HE HAD A PRESS-FIT THE IMPACTING RING DID NOT COME OFF. BY TRYING SEVERAL TIMES AFTERWARD ALSO WITH A VERSAFITCUP CC TRIO HE HAD TO CEMENT A NEW ONE- APRICOT. THE BONE STRUCTURE WAS DAMAGED. THE SURGEON CEMENTED A NEW CUP ON THE BONE DAMAGE.

    Patient problems: No Code Available

    MAUDE report 10487424 ↗
  • InjuryAug 5, 2020Appropriate Device Problem Term/Code Not Available

    THE PATIENT CAME IN REPORTING PAIN DUE TO A MIGRATED CUP THAT AROSE FROM POOR FIXATION 10 YEARS AND 3 MONTHS AFTER THE PRIMARY SURGERY. THE SURGEON REVISED THE CUP AND HEAD AND THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder

    MAUDE report 10366569 ↗
  • InjuryFeb 28, 2019Adverse Event Without Identified Device or Use Problem

    REVISION SURGERY AFTER ONE MONTH AND HALF FROM THE PRIMARY DUE TO SIGNS OF INFECTION (PATHOGEN UNKNOWN). THE CUP AND LINER HAVE BEEN EXPLANTED (AN ANTIBIOTIC SPACER WAS IMPLANTED). THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Unspecified Infection

    MAUDE report 8377329 ↗
  • InjuryNov 24, 2017Device Slipped

    THE PATIENT CAME IN COMPLAINING OF PAIN. THE SURGEON DETERMINED THE CUP WAS LOOSE. THE SURGEON REVISED THE CUP, THE LINER AND THE HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.

    Patient problems: Joint Disorder; Pain

    MAUDE report 7057482 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K083116. One submission often covers many models, so check the product and lot details in each record.

  • Z-2709-2011 · May 23, 2011 · Terminated
    Product Brand Name: Versafitcup Double Mobility Liner Inserter Product Generic Name: Versafitcup Double Mobility Liner Inserter Model Number: 01.26.10.0018 The hip prosthesis is
  • Z-0878-2013 · Jan 15, 2010 · Terminated
    Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.

Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices

All Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices →

Sources and updates