Versafitcup Double Mobility Shells 01.26.58MBRecalls & reported problems
Medacta International SA · Acetabular shell cementless Ø 58 · Class 2 · Implant
All models of Versafitcup Double Mobility Shells (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 07630030808203), received between Nov 24, 2017 and May 13, 2026 — classified as 5 injuries.
The most frequently reported specific device problem is “Device Slipped” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K083116); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Device Slipped1
- Difficult to Remove1
- Malposition of Device1
Patient problems
- Joint Disorder2
- Implant Pain1
- No Code Available1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 13, 2026Malposition of Device
AT 1 YEAR 9 MONTHS FROM MEDACTA PRIMARY SURGERY, THE PATIENT PRESENTED WITH PAIN. PROBABLE ILIOPSOAS IMPINGEMENT WAS SUSPECTED DUE TO A SLIGHTLY OVERHANGING ACETABULAR CUP. THERE WAS NO EVIDENCE OF COMPONENT LOOSENING OR INFECTION. THE SURGEON REVISED THE MEDACTA CUP AND DM LINER WITH A REINFORCEMENT CAGE (SIZE 62) WITH SCREWS, VERSACEM 56 CUP AND DMH LINER.
Patient problems: Implant Pain
MAUDE report 25160331 ↗InjurySep 3, 2020Difficult to Remove
THE SURGEON WANTED TO IMPLANT A VERSAFITCUP DM S58 AND AFTER HE HAD A PRESS-FIT THE IMPACTING RING DID NOT COME OFF. BY TRYING SEVERAL TIMES AFTERWARD ALSO WITH A VERSAFITCUP CC TRIO HE HAD TO CEMENT A NEW ONE- APRICOT. THE BONE STRUCTURE WAS DAMAGED. THE SURGEON CEMENTED A NEW CUP ON THE BONE DAMAGE.
Patient problems: No Code Available
MAUDE report 10487424 ↗InjuryAug 5, 2020Appropriate Device Problem Term/Code Not Available
THE PATIENT CAME IN REPORTING PAIN DUE TO A MIGRATED CUP THAT AROSE FROM POOR FIXATION 10 YEARS AND 3 MONTHS AFTER THE PRIMARY SURGERY. THE SURGEON REVISED THE CUP AND HEAD AND THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder
MAUDE report 10366569 ↗InjuryFeb 28, 2019Adverse Event Without Identified Device or Use Problem
REVISION SURGERY AFTER ONE MONTH AND HALF FROM THE PRIMARY DUE TO SIGNS OF INFECTION (PATHOGEN UNKNOWN). THE CUP AND LINER HAVE BEEN EXPLANTED (AN ANTIBIOTIC SPACER WAS IMPLANTED). THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Unspecified Infection
MAUDE report 8377329 ↗InjuryNov 24, 2017Device Slipped
THE PATIENT CAME IN COMPLAINING OF PAIN. THE SURGEON DETERMINED THE CUP WAS LOOSE. THE SURGEON REVISED THE CUP, THE LINER AND THE HEAD. THE SURGERY WAS COMPLETED SUCCESSFULLY.
Patient problems: Joint Disorder; Pain
MAUDE report 7057482 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K083116. One submission often covers many models, so check the product and lot details in each record.
- Z-2709-2011 · May 23, 2011 · TerminatedProduct Brand Name: Versafitcup Double Mobility Liner Inserter Product Generic Name: Versafitcup Double Mobility Liner Inserter Model Number: 01.26.10.0018 The hip prosthesis is
- Z-0878-2013 · Jan 15, 2010 · TerminatedVersafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.
Other Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.